Image-based monitoring of one-lung ventilation.

S Jean1, I Cinel, I Gratz, C Tay, V Lotano, E Deal, J E Parrillo, R P Dellinger.   

Abstract

BACKGROUND AND OBJECTIVES: With the increasing demand for one-lung ventilation in both thoracic surgery and other procedures, identifying the correct placement becomes increasingly important. Currently, endobronchial intubation is suspected based on a combination of auscultation and physiological findings. We investigated the ability of the visual display of airflow-induced vibrations to detect single-lung ventilation with a double-lumen endotracheal tube.
METHODS: Double-lumen tubes were placed prior to surgery. Tracheal and endobronchial lumens were alternately clamped to produce unilateral lung ventilation of right and left lung. Vibration response imaging, which detects vibrations transmitted to the surface of the thorax, was performed during both right- and left-lung ventilation. Geographical area of vibration response image as well as amount and distribution of lung sounds were assessed.
RESULTS: During single-lung ventilation, the image and video obtained from the vibration response imaging identifies the ventilated lung with a larger and darker image on the ventilated side. During single-lung ventilation, 87.2 +/- 5.7% of the measured vibrations was detected over the ventilated lung and 12.8 +/- 5.7% over the non-ventilated lung (P < 0.0001). It was also noted that during single-lung ventilation, the vibration distribution in the non-ventilated lung had a majority of vibration detected by the medial sensors closest to the midline (P < 0.05) as opposed to the midclavicular sensors when the lung is ventilated.
CONCLUSIONS: During single-lung ventilation, vibration response imaging clearly showed increased vibration in the lung that is being ventilated. Distribution of residual vibration differed in the non-ventilated lung in a manner that suggests transmission of vibrations across the mediastinum from the ventilated lung. The lung image and video obtained from vibration response imaging may provide useful and immediate information to help one-lung ventilation assessment.

Mesh:

Year:  2008        PMID: 18492316     DOI: 10.1017/S0265021508004481

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  3 in total

1.  Detection of endobronchial intubation by monitoring the CO2 level above the endotracheal cuff.

Authors:  Shai Efrati; Israel Deutsch; Nathan Weksler; Gabriel M Gurman
Journal:  J Clin Monit Comput       Date:  2014-05-29       Impact factor: 2.502

2.  Changes in regional distribution of lung sounds as a function of positive end-expiratory pressure.

Authors:  Shaul Lev; Yael A Glickman; Ilya Kagan; David Dahan; Jonathan Cohen; Milana Grinev; Maury Shapiro; Pierre Singer
Journal:  Crit Care       Date:  2009-05-10       Impact factor: 9.097

3.  Vibration response imaging: protocol for a systematic review.

Authors:  Marc P Berry; Luigi Camporota; George Ntoumenopoulos
Journal:  Syst Rev       Date:  2013-09-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.