Literature DB >> 18585199

PVC membrane electrode for the potentiometric determination of Ipratropium bromide using batch and flow injection techniques.

M E M Hassouna1, S A A Elsuccary.   

Abstract

Ipratropium (IP(+)) ion-selective electrode (ISE) has been constructed from poly(vinyl chloride) matrix membrane containing Ipratropium-tetraphenylborate (IP-TPB) as the electroactive component using 2-nitrophenyloctylether as plasticizer. The electrode exhibits near Nernstian response to Ipratropium bromide (IPBr) over the concentration range 10(-5) to 10(-2) mol L(-1) and detection limit 5.1x10(-6) mol L(-1). The electrode offers significant advantages including long lifetime (>2 months), excellent stability and reproducibility, fast response time (<10 s), wide pH working range (pH 2-9), high thermal stability (isothermal coefficient 0.37 mV/degrees C) and superior selectivity for IPBr over a large number of inorganic and organic substances. The electrode was successfully used as indicator electrode in the potentiometric titration of IPBr versus sodium tetraphenylborate (NaTPB) and in the determination of IPBr in Atrovent vials and spiked urine samples applying batch and flow injection techniques, with satisfactory results.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18585199     DOI: 10.1016/j.talanta.2008.01.014

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

Review 1.  Electrochemical sensors.

Authors:  Benjamin J Privett; Jae Ho Shin; Mark H Schoenfisch
Journal:  Anal Chem       Date:  2010-06-15       Impact factor: 6.986

2.  Use of a new ziprasidone-selective electrode in mixed solvents and its application in the analysis of pharmaceuticals and biological fluids.

Authors:  M Soledad García; Joaquín A Ortuño; María Cuartero; Mustafa Salem Abuherba
Journal:  Sensors (Basel)       Date:  2011-09-13       Impact factor: 3.576