Literature DB >> 18377319

Biomechanical comparison of three fixation techniques for unstable thoracolumbar burst fractures. Laboratory investigation.

Frank L Acosta1, Jenni M Buckley, Zheng Xu, Jeffrey C Lotz, Christopher P Ames.   

Abstract

OBJECT: Increased structural stability is considered sufficient justification for higher-risk surgical procedures, such as circumferential fixation after severe spinal destabilization. However, there is little biomechanical evidence to support such claims, particularly after traumatic lumbar burst fracture. The authors sought out to compare the biomechanical performance of the following 3 fixation strategies for spinal reconstruction after decompression for an unstable thoracolumbar burst fracture: 1) short-segment anterolateral fixation; 2) circumferential fixation; and 3) extended anterolateral fixation.
METHODS: Thoracolumbar spines (T10-L4) from 7 donors (mean age at death 64+/-6 years; 1 female and 6 males) were tested in pure moment loading in flexion-extension, lateral bending, and axial rotation. Thoracolumbar burst fractures were surgically induced at L-1, and testing was repeated sequentially for each of the following fixation techniques: short-segment anterolateral, circumferential, and extended anterolateral. Primary and coupled 3D motions were measured across the instrumented site (T12-L2) and compared across treatment groups.
RESULTS: Circumferential and extended anterolateral fixations were statistically equivalent for primary and off-axis range-of-motions in all loading directions, and short-segment anterolateral fixation offered significantly less rigidity than the other 2 methods.
CONCLUSIONS: The results of this study strongly suggest that extended anterolateral fixation is biomechanically comparable to circumferential fusion in the treatment of unstable thoracolumbar burst fractures with posterior column and posterior ligamentous injury. In cases in which an anterior procedure may be favored for load sharing or canal decompression, extension of the anterior instrumentation and fusion one level above and below the unstable segment can result in near equivalent stability to a 2-stage circumferential procedure.

Entities:  

Mesh:

Year:  2008        PMID: 18377319     DOI: 10.3171/SPI/2008/8/4/341

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  9 in total

Review 1.  Role of lumbar interspinous distraction on the neural elements.

Authors:  Alex Alfieri; Roberto Gazzeri; Julian Prell; Christian Scheller; Jens Rachinger; Christian Strauss; Andreas Schwarz
Journal:  Neurosurg Rev       Date:  2012-05-02       Impact factor: 3.042

2.  Contribution of Round vs. Rectangular Expandable Cage Endcaps to Spinal Stability in a Cadaveric Corpectomy Model.

Authors:  Gregory M Mundis; Robert K Eastlack; Payam Moazzaz; Alexander W L Turner; G Bryan Cornwall
Journal:  Int J Spine Surg       Date:  2015-10-22

3.  Biomechanical evaluation of an interfacet joint decompression and stabilization system.

Authors:  Jeremi M Leasure; Jenni Buckley
Journal:  J Biomech Eng       Date:  2014-07       Impact factor: 2.097

4.  OPF/PMMA cage system as an alternative approach for the treatment of vertebral corpectomy.

Authors:  Asghar Rezaei; Hugo Giambini; Alan L Miller; Xifeng Liu; Benjamin D Elder; Michael J Yaszemski; Lichun Lu
Journal:  Appl Sci (Basel)       Date:  2020-10-02       Impact factor: 2.679

5.  A New Vertebral Body Replacement Strategy Using Expandable Polymeric Cages.

Authors:  Xifeng Liu; Alex Paulsen; Hugo Giambini; Ji Guo; A Lee Miller; Po-Chun Lin; Michael J Yaszemski; Lichun Lu
Journal:  Tissue Eng Part A       Date:  2016-12-26       Impact factor: 3.845

6.  Poly(ε-caprolactone) Dendrimer Cross-Linked via Metal-Free Click Chemistry: Injectable Hydrophobic Platform for Tissue Engineering.

Authors:  Xifeng Liu; A Lee Miller; Kevin A Fundora; Michael J Yaszemski; Lichun Lu
Journal:  ACS Macro Lett       Date:  2016-10-31       Impact factor: 7.015

7.  Biomechanical efficacy of monoaxial or polyaxial pedicle screw and additional screw insertion at the level of fracture, in lumbar burst fracture: An experimental study.

Authors:  Hongwei Wang; Changqing Li; Tao Liu; Wei-Dong Zhao; Yue Zhou
Journal:  Indian J Orthop       Date:  2012-07       Impact factor: 1.251

8.  Anterolateral Corrective Lumbar Corpectomy and Interbody Fusion by Using Extended Screw Fixation without Posterior Instrumentation for Posttraumatic Kyphosis.

Authors:  Atsuro Yamazaki; Sumihisa Orita; Takeshi Sainoh; Kazuyo Yamauchi; Miyako Suzuki; Yoshihiro Sakuma; Go Kubota; Yasuhiro Oikawa; Kazuhide Inage; Yukio Nakata; Gen Inoue; Yasuchika Aoki; Tomoaki Toyone; Junichi Nakamura; Masayuki Miyagi; Kazuhisa Takahashi; Seiji Ohtori
Journal:  Case Rep Orthop       Date:  2013-07-18

9.  Biomechanical analysis of the camelid cervical intervertebral disc.

Authors:  Dean K Stolworthy; R Amy Fullwood; Tyler M Merrell; Laura C Bridgewater; Anton E Bowden
Journal:  J Orthop Translat       Date:  2014-12-23       Impact factor: 5.191

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.