Literature DB >> 18554211

Diagnostic accuracy of a new software for complex fractionated electrograms identification in patients with persistent and permanent atrial fibrillation.

Leonardo Calò1, Ermenegildo De Ruvo, Luigi Sciarra, Roberto Gricia, Giovanna Navone, Lucia De Luca, Francesca Nuccio, Antonella Sette, Cristian Pristipino, Alessandro Dulio, Fiorenzo Gaita, Ernesto Lioy.   

Abstract

INTRODUCTION: The elimination of complex fractionated atrial electrograms (CFAEs) has been proposed as a potential target for guiding successful AF substrate ablation. The possibility to efficiently map the atria and rapidly identify CFAEs sites is necessary, before the CFAEs ablation becomes a routine approach. The aims of this study, conducted in patients with persistent and permanent atrial fibrillation (AF), were to analyze by CARTO mapping in the right (RA) and in the left atrium (LA) during AF: (1) the diagnostic accuracy of a new software for CFAEs analysis, (2) the spatial distribution of CFAEs, (3) the regional beat to beat AF intervals (FF). METHODS AND
RESULTS: Twenty-five consecutive patients (four women, 58.8 +/- 11.4 years) undergoing radiofrequency catheter ablation for persistent and permanent AF were enrolled in the study. The CFAE software showed a high sensitivity (90%) and specificity (91%) in the identification of CFAEs, using a specific setting of parameters. The LA had a significantly higher prevalence of CFAEs as compared with the RA (30.5% vs 20.3%, P = 0.016). The CFAEs were mostly present in the septum and in the area of coronary sinus ostium (CS os). The FF intervals were significantly shorter in the LA than in the RA (P < 0.01).
CONCLUSION: CARTO system has a high diagnostic accuracy in the identification of CFAEs. Atrial electrical activity (CFAEs, mean FF intervals) during AF showed a significant spatial inhomogeneity.

Entities:  

Mesh:

Year:  2008        PMID: 18554211     DOI: 10.1111/j.1540-8167.2008.01219.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  10 in total

Review 1.  State-of-the-art and emerging technologies for atrial fibrillation ablation.

Authors:  Jane Dewire; Hugh Calkins
Journal:  Nat Rev Cardiol       Date:  2010-03       Impact factor: 32.419

2.  Long-term outcomes of adjunctive complex fractionated electrogram ablation to pulmonary vein isolation as treatment for non-paroxysmal atrial fibrillation.

Authors:  Alessandro De Bortoli; Ole-Jørgen Ohm; Per Ivar Hoff; Li-Zhi Sun; Peter Schuster; Eivind Solheim; Jian Chen
Journal:  J Interv Card Electrophysiol       Date:  2013-07-07       Impact factor: 1.900

3.  Automated Detection of Complex Fractionated Atrial Electrograms In Substrate-Based Atrial Fibrillation Ablation: Better Discrimination with a New Setting of CARTO® Algorithm.

Authors:  Julien Seitz; Jérôme Horvilleur; Jérôme Lacotte; Yamina Mouhoub; Fiorella Salerno; Anouska Moynagh; Darach O H-Ici; Mehran Monchi; Laurence Curel; Andre Pisapia
Journal:  J Atr Fibrillation       Date:  2013-08-31

Review 4.  Catheter Ablation Targeting Complex Fractionated Atrial Electrogram in Atrial Fibrillation.

Authors:  Dennis H Lau; Stef Zeemering; Bart Maesen; Pawel Kuklik; Sander Verheule; Ulrich Schotten
Journal:  J Atr Fibrillation       Date:  2013-10-31

5.  Autonomic modulation of complex fractionated atrial electrograms in patients with paroxysmal atrial fibrillation.

Authors:  Demosthenes Katritsis; Demetrios Sougiannis; Kimon Batsikas; Eleftherios Giazitzoglou; John Mersinias; George Katritsis; Sunny S Po
Journal:  J Interv Card Electrophysiol       Date:  2011-03-19       Impact factor: 1.900

Review 6.  Elimination Of Triggers Without An Additional Substrate Modification Is Not Sufficient In Patients With Persistent Atrial Fibrillation.

Authors:  Junbeom Park; Hui-Nam Pak
Journal:  J Atr Fibrillation       Date:  2015-02-28

7.  Automatic Extraction of Recurrent Patterns of High Dominant Frequency Mapping During Human Persistent Atrial Fibrillation.

Authors:  Xin Li; Gavin S Chu; Tiago P Almeida; Frederique J Vanheusden; João Salinet; Nawshin Dastagir; Amar R Mistry; Zakariyya Vali; Bharat Sidhu; Peter J Stafford; Fernando S Schlindwein; G André Ng
Journal:  Front Physiol       Date:  2021-03-12       Impact factor: 4.566

8.  Surface ECG-based complexity parameters for predicting outcomes of catheter ablation for nonparoxysmal atrial fibrillation: efficacy of fibrillatory wave amplitude.

Authors:  Jong-Il Park; Seung-Woo Park; Min-Ji Kwon; Jeon Lee; Hong-Ju Kim; Chan-Hee Lee; Dong-Gu Shin
Journal:  Medicine (Baltimore)       Date:  2022-08-05       Impact factor: 1.817

Review 9.  Current hot potatoes in atrial fibrillation ablation.

Authors:  Laurent Roten; Nicolas Derval; Patrizio Pascale; Daniel Scherr; Yuki Komatsu; Ashok Shah; Khaled Ramoul; Arnaud Denis; Frédéric Sacher; Mélèze Hocini; Michel Haïssaguerre; Pierre Jaïs
Journal:  Curr Cardiol Rev       Date:  2012-11

10.  Ablation of complex fractionated atrial electrograms in catheter ablation for AF; where have we been and where are we going?

Authors:  Jane Caldwell; Damian Redfearn
Journal:  Curr Cardiol Rev       Date:  2012-11
  10 in total