Tidal volume threshold for colorimetric carbon dioxide detectors available for use in neonates.

Donna M Garey1, Raymond Ward, Wade Rich, Gregory Heldt, Tina Leone, Neil N Finer.   

Abstract

OBJECTIVE: Colorimetric carbon dioxide detectors are used for confirmation of endotracheal intubation. The colorimetric carbon dioxide detectors that are used for neonates are labeled for use with infants and small children >1 and <15 kg. The objective of this study was to determine the minimal tidal volume that causes a breath-to-breath color change on 2 colorimetric carbon dioxide detectors.
METHODS: Using an artificial-lung model, we determined the tidal volume threshold of 2 colorimetric carbon dioxide detectors (Pedi-Cap [Nellcor, Pleasanton, CA] or Mini StatCO(2) [Mercury Medical, Clearwater, FL]) during ventilation with a T-piece resuscitator or neonatal ventilator. Digital video recordings of the colorimetric carbon dioxide detectors were made during 20 seconds of ventilation at each tidal volume. Seven clinicians who were blinded to the tidal volume reviewed the videos in random order and graded the color change to determine adequacy for clinical application.
RESULTS: The Mini StatCO(2) tidal volume threshold was 0.83 mL, and the Pedi-Cap tidal volume threshold was 1.08 mL.
CONCLUSIONS: The lung model revealed that the tidal volume threshold for the tested colorimetric carbon dioxide detectors is less than the expected tidal volume of a 400-g infant and suggests that these devices are appropriate for use with any neonate to confirm intubation.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18458036     DOI: 10.1542/peds.2007-2708

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  7 in total

1.  Neonatal Resuscitation with an Intact Cord: A Randomized Clinical Trial.

Authors:  Anup Katheria; Debra Poeltler; Jayson Durham; Jane Steen; Wade Rich; Kathy Arnell; Mauricio Maldonado; Larry Cousins; Neil Finer
Journal:  J Pediatr       Date:  2016-08-26       Impact factor: 4.406

2.  Efficacy and user preference of two CO2 detectors in an infant mannequin randomized crossover trial.

Authors:  G A Hawkes; B J O'Connell; V Livingstone; C P Hawkes; C A Ryan; E M Dempsey
Journal:  Eur J Pediatr       Date:  2013-06-12       Impact factor: 3.183

3.  Delivery room management of very low birth weight infants in Germany, Austria and Switzerland--a comparison of protocols.

Authors:  C C Roehr; S Gröbe; M Rüdiger; H Hummler; M Nelle; H Proquitté; H Hammer; G Schmalisch
Journal:  Eur J Med Res       Date:  2010-11-25       Impact factor: 2.175

Review 4.  WITHDRAWN: Techniques to ascertain correct endotracheal tube placement in neonates.

Authors:  Georg M Schmölzer; Charles Christoph C Roehr
Journal:  Cochrane Database Syst Rev       Date:  2018-07-05

Review 5.  Applications of End-Tidal Carbon Dioxide (ETCO2) Monitoring in Emergency Department; a Narrative Review.

Authors:  Hamed Aminiahidashti; Sajad Shafiee; Alieh Zamani Kiasari; Mohammad Sazgar
Journal:  Emerg (Tehran)       Date:  2018-01-15

Review 6.  [Newborn resuscitation and support of transition of infants at birth].

Authors:  John Madar; Charles C Roehr; Sean Ainsworth; Hege Ersda; Colin Morley; Mario Rüdiger; Christiane Skåre; Tomasz Szczapa; Arjan Te Pas; Daniele Trevisanuto; Berndt Urlesberger; Dominic Wilkinson; Jonathan P Wyllie
Journal:  Notf Rett Med       Date:  2021-06-02       Impact factor: 0.892

Review 7.  Non-invasive carbon dioxide monitoring in neonates: methods, benefits, and pitfalls.

Authors:  Deepika Sankaran; Lida Zeinali; Sameeia Iqbal; Praveen Chandrasekharan; Satyan Lakshminrusimha
Journal:  J Perinatol       Date:  2021-06-19       Impact factor: 2.521

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.