Literature DB >> 18377477

Major complications associated with transcatheter atrial septal occluder implantation: a review of the medical literature and the manufacturer and user facility device experience (MAUDE) database.

Jeffrey W Delaney1, Jennifer S Li, John F Rhodes.   

Abstract

OBJECTIVE: To summarize major complications and outcome for patients receiving percutaneous closure of atrial septal communications.
DESIGN: The Medline database and the United States Food and Drug Administration manufacturer and user facility device experience databases (MAUDE) were searched for reports related to complications with atrial septal occluding devices. The medical literature documenting complication rates for these devices were reviewed and summarized. The MAUDE database complication reports were compared with those reported in the medical literature using national implant estimates. OUTCOME: The MAUDE database correlated in the type of complications most frequently encountered with each device. However, based on estimated total implant numbers, there is a higher incidence of major complications, including death. AGA devices had a 0.3% erosion/perforation rate with a higher morbidity and mortality (29%) than previously reported. NMT devices had a lower incidence of erosion/perforation rate of 0.05%. Embolization rates for the NMT devices were also lower than published European studies, possibly reflecting the US restriction of the device for closure of patent foramen ovale. Thrombus was more frequently encountered on the NMT device. Both AGA and NMT devices have been shown to be safe and effective alternatives to cardiac surgery. The MAUDE database correlated, with a very low overall complication rate, but showed a higher estimated major complication rate than the medical literature. These data demonstrate the difficulty in quantifying rare complications in the premarketing analysis and the obligation providers have to report and evaluate complications through vigilant postmarketing surveillance.

Entities:  

Mesh:

Year:  2007        PMID: 18377477     DOI: 10.1111/j.1747-0803.2007.00107.x

Source DB:  PubMed          Journal:  Congenit Heart Dis        ISSN: 1747-079X            Impact factor:   2.007


  11 in total

Review 1.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

2.  Recent changes in instructions for use for the Amplatzer atrial septal defect occluder: how to incorporate these changes while using transesophageal echocardiography or intracardiac echocardiography?

Authors:  Kiran Mallula; Zahid Amin
Journal:  Pediatr Cardiol       Date:  2012-05-11       Impact factor: 1.655

3.  Risk factors and prognosis of atrioventricular block after atrial septum defect closure using the Amplatzer device.

Authors:  Yibin Wang; Yimin Hua; Li Li; Xiaoqin Wang; Lina Qiao; Xiaoqing Shi; Jiping Hua; Yi Qu; Dezhi Mu
Journal:  Pediatr Cardiol       Date:  2013-10-30       Impact factor: 1.655

Review 4.  State-of-the-Art Atrial Septal Defect Closure Devices for Congenital Heart.

Authors:  Michael L O'Byrne; Daniel S Levi
Journal:  Interv Cardiol Clin       Date:  2019-01

5.  Increasing propensity to pursue operative closure of atrial septal defects following changes in the instructions for use of the Amplatzer Septal Occluder device: An observational study using data from the Pediatric Health Information Systems database.

Authors:  Michael L O'Byrne; Russell T Shinohara; Elena K Grant; Joshua P Kanter; Matthew J Gillespie; Yoav Dori; Jonathan J Rome; Andrew C Glatz
Journal:  Am Heart J       Date:  2017-07-19       Impact factor: 4.749

6.  Validation of a Prescreening Program for Transcatheter Atrial Septal Defect Closure.

Authors:  George T Nicholson; Robert N Vincent; Christopher J Petit; Meredith Roman; Michelle Glanville; Dennis W Kim
Journal:  Pediatr Cardiol       Date:  2015-03-20       Impact factor: 1.655

7.  The influence of deficient retro-aortic rim on technical success and early adverse events following device closure of secundum atrial septal defects: An Analysis of the IMPACT Registry®.

Authors:  Michael L O'Byrne; Matthew J Gillespie; Kevin F Kennedy; Yoav Dori; Jonathan J Rome; Andrew C Glatz
Journal:  Catheter Cardiovasc Interv       Date:  2016-05-18       Impact factor: 2.692

Review 8.  Transcatheter device closure of atrial septal defects: more to think about than just closing the hole.

Authors:  Michael L O'Byrne; Andrew C Glatz; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2018-01       Impact factor: 2.161

9.  Prevalence of deficient retro-aortic rim and its effects on outcomes in device closure of atrial septal defects.

Authors:  Michael L O'Byrne; Andrew C Glatz; Sherzana Sunderji; Aswathi E Mathew; David J Goldberg; Yoav Dori; Jonathan J Rome; Matthew J Gillespie
Journal:  Pediatr Cardiol       Date:  2014-05-14       Impact factor: 1.655

10.  Analysis of the US Food and Drug Administration Manufacturer and User Facility Device Experience database for adverse events involving Amplatzer septal occluder devices and comparison with the Society of Thoracic Surgery congenital cardiac surgery database.

Authors:  Daniel J DiBardino; Doff B McElhinney; Aditya K Kaza; John E Mayer
Journal:  J Thorac Cardiovasc Surg       Date:  2009-06       Impact factor: 5.209

View more