The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone.

Suneel Battula1, Andrew J Schoenfeld, Vivek Sahai, Gregory A Vrabec, Jason Tank, Glen O Njus.   

Abstract

BACKGROUND: All surgical screws can experience failure if the torsional, tensile, and flexion loads exerted on the screws are excessively high. The use of self-tapping screws (STS) results in higher insertion torques (IT) as these screws cut their own threads in the pilot hole drilled in the bone. In this study, the torque for inserting the STS into an osteoporotic bone block for different pilot hole sizes (PHS) was measured and the pullout strength (PS) for extraction of the screws was determined for different depths of insertion, 0 mm, 1 mm, and 2 mm beyond the far cortex.
METHODS: Seventy-two Synthes stainless steel STS (40 mm length and 3.5 mm diameter) were inserted into pilot holes of sizes 2.55 (A: 73% OD), 2.50 (B: 71.5%), 2.45 (C: 70%), and 2.8 mm (D: 80%). Using a digital torque screwdriver, screws were inserted to 0 mm, 1 mm or 2 mm past the far cortex. Pullout tests were conducted with an Instron materials testing system. Analysis of variance and Student-Neuman-Keuls tests were performed to determine the effect of DOI and PHS on the loading energy, PS, and IT.
RESULTS: Results demonstrated that IT of the screws inserted into pilot holes A, B, and C were higher than those in D. It was also observed that PS and loading energy for 1 mm and 2 mm penetration past the far cortex were higher than those for 0 mm regardless of PHS. This study also found that an increase in PHS to 2.8 mm will reduce IT but will also reduce the PS relative to a PHS of 2.5 mm, the current standard for 3.5 mm screws.
CONCLUSIONS: The results of previously published studies regarding the effect of pilot hole size on PS in healthy cortical bone cannot be applied to the osteoporotic environment. The findings presented in this research support using PHS no larger than 71.5% of the screw outer diameter (i.e., pilot hole size of 2.5 mm for 3.5 mm screws) and inserting screws at least 2 mm beyond the far cortex to maximize PS and minimize iatrogenic damage in osteoporotic bone.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18404066     DOI: 10.1097/TA.0b013e31802bf051

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  6 in total

1.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

2.  Optimizing Hybrid Plate Fixation with a Locked, Oblique End Screw in Osteoporotic Fractures.

Authors:  Aaron Wynkoop; Osy Ndubaku; Paul M Charpentier; Jeffrey B Peck; Norman E Walter; Patrick Atkinson
Journal:  Iowa Orthop J       Date:  2017

3.  Kirschner wire prepared pilot holes improve screw pullout strength in synthetic osteoporotic-type bone.

Authors:  Hrayr G Basmajian; Joseph N Liu; Travis Scudday; Seth T Campbell; Nirav H Amin
Journal:  J Clin Orthop Trauma       Date:  2019-08-22

4.  Embalmed and fresh frozen human bones in orthopedic cadaveric studies: which bone is authentic and feasible?

Authors:  Tobias Topp; Thorben Müller; Sebastian Huss; Peter Herbert Kann; Eberhard Weihe; Steffen Ruchholtz; Ralph Peter Zettl
Journal:  Acta Orthop       Date:  2012-09-14       Impact factor: 3.717

5.  Physical and mechanical characterization of PLLA interference screws produced by two stage injection molding method.

Authors:  Ali Reza Sadeghi-Avalshahr; Mohammad Khorsand-Ghayeni; Samira Nokhasteh; Amir Mahdi Molavi; Mohammad Sadeghi-Avalshahr
Journal:  Prog Biomater       Date:  2016-10-22

Review 6.  Designs and techniques that improve the pullout strength of pedicle screws in osteoporotic vertebrae: current status.

Authors:  Thomas M Shea; Jake Laun; Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2014-03-03       Impact factor: 3.411