OBJECTIVE: To evaluate the Abbott CELL-DYN Sapphire cytometer for cerebrospinal fluid (CSF) cell count and differentiation. METHODS: One hundred three analyses of CSF cells by the CELL-DYN Sapphire were compared with routine cell count and microscopic differentiation and correlation coefficients calculated. RESULTS: The total cell count of both methods correlated well. The detection of erythrocytes was good (0.898), and a higher content of erythrocytes >100/microL had little effect on total leukocyte count. The correlation between both methods was best with higher leukocyte counts >25/microL (r=0.987), whereas at cell counts <25/microL, the correlation was considerably less precise (r=0.613). For the differentiation of cells, lymphocytes and neutrophils showed moderate correlation. The results for monocytes and eosinophils did not correlate. CONCLUSION: The results for the total cell count in this study are comparable with those achieved with the Bayer Advia 120. While the Abbott CELL-DYN Sapphire yielded slightly better results for erythrocytes and total cell count with a higher erythrocyte content, the Advia 120 achieved slightly better results of lymphocyte and neutrophil count in a previous study.
OBJECTIVE: To evaluate the Abbott CELL-DYN Sapphire cytometer for cerebrospinal fluid (CSF) cell count and differentiation. METHODS: One hundred three analyses of CSF cells by the CELL-DYN Sapphire were compared with routine cell count and microscopic differentiation and correlation coefficients calculated. RESULTS: The total cell count of both methods correlated well. The detection of erythrocytes was good (0.898), and a higher content of erythrocytes >100/microL had little effect on total leukocyte count. The correlation between both methods was best with higher leukocyte counts >25/microL (r=0.987), whereas at cell counts <25/microL, the correlation was considerably less precise (r=0.613). For the differentiation of cells, lymphocytes and neutrophils showed moderate correlation. The results for monocytes and eosinophils did not correlate. CONCLUSION: The results for the total cell count in this study are comparable with those achieved with the Bayer Advia 120. While the Abbott CELL-DYN Sapphire yielded slightly better results for erythrocytes and total cell count with a higher erythrocyte content, the Advia 120 achieved slightly better results of lymphocyte and neutrophil count in a previous study.