Yongwen Qin1, Jinming Chen, Xianxian Zhao, Dening Liao, Ruibin Mu, Shengqiang Wang, Hong Wu, Haoxue Guo.
Abstract
This multicenter study assessed the efficacy and safety of transcatheter closure of perimembranous ventricular septal defect (VSD) using a modified double-disk occluder. In 5 different centers in China, 412 patients with VSD, including 202 men and 210 women, underwent attempted transcatheter closure. The age range was 3 to 65 years (mean 16.4+/-9.1). The diameter of defect was 3 to 15 mm by transthoracic echocardiography and 3 to 18 mm by left ventriculography. The ratio of pulmonary to systemic flow varied from 1.6 to 2.3 (1.9+/-0.4). The device diameter was 4 to 20 mm (7.09+/-3.60). The ventricular septal rim below the aortic valve was 0 to 5 mm. The immediate success rate was 96.6%; 6 cases had third degree atrioventricular block and recovered within 3 weeks. None needed a permanent pacemaker. Dislodgement of the device occurred in 3 patients but the device was recaptured and redeployed in 2 cases. During the follow-up period of 2 years, there was no evidence of residual shunt and device-related complications. In conclusion, transcatheter closure of VSD is safe and effective in most selected patients; the mid-term prognosis of patients with transcatheter closure is good.Entities:
Mesh:
Year: 2008 PMID: 18549859 DOI: 10.1016/j.amjcard.2008.02.069
Source DB: PubMed Journal: Am J Cardiol ISSN: 0002-9149 Impact factor: 2.778