Literature DB >> 18404035

Fit assessment of anatomic plates for the distal medial tibia.

Beat Schmutz1, Martin E Wullschleger, Henry Kim, Hansrudi Noser, Michael A Schütz.   

Abstract

OBJECTIVES: With the development and popularization of minimally invasive surgical methods and implants for fracture fixation, it is increasingly important that the available implants are precontoured to the specific anatomic location for which they are designed. The objective of this study was to develop a noninvasive method and criteria for quantifying the fit of a distal periarticular medial tibia plate and to test the method on a small set of tibia models.
METHODS: The undersurface of the plate was extracted from a digital model of the plate. The surface of the plate was fitted to 21 computer tomography (CT)-based 3-dimensional (3-D) models of human tibiae. Four criteria were defined that constitute an anatomic plate fit and subsequently were applied for the quantitative fit assessment. The fitting of the plate undersurface to the bone was entirely conducted in a virtual environment.
RESULTS: An anatomic fit of the plate was achieved for 4 of the models (19%). The individual categories generated fits of 62% (n = 13) for the proximal end; 43% (n = 9) for the proximal angle; 57% (n = 12) for the middle distance; and 57% (n = 12) for a distal fit.
CONCLUSIONS: Although for the 4 individual criteria plate fits of 43%-62% were achieved, a global/anatomic fit only occurred for 19% of the bone models. This outcome is likely a result of bone morphology variations, which exist in a random population sample combined with the effects of a nonoptimized plate shape. Recommendations for optimizing the fit of the plate are discussed.

Entities:  

Mesh:

Year:  2008        PMID: 18404035     DOI: 10.1097/BOT.0b013e31816b10ba

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  6 in total

1.  Potentialities and limitations of a database constructing three-dimensional virtual bone models.

Authors:  F Radetzki; T Mendel; H Noser; D Stoevesandt; M Röllinghoff; N Gutteck; K S Delank; D Wohlrab
Journal:  Surg Radiol Anat       Date:  2013-04-10       Impact factor: 1.246

2.  Anatomical fitting of a plate shape directly derived from a 3D statistical bone model of the tibia.

Authors:  Beat Schmutz; Kanchana Rathnayaka; Thomas Albrecht
Journal:  J Clin Orthop Trauma       Date:  2019-04-25

3.  Morphometric Analysis of the Clavicles in Chinese Population.

Authors:  Jesse Chieh-Szu Yang; Kun-Jhih Lin; Hung-Wen Wei; Cheng-Lun Tsai; Kang-Ping Lin; Pei-Yuan Lee
Journal:  Biomed Res Int       Date:  2017-04-12       Impact factor: 3.411

4.  Biomechanical evaluation of a novel anatomical plate for oblique lumbar interbody fusion compared with various fixations: a finite element analysis.

Authors:  Weibo Huang; Ye Tian; Hongli Wang; Jianyuan Jiang; Xiaosheng Ma; Feizhou Lv
Journal:  Ann Transl Med       Date:  2022-08

5.  Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs.

Authors:  Beat Schmutz; Jayani Amarathunga; Stanley Kmiec; Prasad Yarlagadda; Michael Schuetz
Journal:  J Orthop Surg Res       Date:  2016-04-27       Impact factor: 2.359

6.  Analyzing the Fitting of Novel Preformed Osteosynthesis Plates for the Reduction and Fixation of Mandibular Fractures.

Authors:  Marc Anton Fuessinger; Mathieu Gass; Caroline Woelm; Carl-Peter Cornelius; Ruediger M Zimmerer; Philipp Poxleitner; Stefan Schlager; Marc Christian Metzger
Journal:  J Clin Med       Date:  2021-12-20       Impact factor: 4.241

  6 in total