| Literature DB >> 18562156 |
Aamer Malik1, Lawrence D Dorr, William T Long.
Abstract
This case report is of a patient with disassociation of the acetabular cup liner caused by impingement. The cup inclination (39 degrees) and anteversion (24 degrees) were good as measured by computer navigation. Impingement occurred because the head-neck ratio was 2.0, and the hip length and offset were short by one head length. Successful revision without intraoperative impingement was accomplished with one size head larger (32 mm; head-neck ratio, 2.3) and one size longer to correct hip length and offset.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18562156 DOI: 10.1016/j.arth.2008.05.023
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757