A comparison of noninvasive blood pressure measurement on the wrist with invasive arterial blood pressure monitoring in patients undergoing bariatric surgery.

Helmut Hager1, Goutham Mandadi, Debra Pulley, J Chris Eagon, Edward Mascha, Benjamin Nutter, Andrea Kurz.   

Abstract

BACKGROUND: In morbidly obese patients, oscillometric blood pressure measurements with an upper-arm cuff are often difficult to perform. The alternative method, invasive blood pressure monitoring, can be difficult to place and is associated with risks. A wrist-mounted blood pressure-monitoring device, the Vasotrac, provides accurate blood pressure measurements in lean patients. Even in the obese, wrist morphology remains relatively unchanged. We thus assessed the degree to which blood pressure measurements with the Vasotrac on the wrist and cuff measurements agree with invasive arterial blood pressure monitoring.
METHODS: We evaluated 22 morbidly obese patients undergoing bariatric surgery lasting 3.8+/-1.1 h. Intraoperative blood pressure was simultaneously measured using the Vasotrac mounted on one wrist; an arterial catheter was inserted in the opposite radial artery, and an oscillometric cuff was positioned on the upper arm. Preoperative patient comfort was evaluated on a scale from 1 to 10, with 10 being most uncomfortable, just after the first oscillometric cuff inflation. Values from the Vasotrac and arterial catheter were recorded at 5-s intervals. Bias, precision, and clinically acceptable agreement were calculated between the two continuous monitoring devices and between the arterial catheter and the cuff measurements, with the arterial catheter providing the reference value.
RESULTS: The patients' age was 44.3+/-9.5 years (mean+/-SD), body mass index was 66.7+/-13.8 kg/m2, and arm circumference was 48.6+/-7.5 cm. Patients found the Vasotrac more comfortable than the oscillometric device [1.7+/-1.8 vs 5.3+/-0.5 (P=0.009)]. A total of 40,411 pairs of values from the Vasotrac and arterial catheter were recorded. Lin's concordance correlation coefficient (95% CI) for mean arterial blood pressure measured between the arterial line and the Vasotrac was 0.74 (0.67, 0.82). The bias (mean error) was -0.25 mmHg; however, the Bland-Altman limits where 95% of individual pressure differences are expected to fall was (-20, 20) mmHg. The precisions for diastolic and systolic pressures were even worse.
CONCLUSION: The Vasotrac was more comfortable than an oscillometric device. Although the average accuracy was good, individual mean Vasotrac and noninvasive blood pressure pressures often differed considerably from arterial values. These results suggest that the Vasotrac monitor should not be substituted for an arterial catheter in super-obese patients.

Entities:  

Mesh:

Year:  2008        PMID: 18618207     DOI: 10.1007/s11695-008-9607-7

Source DB:  PubMed          Journal:  Obes Surg        ISSN: 0960-8923            Impact factor:   4.129


  19 in total

1.  A new noninvasive method to measure blood pressure: results of a multicenter trial.

Authors:  K Belani; M Ozaki; J Hynson; T Hartmann; H Reyford; J M Martino; M Poliac; R Miller
Journal:  Anesthesiology       Date:  1999-09       Impact factor: 7.892

Review 2.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

Review 3.  Treatment of non-alcoholic steatohepatitis.

Authors:  Paul Angulo; Keith D Lindor
Journal:  Best Pract Res Clin Gastroenterol       Date:  2002-10       Impact factor: 3.043

4.  Obesity: preventing and managing the global epidemic. Report of a WHO consultation.

Authors: 
Journal:  World Health Organ Tech Rep Ser       Date:  2000

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Thermoregulatory and anesthetic-induced alterations in the differences among femoral, radial, and oscillometric blood pressures.

Authors:  J M Hynson; D I Sessler; A Moayeri; J A Katz
Journal:  Anesthesiology       Date:  1994-12       Impact factor: 7.892

7.  Blood pressure monitoring during pregnancy. Accuracy of portable devices designed for obese patients.

Authors:  W J Schwartz; W F Rayburn; G L Turnbull; H D Christensen
Journal:  J Reprod Med       Date:  1996-08       Impact factor: 0.142

8.  Blood pressure. Assessment of interarm differences.

Authors:  A J Singer; J E Hollander
Journal:  Arch Intern Med       Date:  1996-09-23

9.  Fat arms, obesity and choice of blood pressure cuff size in diabetic patients.

Authors:  M D Feher; A Cox; J Foxton; A F Lant
Journal:  Br J Clin Pract       Date:  1995 Nov-Dec

Review 10.  Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine.

Authors:  Bernd Scheer; Azriel Perel; Ulrich J Pfeiffer
Journal:  Crit Care       Date:  2002-04-18       Impact factor: 9.097

View more
  6 in total

1.  Effectiveness of Pharmacological Therapies for Intracranial Hypertension in Children With Severe Traumatic Brain Injury--Results From an Automated Data Collection System Time-Synched to Drug Administration.

Authors:  Steven L Shein; Nikki M Ferguson; Patrick M Kochanek; Hülya Bayir; Robert S B Clark; Ericka L Fink; Elizabeth C Tyler-Kabara; Stephen R Wisniewski; Ye Tian; G K Balasubramani; Michael J Bell
Journal:  Pediatr Crit Care Med       Date:  2016-03       Impact factor: 3.624

2.  Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU.

Authors:  Sara Ribezzo; Eleonora Spina; Stefano Di Bartolomeo; Gianfranco Sanson
Journal:  ScientificWorldJournal       Date:  2014-01-30

3.  Investigation of the accuracy of a noninvasive continuous blood pressure device in different age groups and its ability in detecting hypertension and hypotension: an observational study.

Authors:  Ziwei Wang; Guizhen Chen; Kaizhi Lu; Yuan Zhu; Yan Chen
Journal:  BMC Anesthesiol       Date:  2019-12-05       Impact factor: 2.217

4.  Validation of the pulse decomposition analysis algorithm using central arterial blood pressure.

Authors:  Martin C Baruch; Kambiz Kalantari; David W Gerdt; Charles M Adkins
Journal:  Biomed Eng Online       Date:  2014-07-08       Impact factor: 2.819

5.  Cardiovascular Responses to an Isometric Handgrip Exercise in Females with Prehypertension.

Authors:  Vernon Bond; Bryan H Curry; Richard G Adams; Thomas Obisesan; Sudhakar Pemminati; Vasavi R Gorantla; Kishan Kadur; Richard M Millis
Journal:  N Am J Med Sci       Date:  2016-06

6.  Transdermal delivery of Diltiazem HCl from matrix film: Effect of penetration enhancers and study of antihypertensive activity in rabbit model.

Authors:  Rabinarayan Parhi; Padilam Suresh
Journal:  J Adv Res       Date:  2015-09-11       Impact factor: 10.479

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.