Combination balloon-ultrasound imaging catheter for percutaneous transluminal angioplasty. Validation of imaging, analysis of recoil, and identification of plaque fracture.

J M Isner1, K Rosenfield, D W Losordo, L Rose, R E Langevin, S Razvi, B D Kosowsky.   

Abstract

BACKGROUND: We investigated the hypothesis that an ultrasound transducer positioned within an angioplasty balloon could be used to perform quantitative assessment of arterial dimensions before and after percutaneous transluminal angioplasty (PTA) and to identify certain mechanical alterations consequent to PTA, including vascular wall recoil and the initiation of plaque fractures. METHODS AND
RESULTS: A combination balloon-ultrasound imaging catheter (BUIC) that houses a 20-MHz ultrasound transducer within and halfway between the proximal and distal ends of an angioplasty balloon was used to perform PTA in 10 patients with peripheral vascular disease. Each PTA site was also evaluated before and after PTA by standard (nonballoon) intravascular ultrasound (IVUS) technique. In eight patients in whom satisfactory images were recorded with the BUIC before PTA, luminal cross-sectional area (XSA) of stenotic sites (0.10 +/- 0.01 cm2) did not differ significantly from measurements of XSA by IVUS (0.09 +/- 0.01 cm2, p = NS). Likewise, minimum luminal diameter (Dmin) measured by BUIC (0.34 +/- 0.02 cm) was similar to that measured by IVUS (0.33 +/- 0.01 cm, p = NS). In nine patients in whom satisfactory images were recorded with the BUIC after PTA, XSA measured by BUIC (0.29 +/- 0.03 cm2) did not differ significantly from XSA measured by IVUS (0.30 +/- 0.03 cm2, p = NS). Dmin measured by BUIC after PTA (0.57 +/- 0.02 cm) was also similar to Dmin measured by IVUS (0.57 +/- 0.03 cm, p = NS). After PTA, XSA and Dmin measured immediately after deflation were significantly less than balloon XSA and diameter at full inflation, indicating significant elastic recoil of the dilated site. For the nine patients in whom post-PTA images were satisfactory for quantitative analysis, including four patients in whom recoil was 39%, 46%, 50%, and 61%, percent recoil measured 28.6 +/- 7.2%. Finally, plaque fractures were identified on-line in six of 10 patients (60%); in each case, initiation of plaque fracture was observed at inflation pressures of 2 atm or less.
CONCLUSIONS: The results of this preliminary human investigation indicate that an ultrasound transducer positioned within an angioplasty balloon can be used to perform quantitative and qualitative analyses of lumen-plaque-wall alterations immediately preceding, during, and immediately after PTA in patients with peripheral vascular disease.

Entities:  

Mesh:

Year:  1991        PMID: 1860219     DOI: 10.1161/01.cir.84.2.739

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  9 in total

1.  Dynamics of Vascular Remodeling: An Overview and Bibliography.

Authors: 
Journal:  J Thromb Thrombolysis       Date:  1996       Impact factor: 2.300

2.  Ultrasound guided therapeutic catheters: recent developments and clinical results.

Authors:  R J Crowley; M A Hamm; S H Joshi; C D Lennox; G T Roberts
Journal:  Int J Card Imaging       Date:  1991

3.  Influence of plaque morphology on the mechanism of luminal enlargement after directional coronary atherectomy and balloon angioplasty.

Authors:  F Marsico; J Kubica; S De Servi; L Angoli; E Bramucci; A M Costante; G Specchia
Journal:  Br Heart J       Date:  1995-08

Review 4.  Intracoronary ultrasound: current state of the art.

Authors:  P P Kearney; I R Starkey; G R Sutherland
Journal:  Br Heart J       Date:  1995-05

5.  Ultrasound-monitored laser angioplasty: preliminary clinical results.

Authors:  S H Duda; P E Huppert; A Kreis; K R Karsch; C D Claussen
Journal:  Cardiovasc Intervent Radiol       Date:  1993 Mar-Apr       Impact factor: 2.740

Review 6.  Peripheral angioplasty balloon technology.

Authors:  A H Matsumoto; K H Barth; J B Selby; C J Tegtmeyer
Journal:  Cardiovasc Intervent Radiol       Date:  1993 May-Jun       Impact factor: 2.740

Review 7.  Coronary microembolization--its role in acute coronary syndromes and interventions.

Authors:  R Erbel; G Heusch
Journal:  Herz       Date:  1999-11       Impact factor: 1.443

8.  Quantitative assessment of peripheral and coronary artery lesions before and after balloon angioplasty: a comparison of intravascular ultrasound and angiography.

Authors:  M Schartl; W Bocksch; T Linderer; S Beckmann; S Dreysse
Journal:  Heart Vessels       Date:  1994       Impact factor: 2.037

9.  Recombinant Human Elastase Alters the Compliance of Atherosclerotic Tibial Arteries After Ex Vivo Angioplasty.

Authors:  Steven K Burke; Karen Bingham; Emma Moss; Daniel P Gottlieb; Marco D Wong; Kimberly S Bland; F Nicholas Franano
Journal:  J Cardiovasc Pharmacol       Date:  2016-04       Impact factor: 3.105

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.