Literature DB >> 18492306

Long-term complications of bone-anchored hearing aids: a 14-year experience.

K Badran1, A K Arya, D Bunstone, N Mackinnon.   

Abstract

OBJECTIVES: To report cases of long-term surgical complications, implant failure and revision surgery, within a large bone-anchored hearing aid programme. STUDY
DESIGN: Retrospective, case-cohort study.
SETTING: Tertiary referral centre. PATIENTS: One hundred and sixty-five adults and children who had undergone a total of 177 bone-anchored hearing aid implantations. INTERVENTION: Diagnosis and explanation of adverse events and device failure. MAIN OUTCOME MEASURES: Operative complications and survival analysis, surgical challenges related to revisions, and causes of failure.
RESULTS: Twenty-one per cent of patients (3.4 per cent of those observed) suffered from skin reactions; this rate did not increase over time. Seventeen per cent had loss of osseointegration at a median interval of 6.3 months. Loss of osseointegration was observed more frequently in patients with a 3 mm compared with a 4 mm fixture (p < 0.001). Intra-operatively, the only complication was bleeding, occurring in 3 per cent of patients. Post-operative complications included: primary bleeding (2 per cent); severe skin reactions requiring intravenous antibiotics, cautery or grafting (8 per cent); thickening or overgrowth of skin requiring excision (8 per cent); failure of osseointegration requiring a new fixture (18 per cent); and graft necrosis requiring revision (1 per cent). In two patients, it was necessary to explore the area to remove overgrowth of bone. In 16 patients (10 per cent), the bone-anchored hearing aid had to be abandoned due to failure of osseointegration (n = 4), dissatisfaction with the aid (n = 6), intolerable pain (n = 4), hair growth around the abutment (n = 1) or recurrent infections (n = 1). In 12 of these patients, the bone-anchored hearing aid was removed surgically. Overall, 57 patients (34 per cent) underwent revision surgery.
CONCLUSION: Awareness of complications is becoming increasingly important in bone-anchored hearing aid programme. A substantial workload of device maintenance should be anticipated, and patients should be appropriately counselled beforehand. Ninety per cent of our patients chose to persevere with this form of hearing rehabilitation.

Entities:  

Mesh:

Year:  2008        PMID: 18492306     DOI: 10.1017/S0022215108002521

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  17 in total

Review 1.  Human cortical prostheses: lost in translation?

Authors:  Stephen I Ryu; Krishna V Shenoy
Journal:  Neurosurg Focus       Date:  2009-07       Impact factor: 4.047

Review 2.  A systematic review on skin complications of bone-anchored hearing aids in relation to surgical techniques.

Authors:  Shwan Mohamad; Imran Khan; S Y Hey; S S Musheer Hussain
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-12-14       Impact factor: 2.503

3.  Switching from a percutaneous to a transcutaneous bone anchored hearing system: the utility of the fascia temporalis superficialis pedicled flap in case of skin intolerance.

Authors:  A Devèze; S Rossetto; R Meller; M Sanjuan Puchol
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-01-30       Impact factor: 2.503

Review 4.  Postoperative imaging of the internal auditory canal : Visualization of active auditory implants.

Authors:  I Todt; G Rademacher; P Mittmann; S Mutze; A Ernst
Journal:  HNO       Date:  2017-08       Impact factor: 1.284

Review 5.  [Postoperative imaging of the internal auditory canal : Visualization of active auditory implants. German version].

Authors:  I Todt; G Rademacher; P Mittmann; S Mutze; A Ernst
Journal:  HNO       Date:  2017-09       Impact factor: 1.284

6.  Congenital aural atresia treated with floating mass transducer on the round window: 5 years of imaging experience.

Authors:  M Barillari; R Cerini; M Carner; C Cacciatori; F Spagnolli; N Cardobi; M Mandalà; L Colletti; V Colletti; R Pozzi Mucelli
Journal:  Radiol Med       Date:  2011-11-17       Impact factor: 3.469

7.  Long-term audiological benefit with an active transcutaneous bone-conduction device: a retrospective cohort analysis.

Authors:  Julia Hundertpfund; Jens Eduard Meyer; Attila Ovari
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-08-23       Impact factor: 2.503

8.  Modified Baha Punch Technique: Least Invasive, Shortest Time and No Suturing.

Authors:  Hassan Alshehri; Abdulrahman Alsanosi; Osama Majdalawieh
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2015-12-12

9.  Clobetasol's Influence on the Management and Cost of Skin Overgrowth Associated with the Bone-Anchored Hearing Aid.

Authors:  Douglas M Hildrew; Jesse A Guittard; John M Carter; Timothy B Molony
Journal:  Ochsner J       Date:  2015

10.  Implantable Devices for Single-Sided Deafness and Conductive or Mixed Hearing Loss: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2020-03-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.