Literature DB >> 18630838

Automated cerebrospinal fluid cytology.

Tanja Heller1, Ingelore Nagel, Bettina Ehrlich, Mathias Bähr, Herwig Strik.   

Abstract

OBJECTIVE: To evaluate the Abbott CELL-DYN Sapphire cytometer for cerebrospinal fluid (CSF) cell count and differentiation.
METHODS: One hundred three analyses of CSF cells by the CELL-DYN Sapphire were compared with routine cell count and microscopic differentiation and correlation coefficients calculated.
RESULTS: The total cell count of both methods correlated well. The detection of erythrocytes was good (0.898), and a higher content of erythrocytes >100/microL had little effect on total leukocyte count. The correlation between both methods was best with higher leukocyte counts >25/microL (r=0.987), whereas at cell counts <25/microL, the correlation was considerably less precise (r=0.613). For the differentiation of cells, lymphocytes and neutrophils showed moderate correlation. The results for monocytes and eosinophils did not correlate.
CONCLUSION: The results for the total cell count in this study are comparable with those achieved with the Bayer Advia 120. While the Abbott CELL-DYN Sapphire yielded slightly better results for erythrocytes and total cell count with a higher erythrocyte content, the Advia 120 achieved slightly better results of lymphocyte and neutrophil count in a previous study.

Mesh:

Year:  2008        PMID: 18630838

Source DB:  PubMed          Journal:  Anal Quant Cytol Histol        ISSN: 0884-6812            Impact factor:   0.302


  4 in total

Review 1.  [Cytology of cerebrospinal fluid : Standards, importance and modern methods].

Authors:  M Wick; C C Gross; S Isenmann; H Strik
Journal:  Nervenarzt       Date:  2016-12       Impact factor: 1.214

2.  Detection of cancer cells in the cerebrospinal fluid: current methods and future directions.

Authors:  Cody L Weston; Michael J Glantz; James R Connor
Journal:  Fluids Barriers CNS       Date:  2011-03-03

3.  Improvement of cell counting method for Neubauer counting chamber.

Authors:  Minghao Zhang; Lingui Gu; Peihua Zheng; Zhixin Chen; Xinqi Dou; Qizhong Qin; Xiaozhong Cai
Journal:  J Clin Lab Anal       Date:  2019-08-30       Impact factor: 2.352

4.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.