Literature DB >> 18583623

Measurement of urinary free cortisol by tandem mass spectrometry and comparison with results obtained by gas chromatography-mass spectrometry and two commercial immunoassays.

Lisa Wood1, David H Ducroq, Helen L Fraser, Scott Gillingwater, Carol Evans, Alan J Pickett, Derek W Rees, Rhys John, Atilla Turkes.   

Abstract

BACKGROUND: Determination of urinary free cortisol (UFC) is an important adjunct for the assessment of adrenal function. In this study, we have analysed cortisol concentrations in urine samples by gas chromatography-mass spectrometry (GC-MS), liquid chromatography-tandem mass spectrometry (LC-MS/MS) and two immunoassays. The results were compared with GC-MS. The interference of cortisol ring-A metabolites in immunoassays was also assessed.
METHODS: The GC-MS technique involved solvent extraction, LH-20 clean-up and derivatization. Only solid-phase extraction procedure was used for LC-MS/MS. The samples were analysed in positive electro-spray ionization mode, monitoring the transitions for cortisol and deuterated-cortisol at m/z 363.3 > 121.2 and m/z 365.3 > 122.2, respectively. Immunoassays were performed according to the manufacturer's instructions.
RESULTS: When compared with GC-MS results both immunoassays (Coat-A-Count; approximately 1.9-fold, Centaur; approximately 1.6-fold) overestimated UFC concentrations. Cortisol ring-A dihydro- and tetrahydrometabolites contribute significantly to this overestimation. There was no interference by these metabolites in either GC-MS or LC-MS/MS methods. The sensitivity of the LC-MS/MS procedure was 2 nmol/L and the intra- and inter-assay variations were <5% in each quality-control sample. The comparison of the UFC results achieved by assaying the study samples with GC-MS and LC-MS/MS indicated that the agreement between the two methods was excellent (LC-MS/MS = 1.0036GC-MS - 0.0841; r2 = 0.9937).
CONCLUSIONS: The interference of cortisol ring-A metabolites in immunoassays contribute to overestimation of UFC concentrations. The LC-MS/MS procedure had the sensitivity, specificity, linearity, precision and accuracy for the determination of UFC concentrations. The method is suitable for routine use provided that method-dependant reference values are established.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18583623     DOI: 10.1258/acb.2007.007119

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  23 in total

1.  Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry.

Authors:  James J Pitt
Journal:  Clin Biochem Rev       Date:  2009-02

2.  Acute Valproate Exposure Induces Sex-Specific Changes in Steroid Hormone Metabolism in the Cerebral Cortex of Juvenile Mice.

Authors:  Sung-Hee Cho; Jung Hoon Chai; Sung-Youn Chang; Soon Ae Kim
Journal:  Neurochem Res       Date:  2020-06-29       Impact factor: 3.996

3.  LC-MS/MS in the Clinical Laboratory - Where to From Here?

Authors:  Stefan Kg Grebe; Ravinder J Singh
Journal:  Clin Biochem Rev       Date:  2011-02

4.  Urine free cortisol in the diagnosis of Cushing's syndrome: is it worth doing and, if so, how?

Authors:  Hershel Raff; Richard J Auchus; James W Findling; Lynnette K Nieman
Journal:  J Clin Endocrinol Metab       Date:  2014-11-25       Impact factor: 5.958

5.  Measurement of urinary free cortisol by LC-MS-MS: adoption of a literature reference range and comparison with our current immunometric method.

Authors:  L Bianchi; B Campi; M R Sessa; G De Marco; E Ferrarini; R Zucchi; C Marcocci; P Vitti; L Manetti; A Saba; P Agretti
Journal:  J Endocrinol Invest       Date:  2019-04-22       Impact factor: 4.256

6.  Quantification of cortisol in human eccrine sweat by liquid chromatography - tandem mass spectrometry.

Authors:  Min Jia; Wade M Chew; Yelena Feinstein; Perry Skeath; Esther M Sternberg
Journal:  Analyst       Date:  2016-03-21       Impact factor: 4.616

7.  Heat-map visualization of gas chromatography-mass spectrometry based quantitative signatures on steroid metabolism.

Authors:  Ju-Yeon Moon; Hyun-Jin Jung; Myeong Hee Moon; Bong Chul Chung; Man Ho Choi
Journal:  J Am Soc Mass Spectrom       Date:  2009-05-05       Impact factor: 3.109

8.  Accuracy of immunoassay and mass spectrometry urinary free cortisol in the diagnosis of Cushing's syndrome.

Authors:  G Aranda; M Careaga; F A Hanzu; I Patrascioiu; P Ríos; M Mora; B Morales-Romero; W Jiménez; I Halperin; G Casals
Journal:  Pituitary       Date:  2016-10       Impact factor: 4.107

9.  A Pilot Study of the Normative Range of Overnight Urinary Free Cortisol Corrected for Creatinine in Children.

Authors:  Ole D Wolthers; Sabine Mersmann; Sanjeeva Dissanayake
Journal:  Clin Drug Investig       Date:  2018-04       Impact factor: 2.859

Review 10.  Steroid assays in paediatric endocrinology.

Authors:  John W Honour
Journal:  J Clin Res Pediatr Endocrinol       Date:  2010-02-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.