The Leksell Gamma Knife Perfexion and comparisons with its predecessors.

Christer Lindquist1, Ian Paddick.   

Abstract

OBJECTIVE: The objective of this study is to introduce a new radiosurgical device, the Leksell Gamma Knife Perfexion (Elekta Instruments AB, Stockholm, Sweden). Design and performance characteristics are compared with previous models of the gamma knife in a clinical setting.
METHODS: Performance-related features in the design of the new radiosurgical system are described, and the ability to create complex shapes of isodose volumes even with a single isocenter is demonstrated. The concept of "dynamic shaping" of dose distribution is introduced as a means of decreasing the exposure to structures outside the target. Dose plans for the Leksell Gamma Knife Models B, 4C, and Perfexion have been created for target pathologies to illustrate how the software and hardware of the new radiosurgical system can increase treatment quality, decrease treatment time, and increase patient comfort. Radiation doses to critical organs in the rest of the human body are compared.
RESULTS: The automated collimator arrangement in the Perfexion makes it possible to produce more complex treatment volumes than with previous models of the gamma knife. This results from the enhanced ability to shape isodose volumes, even for single isocenters. The collimator arrangement and the patient positioning system also allow shorter patient transit times, reducing unwanted radiation exposure during movement between isocenters.
CONCLUSION: The Perfexion exceeds the capabilities of previous gamma knife models in terms of treatment efficiency, conformity, and radiation protection. These qualities enable treatment of larger target volumes, especially close to eloquent areas. Pathologies previously inaccessible in the head and neck are now treatable due to the increased volume of the radiation cavity.

Entities:  

Year:  2008        PMID: 18596433     DOI: 10.1227/01.neu.0000316276.20586.dd

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  11 in total

1.  Stereotactic radiosurgery of glomus jugulare tumors: current concepts, recent advances and future perspectives.

Authors:  Omer Sager; Ferrat Dincoglan; Murat Beyzadeoglu
Journal:  CNS Oncol       Date:  2015

2.  Radiosurgery technology development and use.

Authors:  Sanford L Meeks; Jason Pukala; Naren Ramakrishna; Twyla R Willoughby; Francis J Bova
Journal:  J Radiosurg SBRT       Date:  2011

Review 3.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

4.  Comparison of plan quality and delivery time between volumetric arc therapy (RapidArc) and Gamma Knife radiosurgery for multiple cranial metastases.

Authors:  Evan M Thomas; Richard A Popple; Xingen Wu; Grant M Clark; James M Markert; Barton L Guthrie; Yu Yuan; Michael C Dobelbower; Sharon A Spencer; John B Fiveash
Journal:  Neurosurgery       Date:  2014-10       Impact factor: 4.654

5.  Target and peripheral dose from radiation sector motions accompanying couch repositioning of patient coordinates with the Gamma Knife(®) Perfexion(™).

Authors:  Tuan-Anh Tran; Vincent Wu; Harish Malhotra; James P Steinman; Dheerendra Prasad; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2011-04-23       Impact factor: 2.991

6.  Feasibility of identification of gamma knife planning strategies by identification of pareto optimal gamma knife plans.

Authors:  C A Giller
Journal:  Technol Cancer Res Treat       Date:  2011-12

7.  Simulational study of a dosimetric comparison between a Gamma Knife treatment plan and an intensity-modulated radiotherapy plan for skull base tumors.

Authors:  Hisato Nakazawa; Yoshimasa Mori; Masataka Komori; Takahiko Tsugawa; Yuta Shibamoto; Tatsuya Kobayashi; Chisa Hashizume; Yukio Uchiyama; Masahiro Hagiwara
Journal:  J Radiat Res       Date:  2013-12-17       Impact factor: 2.724

8.  Useful base plate to support the head during Leksell skull frame placement in gamma knife perfexion radiosurgery.

Authors:  Hisato Nakazawa; Yoshimasa Mori; Masahiro Hagiwara; Takahiko Tsugawa; Yuta Shibamoto; Tatsuya Kobayashi; Chisa Hashizume
Journal:  Nagoya J Med Sci       Date:  2014-02       Impact factor: 1.131

9.  Comparison of volumetric-modulated arc therapy and dynamic conformal arc treatment planning for cranial stereotactic radiosurgery.

Authors:  Jessica Molinier; Christine Kerr; Sebastien Simeon; Norbert Ailleres; Marie Charissoux; David Azria; Pascal Fenoglietto
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

10.  Dosimetry and Treatment Descriptions Using the First Completely Automated Stereotactic Intracranial Radiosurgery Rotating Gamma Ray Unit in America.

Authors:  Eduardo E Lovo; Fidel J Campos; Victor E Caceros; Mario H Minervini; William A Reyes
Journal:  Cureus       Date:  2018-04-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.