Abstract

The Airtraq laryngoscope has an oropharyngeal airway-shaped blade that provides a non-line-of-sight view of the glottis. The configuration of the blade should mean that less movement of the cervical spine is required during laryngeal visualisation. We compared the degree of cervical spine movement in laryngoscopy performed using the Airtraq and conventional Macintosh laryngoscope. In 20 patients requiring general anaesthesia and tracheal intubation, we measured cervical spine movement using radiography in the same patient during consecutive procedures using the two laryngoscopes. Although significant movement of the cervical spine from baseline was noted during all procedures (p < 0.05), cervical spinal extension with the Airtraq was 29% less than that measured during Macintosh laryngoscopy between the occiput and C4, and 44% less at the C3/C4 motion segment (p < 0.05). Anterior deviations of the vertebral bodies from baseline were 32%, 35%, 38% and 40% less at the atlas, C2, C3, and C4 vertebrae, respectively, during Airtraq laryngoscopy than those measured during Macintosh laryngoscopy (p < 0.01). Our study demonstrated that laryngoscopy using the Airtraq laryngoscope involves less movement of the cervical spine compared to conventional procedures using a Macintosh laryngoscope.

Entities:  

Mesh:

Year:  2008        PMID: 18477276     DOI: 10.1111/j.1365-2044.2008.05480.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  20 in total

Review 1.  [Indirect laryngoscopy : Alternatives to securing the airway].

Authors:  R R Noppens; C Werner; T Piepho
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

2.  A comparison of video laryngoscopy to direct laryngoscopy for the emergency intubation of trauma patients.

Authors:  Maria Michailidou; Terence O'Keeffe; Jarrod M Mosier; Randall S Friese; Bellal Joseph; Peter Rhee; John C Sakles
Journal:  World J Surg       Date:  2015-03       Impact factor: 3.352

3.  Comparison of Airtraq optical laryngoscope and Storz video laryngoscope in a cadaver model.

Authors:  Michael C Wadman; Travis W Dierks; Chad E Branecki; Claudia L Barthold; Lance H Hoffman; Lina Lander; Carol S Lomneth; Richard A Walker
Journal:  World J Emerg Med       Date:  2011

4.  Distortion of anterior airway anatomy during laryngoscopy with the GlideScope videolaryngoscope.

Authors:  Yoshihiro Hirabayashi; Akifumi Fujita; Norimasa Seo; Hideharu Sugimoto
Journal:  J Anesth       Date:  2010-04-06       Impact factor: 2.078

5.  Advances in laryngoscopy: rigid indirect laryngoscopy.

Authors:  Deanne R Cheyne; Patrick Doyle
Journal:  F1000 Med Rep       Date:  2010-08-19

6.  A comparison of 4 airway devices on cervical spine alignment in cadaver models of global ligamentous instability at c1-2.

Authors:  Adam L Wendling; Patrick J Tighe; Bryan P Conrad; Tezcan Ozrazgat Baslanti; Marybeth Horodyski; Glenn R Rechtine
Journal:  Anesth Analg       Date:  2013-01-25       Impact factor: 5.108

7.  Face-to-face tracheal intubation in adult patients: a comparison of the Airtraq™, Glidescope™ and Fastrach™ devices.

Authors:  Zehra Ipek Arslan; Volkan Alparslan; Pınar Ozdal; Kamil Toker; Mine Solak
Journal:  J Anesth       Date:  2015-07-29       Impact factor: 2.078

8.  Intubation of a Paediatric Manikin in Tongue Oedema and Face-to-Face Simulations by Novice Personnel: a Comparison of Glidescope, Airtraq and Direct Laryngoscopy.

Authors:  Zehra İpek Arslan; Canan Turna; Nevin Esra Gümüş; Kamil Toker; Mine Solak
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-04-01

9.  Intubation biomechanics: laryngoscope force and cervical spine motion during intubation with Macintosh and Airtraq laryngoscopes.

Authors:  Bradley J Hindman; Brandon G Santoni; Christian M Puttlitz; Robert P From; Michael M Todd
Journal:  Anesthesiology       Date:  2014-08       Impact factor: 7.892

Review 10.  Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation.

Authors:  Sharon R Lewis; Andrew R Butler; Joshua Parker; Tim M Cook; Andrew F Smith
Journal:  Cochrane Database Syst Rev       Date:  2016-11-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.