Literature DB >> 18334858

Assessment of contact location in subthalamic stimulation for Parkinson's disease by co-registration of computed tomography images.

Fumiaki Yoshida1, Yasushi Miyagi, Takato Morioka, Kimiaki Hashiguchi, Nobuya Murakami, Kenichi Matsumoto, Shinji Nagata, Tomio Sasaki.   

Abstract

AIMS: To study the validity of a co-registration method for postimplantation computed tomography (CT) images for localizing the location of an intracranial deep brain stimulator.
METHODS: Three-dimensional reconstruction images of postimplantation CTs were co-registered with preoperative CTs by stereotactic planning software and used to localize 18 leads in the subthalamic nuclei of 10 patients with Parkinson's disease. Our conventional method using superimposition of sagittal postimplantation magnetic resonance (MR) images were employed as a comparison.
RESULTS: The co-registered CT images separately visualized four individual contacts; on the other hand, the MR superimposition method demonstrated the leads as a group of no MR signal areas. Although laterality of the distal contact did not differ between two methods, the distal contact was located more anteriorly and inferiorly by the MR superimposition method than by the CT co-registration method.
CONCLUSION: The CT co-registration method is superior to MR in visualizing deep brain stimulation contacts and resolves problems of MR safety in patients treated with a neurostimulator. 2008 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2008        PMID: 18334858     DOI: 10.1159/000120428

Source DB:  PubMed          Journal:  Stereotact Funct Neurosurg        ISSN: 1011-6125            Impact factor:   1.875


  3 in total

1.  Evaluation of electrode position in deep brain stimulation by image fusion (MRI and CT).

Authors:  I Barnaure; P Pollak; S Momjian; J Horvath; K O Lovblad; C Boëx; J Remuinan; P Burkhard; M I Vargas
Journal:  Neuroradiology       Date:  2015-05-29       Impact factor: 2.804

Review 2.  Post-operative imaging in deep brain stimulation: A controversial issue.

Authors:  Christian Saleh; Georges Dooms; Christophe Berthold; Frank Hertel
Journal:  Neuroradiol J       Date:  2016-03-30

3.  Comparison of Intraoperative 3-Dimensional Fluoroscopy With Standard Computed Tomography for Stereotactic Frame Registration.

Authors:  Terrance Peng; Daniel R Kramer; Morgan B Lee; Michael F Barbaro; Li Ding; Charles Y Liu; Spencer Kellis; Brian Lee
Journal:  Oper Neurosurg (Hagerstown)       Date:  2020-06-01       Impact factor: 2.703

  3 in total