Literature DB >> 18425828

Single-breathhold four-dimensional assessment of left ventricular volumes and function using k-t BLAST after application of extracellular contrast agent at 3 Tesla.

Ashraf Hamdan1, Sebastian Kelle, Bernhard Schnackenburg, Ernst Wellnhofer, Eckart Fleck, Eike Nagel.   

Abstract

PURPOSE: To prospectively determine the feasibility and accuracy of a four-dimensional (4D) k-space over time broad-use linear acquisition speed-up technique (k-t BLAST) for the evaluation of left ventricular (LV) volumes in comparison to standard multiple-breathhold cine imaging, using a 3.0 Tesla (3T) MR system.
MATERIALS AND METHODS: In 23 subjects, short-axis cine loops completely covering the LV were acquired using conventional turbo gradient echo (GRE) imaging. Immediately after administration of gadobenate dimeglumine, a rapid single-breathhold k-t BLAST 4D dataset with the same coverage was acquired and reconstructed to short-axis views. Quantitative aortic flow measurement for LV stroke volume (LVSV) was used to calibrate both techniques. For GRE and k-t BLAST cine imaging: LV volumes, ejection fraction (EF), and blood-to-myocardium-contrast (BMC) were determined.
RESULTS: k-t BLAST and GRE sequences showed a strong correlation for LV volumes and EF (r = 0.97-0.99; P < 0.001). Excellent agreement was also found between the LVSV determined by aortic flow measurements and LVSV assessed using GRE sequence and k-t BLAST sequence. BMC of GRE was similar to that of k-t BLAST cine imaging.
CONCLUSION: The use of the single-breathhold 4D k-t BLAST technique for the assessment of LV volume is feasible and accurate in 3T MRI. (c) 2008 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18425828     DOI: 10.1002/jmri.21332

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  8 in total

1.  Functional cardiac MR imaging with true fast imaging with steady-state free precession before and after intravenous injection of contrast medium: comparison of image quality and accuracy.

Authors:  Gabriele A Krombach; Thorsten Plum; Ralf Koos; Rainer Hoffmann; Ertunc Altiok; Nils A Krämer; Rolf W Günther; Felix Schoth
Journal:  Eur Radiol       Date:  2010-10-03       Impact factor: 5.315

2.  Free-breathing whole-heart 3D cine magnetic resonance imaging with prospective respiratory motion compensation.

Authors:  Mehdi H Moghari; Ashita Barthur; Maria E Amaral; Tal Geva; Andrew J Powell
Journal:  Magn Reson Med       Date:  2017-12-08       Impact factor: 4.668

3.  Three-dimensional cine MRI in free-breathing infants and children with congenital heart disease.

Authors:  Achim Seeger; Michael C Fenchel; Gerald F Greil; Petros Martirosian; Ulrich Kramer; Christiane Bretschneider; Joerg Doering; Claus D Claussen; Ludger Sieverding; Stephan Miller
Journal:  Pediatr Radiol       Date:  2009-10-02

4.  Single-Breathhold Four-Dimensional Assessment of Left Ventricular Morphological and Functional Parameters by Magnetic Resonance Imaging Using the VAST Technique.

Authors:  Carlos Eduardo Rochitte; Clerio F Azevedo; Miguel A Rosário; Maria H R Siqueira; Victor Monsão; Manoj Saranathan; Thomas K Foo; Roberto Kalil Filho; Giovanni G Cerri; José A F Ramires
Journal:  Open Cardiovasc Med J       Date:  2011-03-22

5.  Acceleration of tissue phase mapping by k-t BLAST: a detailed analysis of the influence of k-t-BLAST for the quantification of myocardial motion at 3T.

Authors:  Anja Lutz; Axel Bornstedt; Robert Manzke; Patrick Etyngier; G Ulrich Nienhaus; Volker Rasche
Journal:  J Cardiovasc Magn Reson       Date:  2011-01-11       Impact factor: 5.364

6.  Single breath-hold assessment of cardiac function using an accelerated 3D single breath-hold acquisition technique--comparison of an intravascular and extravascular contrast agent.

Authors:  Marcus R Makowski; Andrea J Wiethoff; Christian H P Jansen; Sergio Uribe; Victoria Parish; Andreas Schuster; Rene M Botnar; Aaron Bell; Christoph Kiesewetter; Reza Razavi; Tobias Schaeffter; Gerald F Greil
Journal:  J Cardiovasc Magn Reson       Date:  2012-07-31       Impact factor: 5.364

7.  Population-based studies of myocardial hypertrophy: high resolution cardiovascular magnetic resonance atlases improve statistical power.

Authors:  Antonio de Marvao; Timothy J W Dawes; Wenzhe Shi; Christopher Minas; Niall G Keenan; Tamara Diamond; Giuliana Durighel; Giovanni Montana; Daniel Rueckert; Stuart A Cook; Declan P O'Regan
Journal:  J Cardiovasc Magn Reson       Date:  2014-02-03       Impact factor: 5.364

8.  Left Ventricular Function Evaluation on a 3T MR Scanner with Parallel RF Transmission Technique: Prospective Comparison of Cine Sequences Acquired before and after Gadolinium Injection.

Authors:  Thibault Caspar; Anthony Schultz; Mickaël Schaeffer; Aïssam Labani; Mi-Young Jeung; Paul Thomas Jurgens; Soraya El Ghannudi; Catherine Roy; Mickaël Ohana
Journal:  PLoS One       Date:  2016-09-26       Impact factor: 3.240

  8 in total