Corneal power measurements with the Pentacam Scheimpflug camera after myopic excimer laser surgery.

Giacomo Savini1, Piero Barboni, Vincenzo Profazio, Maurizio Zanini, Kenneth J Hoffer.   

Abstract

PURPOSE: To evaluate corneal power measurements by a rotating Scheimpflug camera (Pentacam, Oculus) in eyes that have had myopic excimer laser surgery.
SETTING: Private practice, Bologna, Italy.
METHODS: This prospective comparative interventional case series comprised 16 eyes of 16 patients who had myopic excimer laser surgery and for whom all perioperative data were available. Four corneal power measurements obtained with the Pentacam (simulated keratometry, true net power, equivalent K reading, and BESSt formula) were analyzed and compared with values derived using the clinical history method and 2 other formulas for calculating corneal power after refractive surgery (modified keratometric refractive index according to Savini et al. and separate consideration of the anterior and posterior corneal curvatures according to Speicher).
RESULTS: Analysis of variance showed a statistically significant difference between all methods (P<.0001). Bonferroni multiple comparison tests showed that the only Pentacam measurements not statistically different from the corneal power values derived using the clinical history method were the equivalent K readings at 1.0 mm, 2.0 mm, and 3.0 mm and those derived with the BESSt formula; however, considerably large 95% limits of agreement (LoA) were calculated between each of these values and those obtained with the clinical history method.
CONCLUSIONS: The Pentacam device gave corneal power measurements that did not statistically significantly had differ from those predicted by the clinical history method in eyes that had previous myopic excimer laser surgery. Wide LoA are a potential source of error in intraocular lens power calculation in such patients.

Entities:  

Mesh:

Year:  2008        PMID: 18471637     DOI: 10.1016/j.jcrs.2008.01.012

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  13 in total

1.  Orbscan II and double-K method for IOL calculation after refractive surgery.

Authors:  Sérgio Kwitko; Diane R Marinho; Samuel Rymer; Norton Severo; Carlos G Arce
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-03-29       Impact factor: 3.117

2.  Practical method to calculate post-LASIK corneal power: the Actual K(a+p) method.

Authors:  Yi-Qian Hu; Xiang-Yu Ye; Xiu-Li Zhou; Yin Li; Cheng-Hui Xu; Lin-Lu Tian; Huang Zhu; Yao-Hua Sheng
Journal:  Int J Ophthalmol       Date:  2010-12-18       Impact factor: 1.779

3.  Assessment of total corneal power after myopic corneal refractive surgery in Chinese eyes.

Authors:  Yanjun Hua; Chao Pan; Qinmei Wang
Journal:  Int Ophthalmol       Date:  2019-03-01       Impact factor: 2.031

4.  Corneal power measurement with a new aberrometer/corneal topographer in eyes after small incision lenticule extraction for myopia.

Authors:  Chao Pan; Weina Tan; Yanjun Hua; Xiaohua Lei
Journal:  Int Ophthalmol       Date:  2019-05-27       Impact factor: 2.031

5.  Excimer Laser Corneal Refractive Surgery in the Clinic: A Systematic Review and Meta-analysis.

Authors:  Huang Zhang; Mingming Li; Zhimin Cen
Journal:  Comput Math Methods Med       Date:  2022-06-15       Impact factor: 2.809

6.  Day to Day Clinically Relevant Corneal Elevation, Thickness, and Curvature Parameters Using the Orbscan II Scanning Slit Topographer and the Pentacam Scheimpflug Imaging Device.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

7.  Agreement between clinical history method, Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery.

Authors:  Kaevalin Lekhanont; Manachai Nonpassopon; Khemruetai Wannarosapark; Varintorn Chuckpaiwong
Journal:  PLoS One       Date:  2015-04-08       Impact factor: 3.240

8.  Corneal power evaluation after myopic corneal refractive surgery using artificial neural networks.

Authors:  Robert Koprowski; Michele Lanza; Carlo Irregolare
Journal:  Biomed Eng Online       Date:  2016-11-15       Impact factor: 2.819

Review 9.  Intraocular lens power calculation following laser refractive surgery.

Authors:  Christopher Hodge; Colm McAlinden; Michael Lawless; Colin Chan; Gerard Sutton; Aifric Martin
Journal:  Eye Vis (Lond)       Date:  2015-04-02

10.  Evaluation of Equivalent Keratometry Readings Obtained by Pentacam HR (High Resolution).

Authors:  Yanjun Hua; Xiaolan Zhang; Tor Paaske Utheim; Jinhai Huang; Chao Pan; Weina Tan; Qinmei Wang
Journal:  PLoS One       Date:  2016-03-07       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.