| Literature DB >> 18373964 |
Kamila Syslová1, Petr Kacer, Marek Kuzma, Pavlína Klusácková, Zdena Fenclová, Jindriska Lebedová, Daniela Pelclová.
Abstract
Rapid and precise method for the determination of 8-iso-prostaglandin F(2alpha), an essential marker of the oxidative stress, in exhaled breath condensate (EBC) was developed. The protocol consisted of stable isotope dilution, immunoseparation combined with selective and sensitive LC-ESI-MS/MS operated in multiple reaction monitoring (MRM) mode. The imprecision of the developed method was below 8.8%, the parameter of mean inaccuracy was determined as <9.6% (0-250pg of 8-iso-prostaglandin F(2alpha)/ml EBC). The limit of detection (LOD) was 1 pg/ml EBC and limit of quantification (LOQ) 5 pg/ml EBC. A significant difference in 8-iso-prostaglandin F(2alpha) content between the group of asbestosis patients and healthy volunteers was found.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18373964 DOI: 10.1016/j.jchromb.2008.02.019
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205