Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer?

Claude Nahmias1, Catherine Lemmens, David Faul, Eric Carlson, Misty Long, Todd Blodgett, Johan Nuyts, David Townsend.   

Abstract

UNLABELLED: In patients with oral head and neck cancer, the presence of metallic dental implants produces streak artifacts in the CT images. These artifacts negate the utility of CT for the spatial localization of PET findings and may propagate through the CT-based attenuation correction into the PET images. In this study, we evaluated the efficacy of an algorithm that reduces metallic artifacts in CT images and the impact of this approach on the quantification of PET images.
METHODS: Fifty-one patients with and 9 without dental implants underwent a PET/CT study. CT images through the patient's dental implants were reconstructed using both standard CT reconstruction and an algorithm that reduces metallic artifacts. Attenuation correction factors were calculated from both sets of CT images and applied to the PET data. The CT images were evaluated for any reduction of the artifacts. The PET images were assessed for any quantitative change introduced by metallic artifact reduction.
RESULTS: For each reconstruction, 2 regions of interest were defined in areas where the standard CT reconstruction overestimated the Hounsfield units (HU), 2 were defined in underestimated areas, and 1 was defined in a region unaffected by the artifacts. The 5 regions of interest were transferred to the other 3 reconstructions. Mean HU or mean Bq/cm(3) were obtained for all regions. In the CT reconstructions, metallic artifact reduction decreased the overestimated HUs by approximately 60% and increased the underestimated HUs by approximately 90%. There was no change in quantification in the PET images between the 2 algorithms (Spearman coefficient of rank correlation, 0.99). Although the distribution of attenuation (HU) changed considerably in the CT images, the distribution of activity did not change in the PET images.
CONCLUSION: Our study demonstrated that the algorithm can enhance the structural and spatial content of CT images in the presence of metallic artifacts. The CT artifacts do not propagate through the CT-based attenuation correction into the PET images, confirming the robustness of CT-based attenuation correction in the presence of metallic artifacts. The study also demonstrated that considerable changes in CT images do not change the PET images.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18552154     DOI: 10.2967/jnumed.107.049858

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  16 in total

1.  Comparative assessment of energy-mapping approaches in CT-based attenuation correction for PET.

Authors:  Mohammad R Ay; Maryam Shirmohammad; Saeed Sarkar; Arman Rahmim; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2011-02       Impact factor: 3.488

2.  Qualitative and quantitative assessment of metal artifacts arising from implantable cardiac pacing devices in oncological PET/CT studies: a phantom study.

Authors:  Mohammad R Ay; Abolfazl Mehranian; Mehrsima Abdoli; Pardis Ghafarian; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2011-12       Impact factor: 3.488

3.  Successive iterative restoration applied to streak artifact reduction in X-ray CT image of dento-alveolar region.

Authors:  Jian Dong; Atsushi Kondo; Kosuke Abe; Yoshihiko Hayakawa
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-05       Impact factor: 2.924

4.  Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny.

Authors:  Doumit Daou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-09-12       Impact factor: 9.236

5.  Statistical iterative reconstruction for streak artefact reduction when using multidetector CT to image the dento-alveolar structures.

Authors:  J Dong; Y Hayakawa; C Kober
Journal:  Dentomaxillofac Radiol       Date:  2014-04-22       Impact factor: 2.419

6.  Reduction of artefacts caused by hip implants in CT-based attenuation-corrected PET images using 2-D interpolation of a virtual sinogram on an irregular grid.

Authors:  Mehrsima Abdoli; Johan R de Jong; Jan Pruim; Rudi A J O Dierckx; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-08-18       Impact factor: 9.236

7.  Clinical evaluation of the normalized metal artefact reduction algorithm caused by dental fillings in CT.

Authors:  X-Y Gong; E Meyer; X-J Yu; J-H Sun; L-P Sheng; K-H Huang; R-Z Wu
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

Review 8.  Accuracy of imaging methods for detection of bone tissue invasion in patients with oral squamous cell carcinoma.

Authors:  S Uribe; L A Rojas; C F Rosas
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

Review 9.  Application of positron emission tomography/computed tomography in radiation treatment planning for head and neck cancers.

Authors:  Musaddiq J Awan; Farzan Siddiqui; David Schwartz; Jiankui Yuan; Mitchell Machtay; Min Yao
Journal:  World J Radiol       Date:  2015-11-28

Review 10.  Post-therapeutic surveillance schedule for oral cancer: is there agreement?

Authors:  Guicai Liu; Eric J Dierks; R Bryan Bell; Tuan G Bui; Bryce E Potter
Journal:  Oral Maxillofac Surg       Date:  2012-09-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.