A new approach for assessment of wear in metal-backed acetabular cups using computed tomography: a phantom study with retrievals.

Anneli Jedenmalm1, Marilyn E Noz, Henrik Olivecrona, Lotta Olivecrona, Andre Stark.   

Abstract

BACKGROUND AND
PURPOSE: Polyethylene wear is an important cause of aseptic loosening in hip arthroplasty. Detection of significant wear usually happens late on, since available diagnostic techniques are either not sensitive enough or too complicated and expensive for routine use. This study evaluates a new approach for measurement of linear wear of metal-backed acetabular cups using CT as the intended clinically feasible method.
MATERIAL AND METHODS: 8 retrieved uncemented metal-backed acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral head into the cup was measured in the CT volumes using dedicated software. Landmark points were placed on the CT images of cup and head, and also on a reference plane in order to calculate the wear vector magnitude and angle to one of the axes. A coordinate-measuring machine was used to test the accuracy of the proposed CT method. For this purpose, the head diameters were also measured by both methods.
RESULTS: Accuracy of the CT method for linear wear measurements was 0.6 mm and wear vector angle was 27 degrees . No systematic difference was found between CT scans.
INTERPRETATION: This study on explanted acetabular cups shows that CT is capable of reliable measurement of linear wear in acetabular cups at a clinically relevant level of accuracy. It was also possible to use the method for assessment of direction of wear.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18484247     DOI: 10.1080/17453670710015003

Source DB:  PubMed          Journal:  Acta Orthop        ISSN: 1745-3674            Impact factor:   3.717


  6 in total

1.  Distal femoral stem-bone anchorage of a cementless revision total hip arthroplasty: evaluation of 14 patients by CT.

Authors:  Rüdiger J Weiss; Fredrik Strömwall; Mats O Beckman; Karl A Hansson; André Stark
Journal:  Acta Orthop       Date:  2009-06       Impact factor: 3.717

2.  Validation of a 3D CT method for measurement of linear wear of acetabular cups.

Authors:  Anneli Jedenmalm; Fritjof Nilsson; Marilyn E Noz; Douglas D Green; Ulf W Gedde; Ian C Clarke; Andreas Stark; Gerald Q Maguire; Michael P Zeleznik; Henrik Olivecrona
Journal:  Acta Orthop       Date:  2011-02-01       Impact factor: 3.717

3.  Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique.

Authors:  Eric Vandenbussche; Mohammed Saffarini; Ulrich Hansen; Fabienne Taillieu; Céline Mutschler; Bernard Augereau; Thomas M Gregory
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

4.  A CT method for following patients with both prosthetic replacement and implanted tantalum beads: preliminary analysis with a pelvic model and in seven patients.

Authors:  Henrik Olivecrona; Gerald Q Maguire; Marilyn E Noz; Michael P Zeleznik; Uldis Kesteris; Lars Weidenhielm
Journal:  J Orthop Surg Res       Date:  2016-02-24       Impact factor: 2.359

5.  Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography.

Authors:  Lars Weidenhielm; Henrik Olivecrona; Gerald Q Maguire; Marilyn E Noz
Journal:  Skeletal Radiol       Date:  2018-01-23       Impact factor: 2.199

6.  A simple method of measuring the wear of explanted acetabular component inserts.

Authors:  L Krakow; A Klockow; E Roehner; S Brodt; H Eijer; J Bossert; G Matziolis
Journal:  Bone Joint Res       Date:  2017-09       Impact factor: 5.853

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.