Literature DB >> 18363000

Mainstream end-tidal carbon dioxide monitoring in ventilated neonates.

Y R Bhat1, N Abhishek.   

Abstract

INTRODUCTION: Continuous noninvasive monitoring of the partial pressure of arterial carbon dioxide (PaCO2) in ventilated neonates would help clinicians to reduce arterial blood sampling. Our objective was to determine the correlation and agreement between end-tidal carbon dioxide (EtCO2) and PaCO2 in newborns ventilated for various clinical situations.
METHODS: This prospective study was undertaken over 15 months in a teaching hospital. Simultaneous end-tidal and arterial CO2 pairs were obtained from ventilated neonates who were monitored by mainstream capnography and had indwelling arterial catheter. The correlation coefficient and degree of bias between EtCO2 and PaCO2 were assessed for various clinical situations.
RESULTS: A total of 133 end-tidal and arterial CO2 pairs were analysed from 32 ventilated newborns. The mean gestational age was 34.6 +/- 3.8 weeks and birth weight was 2,200 +/- 780 g. The overall coefficient of correlation (r) was 0.73 (p-value is less than 0.001). The EtCO2 value was lower than the corresponding PaCO2 value in 86.5 percent pairs, with a mean bias of -6.65 +/- 7.54 mmHg (95 percent CI, - 7.9 to - 5.35). The r-value was more than or equal to 0.92 in neonates ventilated for sepsis, asphyxia and apnoea of prematurity, 0.67 in hyaline membrane disease (HMD) and 0.69 in meconium-aspiration syndrome. In HMD, neonates who received surfactant had a better r-value than those who did not (0.76 vs. 0.6).
CONCLUSION: The correlation between mainstream EtCO2 and PaCO2 is good. Neonates with pulmonary disease will have a lower correlation. Surfactant therapy improves the correlation. EtCO2 monitoring is helpful in trending or screening for abnormal PaCO2 values.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18363000

Source DB:  PubMed          Journal:  Singapore Med J        ISSN: 0037-5675            Impact factor:   1.858


  12 in total

1.  Mainstream time-capnography: an aid to select an appropriate uncuffed endotracheal tube in small children.

Authors:  Praveen Kumar Neema; Aveek Jayant; Manikandan Sethuraman; Ramesh C Rathod
Journal:  J Clin Monit Comput       Date:  2008-12-13       Impact factor: 2.502

2.  Beyond Anesthesia Toxicity: Anesthetic Considerations to Lessen the Risk of Neonatal Neurological Injury.

Authors:  Mary Ellen McCann; Jennifer K Lee; Terrie Inder
Journal:  Anesth Analg       Date:  2019-11       Impact factor: 5.108

Review 3.  A review of pediatric capnography.

Authors:  Naveen Eipe; Dermot R Doherty
Journal:  J Clin Monit Comput       Date:  2010-07-16       Impact factor: 2.502

4.  Detection of carbon dioxide thresholds using low-flow sidestream capnography in ventilated preterm infants.

Authors:  Emmanuel Lopez; Sophie Grabar; Alexandre Barbier; Baruch Krauss; Pierre-Henri Jarreau; Guy Moriette
Journal:  Intensive Care Med       Date:  2009-11       Impact factor: 17.440

Review 5.  Correlation of End-Tidal Carbon Dioxide with Arterial Carbon Dioxide in Mechanically Ventilated Neonates: A Scoping Review.

Authors:  Ingra Pereira Monti Martins; Adriane Muller Nakato; Paula Karina Hembecker; Sérgio Ossamu Ioshii; Percy Nohama
Journal:  Glob Pediatr Health       Date:  2021-05-17

6.  Correlation of end tidal and arterial carbon dioxide levels in critically Ill neonates and children.

Authors:  Hiren Mehta; Rahul Kashyap; Sangita Trivedi
Journal:  Indian J Crit Care Med       Date:  2014-06

7.  End-tidal carbon dioxide measurement in preterm infants with low birth weight.

Authors:  Hsin-Ju Lin; Ching-Tzu Huang; Hsiu-Feng Hsiao; Ming-Chou Chiang; Mei-Jy Jeng
Journal:  PLoS One       Date:  2017-10-17       Impact factor: 3.240

8.  Effect of corrective or palliative procedures on arterial to end-tidal carbon dioxide pressure difference in pediatric cardiac surgery.

Authors:  Eissa Bilehjani; Solmaz Fakhari; Alireza Yaghoubi; Yashar Eslampoor; Simin Atashkhoei; Mousa Mirinajad
Journal:  Afr J Paediatr Surg       Date:  2018 Apr-Jun

9.  Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children.

Authors:  Muhterem Duyu; Anıl Dogan Bektas; Zeynep Karakaya; Meral Bahar; Aybuke Gunalp; Yasemin Mocan Caglar; Meryem Nihal Yersel; Ozlem Bozkurt
Journal:  Sci Rep       Date:  2020-12-16       Impact factor: 4.379

10.  Accuracy of a battery-powered portable capnometer in premature infants.

Authors:  Eiji Hirakawa; Satoshi Ibara
Journal:  J Clin Monit Comput       Date:  2021-01-01       Impact factor: 1.977

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.