Comparison of rebound and applanation tonometry in the management of patients treated for glaucoma or ocular hypertension.

Jeannette B Rehnman1, Lene Martin.   

Abstract

AIM: To compare intraocular pressure (IOP) values obtained with the iCare rebound tonometer (RBT) with values obtained with the Goldmann applanation tonometer (GAT) in patients with glaucoma and ocular hypertension, and to evaluate the impact of possible differences on management decisions.
METHODS: The IOP was measured with the iCare RBT and GAT in 45 consecutive patients in a nurse-led glaucoma unit. The outcome measures were the IOP difference between the methods and the 95% limits of agreement (LoA) for IOP measurements. In all subjects where the IOP readings differed by >3 mmHg between the RBT and GAT, two ophthalmologists were independently asked to recommend treatment and/or follow-up intervals, based on the patients' clinical data and the IOP values.
RESULTS: The RBT overestimated the IOP compared with GAT in 67% of the cases (30/45). In 78% (35/45) the IOP difference was within +/-3 mmHg. When an extreme value of 15 mmHg difference was excluded, the 95% LoA was +/-4.31 mmHg. In five of the 10 patients with an IOP difference of > +/- 3 mmHg the difference would have led to change in therapy; in three more patients the follow-up interval would have been changed. In the remaining two patients the difference in IOP would not have led to any change in management.
CONCLUSIONS: The mean difference between RBT and GAT was small (1.5 +/- 3 mmHg) compared with the published repeatability of GAT values. Yet, the difference was found to have clinical implications in 18% (8/45) of the patients.

Entities:  

Mesh:

Year:  2008        PMID: 18565094     DOI: 10.1111/j.1475-1313.2008.00571.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  6 in total

1.  Does rebound tonometry probe misalignment modify intraocular pressure measurements in human eyes?

Authors:  Ian G Beasley; Deborah S Laughton; Benjamin J Coldrick; Thomas E Drew; Marium Sallah; Leon N Davies
Journal:  J Ophthalmol       Date:  2013-08-29       Impact factor: 1.909

2.  The utility of rebound tonometer in late elderly subjects.

Authors:  Rei Sakata; Jiro Numaga
Journal:  Clin Interv Aging       Date:  2014-01-22       Impact factor: 4.458

3.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

4.  Comparison of Icare Pro Tonometry and Icare One Tonometry Measurements in Healthy Eyes

Authors:  Hüseyin Mayalı; Çağlar Sarıgül; Emin Kurt; Özcan Rasim Kayıkçıoğlu; Süleyman Sami İlker
Journal:  Turk J Ophthalmol       Date:  2019-06-27

5.  Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study.

Authors:  Annegret Hella Dahlmann-Noor; Renata Puertas; Shenille Tabasa-Lim; Ahmed El-Karmouty; Mustafa Kadhim; Nicholas Kloster Wride; Amanda Lewis; Dawn Grosvenor; Poornima Rai; Maria Papadopoulos; John Brookes; Catey Bunce; Peng Tee Khaw
Journal:  BMJ Open       Date:  2013-04-02       Impact factor: 2.692

6.  Intraocular pressure in a cohort of healthy eastern European schoolchildren: variations in method and corneal thickness.

Authors:  Patrycja Krzyżanowska-Berkowska; Magdalena Asejczyk-Widlicka; Barbara Pierscionek
Journal:  BMC Ophthalmol       Date:  2012-12-02       Impact factor: 2.209

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.