Literature DB >> 18606631

Disposable reagentless electrochemical immunosensor array based on a biopolymer/sol-gel membrane for simultaneous measurement of several tumor markers.

Jie Wu1, Feng Yan, Xiaoqing Zhang, Yuetian Yan, Jinhai Tang, Huangxian Ju.   

Abstract

BACKGROUND: A reagentless sensor array for simultaneous multianalyte testing (SMAT) may enable accurate diagnosis and be applicable for point-of-care testing. We developed a disposable reagentless immunosensor array for simple immunoassay of panels of tumor markers.
METHODS: We carried out SMAT with a direct capture format, in which colloidal gold nanoparticles with bound horseradish peroxidase (HRP)-labeled antibodies were immobilized on screen-printed carbon electrodes with biopolymer/sol-gel to trap their corresponding antigens from sample solution. Upon formation of immunocomplex, the direct electrochemical signal of the HRP decreased owing to increasing spatial blocking, and the analytes could be simultaneously determined by monitoring the signal changes.
RESULTS: The proposed reagentless immunosensor array allowed simultaneous detection of carcinoma antigen 153, carcinoma antigen 125, carbohydrate antigen 199, and carcinoembryonic antigen in clinical serum samples in the ranges of 0.4-140 kU/L, 0.5-330 kU/L, 0.8-190 kU/L, and 0.1-44 microg/L, respectively, with detection limits of 0.2 kU/L, 0.5 kU/L, 0.3 kU/L, and 0.1 microg/L corresponding to the signals 3 SD above the mean of a zero standard. The interassay imprecision of the arrays was <9.5%, and they were stable for 35 days. The positivity detection rate of panels of tumor markers was >95.5% for 95 cases of cancer-positive sera.
CONCLUSIONS: The immunosensor array provides a SMAT with short analytical time, small sampling volume, no need for substrate, and, no between-electrode cross-talk. This method not only proved the capability of the array in point-of-care testing, but also allowed simultaneous testing of several tumor markers.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18606631     DOI: 10.1373/clinchem.2007.102350

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  6 in total

Review 1.  Gold nanoparticles in chemical and biological sensing.

Authors:  Krishnendu Saha; Sarit S Agasti; Chaekyu Kim; Xiaoning Li; Vincent M Rotello
Journal:  Chem Rev       Date:  2012-02-02       Impact factor: 60.622

Review 2.  Targeted hyperthermia using metal nanoparticles.

Authors:  Paul Cherukuri; Evan S Glazer; Steven A Curley
Journal:  Adv Drug Deliv Rev       Date:  2009-11-10       Impact factor: 15.470

Review 3.  Electrochemistry of nonconjugated proteins and glycoproteins. Toward sensors for biomedicine and glycomics.

Authors:  Emil Paleček; Jan Tkáč; Martin Bartošík; Tomáš Bertók; Veronika Ostatná; Jan Paleček
Journal:  Chem Rev       Date:  2015-02-09       Impact factor: 60.622

Review 4.  Gold Nanoparticles for Diagnostics: Advances towards Points of Care.

Authors:  Mílton Cordeiro; Fábio Ferreira Carlos; Pedro Pedrosa; António Lopez; Pedro Viana Baptista
Journal:  Diagnostics (Basel)       Date:  2016-11-22

Review 5.  Nanostructures for Biosensing, with a Brief Overview on Cancer Detection, IoT, and the Role of Machine Learning in Smart Biosensors.

Authors:  Aishwaryadev Banerjee; Swagata Maity; Carlos H Mastrangelo
Journal:  Sensors (Basel)       Date:  2021-02-10       Impact factor: 3.576

Review 6.  Antibody-Electroactive Probe Conjugates Based Electrochemical Immunosensors.

Authors:  Mateusz Kondzior; Iwona Grabowska
Journal:  Sensors (Basel)       Date:  2020-04-03       Impact factor: 3.576

  6 in total