Literature DB >> 18562156

Impingement as a mechanism of dissociation of a metasul metal-on-metal liner.

Aamer Malik1, Lawrence D Dorr, William T Long.   

Abstract

This case report is of a patient with disassociation of the acetabular cup liner caused by impingement. The cup inclination (39 degrees) and anteversion (24 degrees) were good as measured by computer navigation. Impingement occurred because the head-neck ratio was 2.0, and the hip length and offset were short by one head length. Successful revision without intraoperative impingement was accomplished with one size head larger (32 mm; head-neck ratio, 2.3) and one size longer to correct hip length and offset.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18562156     DOI: 10.1016/j.arth.2008.05.023

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger.

Authors:  Moritz M Innmann; Tobias Gotterbarm; Jan Philippe Kretzer; Christian Merle; Volker Ewerbeck; Stefan Weiss; Peter R Aldinger; Marcus R Streit
Journal:  Int Orthop       Date:  2013-12-20       Impact factor: 3.075

2.  Polyethylene Liner Dissociation Is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases.

Authors:  Andrew Yun; Emmanuel N Koli; John Moreland; Richard Iorio; John F Tilzey; J Wesley Mesko; Gwo-Chin Lee; Mark Froimson
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

3.  Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Authors:  Ian C Clarke; Jean-Yves Lazennec; Adrien Brusson; Christina Savisaar; John G Bowsher; Michelle Burgett; Thomas K Donaldson
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  Severe impingement of lumbar disc replacements increases the functional biological activity of polyethylene wear debris.

Authors:  Ryan M Baxter; Daniel W Macdonald; Steven M Kurtz; Marla J Steinbeck
Journal:  J Bone Joint Surg Am       Date:  2013-06-05       Impact factor: 5.284

5.  Long-Term Results of a Second-Generation, Small-Diameter, Metal-On-Metal Bearing in Primary Total Hip Arthroplasty at 14-year Follow-Up.

Authors:  Tobias Reiner; Matthias C Klotz; Kirsten Seelmann; Fabian Hertzsch; Moritz M Innmann; Marcus R Streit; Timo A Nees; Babak Moradi; Christian Merle; Jan Philippe Kretzer; Tobias Gotterbarm
Journal:  Materials (Basel)       Date:  2020-01-24       Impact factor: 3.623

6.  Impingement Resulting in Femoral Notching and Elevated Metal-Ion Levels After Dual-Mobility Total Hip Arthroplasty.

Authors:  Chelsea Matzko; Brandon Naylor; Ryan Cummings; Yevgeniy Korshunov; H John Cooper; Matthew S Hepinstall
Journal:  Arthroplast Today       Date:  2020-12-04

7.  10-year results of the uncemented Allofit press-fit cup in young patients.

Authors:  Marcus R Streit; Stefan Weiss; Franziska Andreas; Thomas Bruckner; Tilman Walker; J Philippe Kretzer; Volker Ewerbeck; Christian Merle
Journal:  Acta Orthop       Date:  2014-05-30       Impact factor: 3.717

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.