ECG-edit function in multidetector-row computed tomography coronary arteriography for patients with arrhythmias.

Hideyuki Matsutani1, Tomonari Sano, Takeshi Kondo, Hitomi Morita, Takehiro Arai, Takako Sekine, Shinichi Takase, Akitsugu Oida, Hiroshi Fukazawa, Masahiko Suguta, Makoto Kondo, Takahide Kodama, Tadaaki Orihara, Norikazu Yamada, Masaharu Tsuyuki, Jagat Narula.   

Abstract

BACKGROUND: ECG-gating is necessary for cardiac computed tomography, but is not suitable for arrhythmias, so the aim of this study was to evaluate the usefulness of the ECG-edit function for this purpose. METHODS AND
RESULTS: Of 1,221 patients undergoing 64-row multidetector-row computed tomography coronary angiography (coronary MDCT), 123 patients (28 atrial fibrillation (Af), 39 premature atrial contractions (PAC), 42 premature ventricular contractions (PVC), 3 PAC + PVC, 10 sinus arrhythmias (SA), and a second-degree atrioventricular block (2 degrees AVB)) had arrhythmias necessitating the ECG-edit function. Short R-R interval was deleted and mid-diastolic phases were selected from the long R-R intervals using the "R+absolute time" method. In the present study, the reconstructed images were evaluated using a triple-grade scale A-C, representing excellent, acceptable, and unacceptable image quality. Image quality, categorized as A, B and C, respectively, was 50%, 36% and 14% for the 28 patients with Af; 56%, 36% and 8% for the 39 PAC patients, and 65%, 33% and 2% in the 42 PVC patients. None of the scans of the PAC + PVC, SA, and 2 degrees AVB patients was ranked as C.
CONCLUSIONS: The ECG-edit function is useful for reconstructing coronary MDCT images in many arrhythmias, and provides clinically acceptable images in most cases.

Entities:  

Mesh:

Year:  2008        PMID: 18577814     DOI: 10.1253/circj.72.1071

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  3 in total

1.  Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography.

Authors:  Yuko Kawaguchi; Shinichiro Fujimoto; Kazuhisa Takamura; Etsuro Kato; Shoko Suda; Rie Matsumori; Makoto Hiki; Kanako K Kumamaru; Hiroyuki Daida
Journal:  Int J Cardiovasc Imaging       Date:  2017-09-12       Impact factor: 2.357

2.  Preliminary study of prospective ECG-gated 320-detector CT coronary angiography in patients with ventricular premature beats.

Authors:  Tong Zhang; Jinquan Bai; Wei Wang; Dan Wang; Baozhong Shen
Journal:  PLoS One       Date:  2012-06-06       Impact factor: 3.240

3.  Lidocaine bolus may facilitate computed tomographic coronary angiography in patients with frequent premature ventricular contractions.

Authors:  Jerzy Pręgowski; Jan Jastrzębski; Cezary Kępka; Mariusz Kruk; Marcin Demkow; Lukasz Kalińczuk; Rafał Wolny; Michał Ciszewski; Ilona Michałowska; Adam Witkowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2013-09-16       Impact factor: 1.426

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.