Leakage correction estimate for electret ion chamber dielectric material used for long-term environmental gamma monitoring.

David F Jones1, L R Paulus.   

Abstract

The Idaho Department of Environmental Quality INL Oversight Program (DEQ-INL) operates an environmental gamma radiation detection network consisting of a series of high-pressure ion chambers to provide real-time ambient radiation measurements and a series of passive environmental electret ion chambers to increase coverage area and measure cumulative dose over a calendar quarter. DEQ-INL has identified a consistent over-response of approximately 40% by the electret ion chambers with respect to co-located high-pressure ion chambers since 1998. DEQ-INL conducted a series of three investigations to quantify this over-response. The over-response is likely attributable to a number of factors, including inherent voltage loss by the electret material not due to ionization within the chamber. One aspect of the investigation verified the manufacturer's calibration factor used to convert decrease in voltage to exposure. Additional investigations were performed that identified an average electret voltage loss of 0.2 V d(-1). When this voltage correction was applied to historical environmental data, electret ion chamber response was within 10% of the co-located high-pressure ion chamber response.

Entities:  

Mesh:

Year:  2008        PMID: 18403969     DOI: 10.1097/01.HP.0000299276.52267.3b

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   1.316


  1 in total

1.  PROCEDURES TO MEASURE MEAN AMBIENT DOSE EQUIVALENT RATES USING ELECTRET ION CHAMBERS.

Authors:  F Leontaris; A Boziari; A Clouvas; M Kolovou; J Guilhot
Journal:  Radiat Prot Dosimetry       Date:  2020-08-03       Impact factor: 0.972

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.