Catastrophic failure of ceramic-polyethylene bearing total hip arthroplasty.

Justin Needham1, Travis Burns, Tad Gerlinger.   

Abstract

Complications of ceramic-polyethylene bearing total hip arthroplasty (THA) include osteolysis, loosening, dislocation, and component failure. Catastrophic acetabular component failure involves severe damage to both the polyethylene liner and metal shell. This case study presents the first reported complete wear-through of the acetabular portion of a ceramic-polyethylene arthroplasty presenting as a dislocation and a review of the literature. In this study, a patient's alumina ceramic femoral head penetrated the polyethylene liner and titanium shell and presented as a dislocated THA. The contributing factors for this catastrophic failure include young patient age, high activity level, thin polyethylene liner, backside wear, component positioning, polyethylene sterilization with gamma irradiation in air, and lack of appropriate follow-up. Revision THA was performed without complications.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18514888     DOI: 10.1016/j.arth.2007.04.026

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Use of morphometry to quantify osteolysis after total hip arthroplasty.

Authors:  Lindsay K Smith; Fiona Cramp; Shea Palmer; Nikki Coghill; Robert F Spencer
Journal:  Clin Orthop Relat Res       Date:  2010-05-12       Impact factor: 4.176

2.  Catastrophic failure of ceramic-on-ceramic total hip arthroplasty presenting as squeaking hip.

Authors:  David Malem; Mathias Thomas Nagy; Sabyasachi Ghosh; Bhavik Shah
Journal:  BMJ Case Rep       Date:  2013-02-20

3.  Catastrophic wear in uncemented acetabular cups after Total Hip Arthroplasty. A case series.

Authors:  Salomon Jasqui Remba; Ricardo Esquivel Gomez; Luis Gutiérrez Ambriz; Daniel Portman Santos; Isaac Baley Amiga
Journal:  Int J Surg Case Rep       Date:  2021-03-30

4.  Early catastrophic acetabular failure in Furlong total hip replacements.

Authors:  David Knox; Steven W Hamilton; Douglas Wardlaw; Peter H Gibson
Journal:  J Orthop Traumatol       Date:  2009-01-31

5.  Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty.

Authors:  Givenchy Manzano; Rayna A C Levin; Michael B Mayor; Ran Schwarzkopf
Journal:  J Orthop Case Rep       Date:  2014 Apr-Jun

6.  Prosthetic femoral head erosion through an acetabular component treated with revision and implant preservation.

Authors:  Mark Daniel Kohn; Navin Fernando
Journal:  Arthroplast Today       Date:  2020-01-14

7.  Do Not Postpone Revision of Worn Conventional Liners in Ceramic-on-Polyethylene Total Hip Arthroplasty: A New Dramatic Failure.

Authors:  Thorsten Gehrke; Mustafa Citak; Hussein Abdelaziz
Journal:  Arthroplast Today       Date:  2021-07-19
  7 in total