Qualitative dosimetric and radiobiological evaluation of high - dose - rate interstitial brachytherapy implants.

Than S Kehwar1, Syed F Akber, Kamlesh Passi.   

Abstract

Radiation quality indices (QI), tumor control probability (TCP), and normal tissue complication probability(NTCP) were evaluated for ideal single and double plane HDR interstitial implants. In the analysis, geometrically-optimized at volume (GOV) treatment plans were generated for different values of inter-source-spacing (ISS) within the catheter, inter-catheter-spacing (ICS), and inter-plane-spacing (IPS) for single - and double - plane implants. The dose volume histograms (DVH) were generated for each plan, and the coverage volumes of 100%, 150%, and 200% were obtained to calculate QIs, TCP, and NTCP. Formulae for biologically effective equivalent uniform dose (BEEUD), for tumor and normal tissues, were derived to calculate TCP and NTCP. Optimal values of QIs, except external volume index (EI), and TCP were obtained at ISS = 1.0 cm, and ICS = 1.0 cm, for single-plane implants, and ISS = 1.0 cm, ICS = 1.0 cm, and IPS = 0.75 to 1.25 cm, for double - plane implants. From this study, it is assessed that ISS = 1.0 cm, ICS = 1.0 cm, for single - plane implant and IPS between 0.75 cm to 1.25 cm provide better dose conformity and uniformity.

Entities:  

Keywords:  HDR interstitial implants; biologically effective equivalent uniform dose.; geometrical – optimization at volume; inter-catheter-spacing; inter-source-spacing; quality indices

Mesh:

Year:  2008        PMID: 18311328      PMCID: PMC2253950          DOI: 10.7150/ijms.5.41

Source DB:  PubMed          Journal:  Int J Med Sci        ISSN: 1449-1907            Impact factor:   3.738


INTRODUCTION

Use of computerized, remote controlled, high-dose-rate (HDR) brachytherapy units, and treatment planning systems provide conformal dose coverage to the target volume and minimum possible dose to surrounding normal tissues / critical organs. However, the basic principles of dosimetry systems 1 - 4 still influence the criteria of the source placement (activity distribution) and dose distributions in brachytherapy applications. In the HDR brachytherapy applications, such as in the treatment of carcinoma of the cervix (Ca.Cx.), the basic rules of the Manchester system 1 are still followed in many clinics World wide. In the HDR interstitial brachytherapy (ISBT) implants none of the classical dosimetry system 1 - 4 is followed. This is because modern HDR units have a high activity miniature type single stepping source, which offers an advantage of varying source positions (dwell positions) and time (dwell time) to a particular dwell position to obtain an appropriate dose distribution and isodose geometry. For HDR implants, a new dosimetry system, known as stepping source dosimetry system (SSDS) 5, has been devised in which source and dose distribution rules were formed using the selected basic rules of the Paris and the Manchester dosimetry systems with some modifications. Kwan et al 6 have done a computerized dosimetric study to determine optimal source and ribbon separation for single – plane implants, and the ribbon and plane separation of for double plane implants were studied with respect to the dose homogeneity, for single – and double – plane iridium – 192 (Ir – 192) implants. In another study of Quimby type breast implants, interplanar spacing, based on the implant sizes, was studied 7. None of the study has so far able to optimize these parameter for HDR single – and double – plane implants. In this work, we performed a computerized dosimetric study of HDR implants to find out optimal values of inter – source – spacing (ISS), within the catheter, and inter – catheter – spacing (ICS), within the target volume (TV), for ideal single plane implants. This was done by computing various radiation quality indices (QI) for geometrically optimized at volume (GOV) treatment plans. The GOV mode of optimization was chosen due to its simplicity, otherwise reader can choose any other suitable mode of optimization in practice. The inter – plane – spacing (IPS) for ideal double plane implants has also been determined using optimal values of ISS and ICS, obtained from single plane implants, by computing above said QIs for GOV treatment plans. The concept of Biologically Effective Equivalent Uniform Dose (BEEUD) has been introduced to calculate the tumor control probability (TCP) 8, 9 and normal tissue complications probability (NTCP) 10 for these HDR plans.

METHODS AND MATERIALS

Dosimetric parameters

To investigate ideal values of ISS, ICS and IPS in HDR interstitial implants, for which nearly ideal dose distribution within the target volume and maximum sparing of the surrounding normal tissues / organs, can be achieved, a quantitative analysis of ideal single and double plane implants has been done for different ISS and ICS values. The analysis is performed by computing and comparing different QIs, TCP, and NTCP for GOV treatment plans of these implants. The quality indices used in this study are: the coverage index (CI), the external volume index (EI), the relative dose homogeneity index (DHI), the overdose index (ODI), and the dose non-uniformity ratio (DNR), and are defined as: 1. : The fraction of the target volume that receives a dose equal to or greater than the reference dose 11. 2. : The ratio of the volume of normal tissue that receives a dose equal to or greater than the reference dose to the volume of the target 11. 3. : This is defined as the ratio of the target volume which receives a dose in the range of 1.0 to 1.5 times of the reference dose to the volume of the target that receives a dose equal to or greater than the reference dose 11. 4. This is the ratio of the target volume which receives a dose equal to or more than 2.0 times of the reference dose to the volume of the target that receives a dose equal to or greater than the reference dose 11. 5. : This is the ratio of the target volume which receives a dose equal to or greater than 1.5 times of the reference dose to the volume of the target which receives a dose equal to or greater than the reference dose 12. Conditions for an ideal implant are where the values of QIs should be as follows To compute above defined QIs, for single plane implants, ideal targets of target volumes of the dimensions of Length (L= 6.0 cm) × Width (W= 5.0 cm ) × Thickness (T = 1.0 cm) have been taken into account. While changing the values of ISS and / or ICS, sometimes extra length and width of target volume were also added to keep constant distance between target surface and peripheral dwell positions and / or target surface and peripheral catheters. The catheters and peripheral dwell positions were placed within 0.5 cm of the boundary of the target volume. Treatment plans were generated using PLATO (Nucletron BV, Veenendaal, The Netherlands) 3 – D treatment planning system. The dose points were placed on the surface of the target volume relative to the active dwell positions. All the dose points, in the implant, were used for dose normalization for the total dose of 42 Gy with 3.5 Gy per fraction. The Cumulative DVH (cDVH) for GOV treatment plans were generated for different values of ISS and ICS. The values of ICS vary from 0.5 cm to 2.0 cm, in steps of 0.25 cm. For each ICS values, the ISS varies from 0.25 cm to 2.0 cm in steps of 0.25 cm. In each treatment plan, the isodose surfaces of 100% (42 Gy), 150% (63 Gy) and 200% (84 Gy) were generated to find out the respective dose coverage volumes. By comparing the QIs for all treatment plans, the optimal values of ISS and ICS were obtained for which QIs to be the closest values of that of an ideal implant. The GOV treatment plans were also generated for double plane implants using optimal values of ISS and ICS, obtained from single plane implants QI analysis. The cDVHs were generated for inter – plane –spacing (IPS) vary from 0.5 cm to 2.0 cm in steps of 0.25 cm and the coverage volumes for the isodose surfaces of 100%, 150% and 200% were obtained from the cDVHs, as calculated for single plan implants, to compute the above said QIs for each treatment plan with different IPS value.

Radiobiological models

The linear quadratic (LQ) model provides a simple way to describe dose – response of different fractionation schemes, in terms of the Biologically Effective Dose (BED) 13. The BED for HDR ISBT 9 for a total dose of D (Gy) delivered with dose d (Gy) per fraction can be written by Where α/ß ratio is the tissue specific parameter and is the ratio of the coefficients of lethal damage to the sublethal damage, and G is the factor accounting for incomplete repair of sublethal damage during interfraction interval between the fractions. In this study, it is assumed that the time interval between the fractions is sufficient enough to allow the full repair of the sublethal damage, hence G is taken as 1. The tumor control probability (TCP) 8, 9 for uniform dose distribution within the target volume is given by Where ρ, V, α, and BEDt are the clonogenic cell density, target volume, coefficient of lethal damage (radio – sensitivity of lethal damage), and BED for the target, respectively. The dose distribution of HDR ISBT within target volume is highly non – uniform and has high dose gradient, hence equation (7) can not be directly applied to compute accurate TCP. Hence, to get an appropriate expression of TCP for HDR ISBT implant different regions of HDR ISBT implant have been considered (Figure 1). It is also shown that target volume is divided into four regions which are (1) the region which receives a dose less than the reference dose, (2) the region which receives a dose in the range of 1.0 to 1.5 times of the reference dose, (3) the region which receives a dose in the range of 1.5 to 2.0 times of the reference dose, and (4) the region which receives a dose equal to or more than 2.0 times of the reference dose. Each region of target volume has its own BEEUD. The expression of BEEUD, for tumor, is derived in Apendix – A, where it is considered that there is a non – uniform dose distribution within the target volume. The target volume is divided into 'n' number of voxels of small enough volume. So it can be assumed that the dose distribution within the voxel is uniform. The expression for BEEUD, given in equation (e) of Appendix – A, is written as
Figure 1

Schematic diagram showing target volume (TV), portion of target volume (TVDref) that receives dose equal to or more than the reference dose Dref, the isodose surface that receives 1.5 time of the reference dose (1.5 Dref), and that receives 2.0 times of the reference dose (2.0 Dref).

Where V is the target volume, vi is the volume of ith voxel of the target volume, and BEDti is the BED of the ith voxel of the target volume. The subscript 't' denotes the target volume. With the use of BEEUD of each region, shown in Figure 1, the TCP may be written as Where the terms TCP1, TCP2, TCP3, and TCP4 are the TCPs of above defined regions of the target volume, respectively. The expressions of these terms are given as follows 1. The TCP for the region of target volume which receives a dose less than the reference dose By rearranging and using the value of equation (1), we may write 2. The TCP for the region of target volume that receives a dose in the range of 1.0 to 1.5 times of the reference dose Using the value of equation (3), we may write 3. The TCP for the region of target volume that receives a dose in the range of 1.5 to 2.0 times of the reference dose By rearranging and using the values of equations (4) & (5), we may write 4. The TCP for the region of target volume that receives a dose equal to or greater than 2 times of the reference dose By using the value of equation (4), we may have the form of TCP4 Now multiplying and rearranging equations 9(a) – 9(d), the expression of net TCP may be given by Probably three radiobiological parameters, considered in the TCP formulation, such as clonogenic cell density (ρ), radio-sensitivity (α), and cell proliferation rate (Tp) influence the TCP phenomenological and are voxel dependent. In this work, it is assumed that first two parameters are constant throughout the target volume and influence of the cell proliferation rate is negligible. The radiobiologically based expression of normal tissue complication probability (NTCP) for uniform dose distribution within normal tissue / organ, was initially proposed by Kallman, et al 14 and was modified by Zaider and Amols 15. Kehwar and Sharma 16 and Kehwar 10 have further extended this model for the multiple component (MC) and the linear quadratic (LQ) models, respectively. These extended forms, of the NTCP model for MC and LQ models, were fitted to the normal tissue tolerance doses reported by Emami et al 17 at TD5/5 and TD50/5 for partial volumes of different normal tissues / organs. Kehwar's 10 NTCP equation of LQ model may be written as Where v and BEDn are the fractional partial volume (v=V/V0, here V and V0 are the partial volume and the reference volume of the normal tissue / organ, respectively) and BED of normal tissue / organ. The N0 and k are tissue-specific, non-negative adjustable parameters. The dose distribution outside the target volume within the adjacent normal tissue is highly non-uniform, hence equation (11) can not be applied to calculate NTCP for such a high dose gradient. For the purpose, entire volume of the normal tissue / organ is divided into two regions, viz (1) the region that receives a dose less than the reference dose, and (2) the region that receives a dose equal to or greater than the reference dose. Each region of normal tissue has its own BEEUD. The expression of BEEUD for normal tissue is derived in Apendix – B, where it is considered that there is a non-uniform dose distribution within the normal tissue, and is divided into 'n' number of very-very small sub-volumes (voxels). It has also been assumed that the dose distribution within a sub-volume is uniform. From equation (11), it is seen that the equation of NTCP is not an additive term of the volume, as TCP for TV, so the NTCP of voxels can not provide net NTCP of entire normal tissue. Therefore, equation (11) has been modified to account for addition of the volumes of the voxels, and the new term is known as the NTCP factor (NTCPF) which is written as Where V0 is the reference volume of the normal tissue / organ and Vi is the volume of ith voxel in the normal tissue / organ. The expression of BEEUD, from Appendix – B, for normal tissue is written by With the use of BEEUD of each region of normal tissue / organ, the NTCPF may be written as Where the terms NTCPFn1, and NTCPFn2 are the NTCPFs of above defined two normal tissue regions, respectively. The expressions of these terms are given as follows 1. The NTCPF for the region of normal tissue / organ which receives a dose less than the reference dose Where, V, is the normal tissue volume of the normal tissue / organ. By rearranging and using the value of equation (2), we may write 2. The NTCPF for the region of normal tissue / organ that receives a dose equal to or greater than the reference dose Using the value of equation (3), we may write By adding and rearranging equations (14a) and (14b), the net NTCPF will be written as The net NTCP from equation (15) is written by For statistical comparison, two tail unpaired t-student test is employed to the results of NO and GOV plans.

RESULTS

Dosimetric Analysis

a) Single Plane Implant

The curves were plotted for single plane implants between ISS and IQs, which are shown in Figures 2 to Figure 6. Figure 2 shows that the CI decreases from 0.98 to 0.97 for the values of ISS, which may be considered almost constant. The slope of the linear lines is -0.006 for all ICS values. The CI at ISS = 1.0 cm and ICS = 1.0 cm are 0.98 these plans.
Figure 2

A quantitative comparison of CI calculated for varying ISS and ICS values for NO and GOV plans for ideal HDR single plane interstitial implants.

Figure 6

Comparison of calculated DNR for varying ISS and ICS for GOV plans, of ideal HDR single plane interstitial implants.

Figure 3 shows that the value of EI increases in a linear trend insignificantly for all ICS values, and for any value of ISS. In these plans, the slopes of all linear lines remain almost constant with an average of 0.0012 (0.0012, 0.0013).
Figure 3

Comparison of calculated EI for varying ISS and ICS for GOV plans, of ideal HDR single plane interstitial implants.

It is clear from Figure 4 that initially the value of DHI increases with increasing ISS and ICS and reaches to a maximum value at ISS = 1.0 cm and ICS = 1.0 cm, and then decreases with ISS. The values of DHI at ISS = 1.0 cm and ICS = 1.0 cm are 0.851 for these plans.
Figure 4

Comparison of calculated DHI for varying ISS and ICS for GOV plans, of ideal HDR single plane interstitial implants.

The relation between ODI and ISS for different ICS values is given in Figure 5. It appears that the value of ODI decreases with increasing ISS and ICS and reaches to a minimum at ISS = 1.0 cm & ICS = 1.0 cm, thereafter it starts increasing with ISS and ICS. The values of ODI at ISS = 1.0 cm and ICS = 1.0 cm are 0.079 for these plans. Figure 6 shows similar relation between DNR and ISS as between ODI and ISS.
Figure 5

Comparison of calculated ODI for varying ISS and ICS for GOV plans, of ideal HDR single plane interstitial implants.

The calculated QIs for an ideal HDR implant reveals that at ISS = 1.0 cm and ICS = 1.0 cm, the values of DHI, ODI and DNR attain an optimal level. In this study, the values of QIs for single - plane implant at ISS =1.0 cm and ICS = 1.0 cm are CI = 0.98; EI = 0.062; DHI = 0.851; ODI = 0.079, and DNR = 0.149, respectively.

b) Double Plane Implant

For simplicity of the study, the best suitable values of ISS and ICS (ISS = 1.0 cm & ICS = 1.0 cm) for which DHI, ODI and DNR attain optimal values in single - plane implants, were used to construct the double plane implant. These values of ISS and ICS may not be optimal for double plane implants. The implant length and width were kept constant while the interplane separation (IPS) allowed to vary from 0.5 cm to 2.0 cm in steps of 0.25 cm. The GOV plans were generated for each IPS and to find out the volume coverage for 100%, 150% and 200% isodose surfaces the DVHs were generated. From above determined volumes, the QIs were computed and found that the variation of QIs with IPS is similar to that as of IQs with ISS. For IPS = 0.75 cm to 1.25 cm the QIs are optimal to treat a target of thickness from 1.75 to 2.25 cm. The values of QIs for IPS = 1.0 cm are CI = 0.978, EI = 0.08, DHI = 0.88, ODI = 0.09 and DNR = 0.29. If the IPS is further increased beyond 1.25 cm the DHI decreases and ODI increases, and a cold spot is generated between the two planes. Plots between QIs and IPS, for both type of plans, were similar to that for single plan implants, hence to avoid repetition of the figures, we have not included in this paper.

Radiobiological Analysis

BEEUDs have been calculated using equation (8) for different portions of the TV with the use of α/β = 10 Gy, α = 0.35 Gy-1 13, and clonogenic cell density ρ = 107 18 to calculate the net value of TCP using equation (10) for entire TV. In the calculation of BEEUD, for a particular region of the TV, the volume of that region is subdivided into very small sub-volumes and it is assumed that there is a uniform dose distribution within each sub-volume. The plots between net TCPs and ISS are shown in Figure 7, where the TCP for ICS = 1.0 cm and ISS = 1.0 cm implant is higher compared to other ICS and ISS settings.
Figure 7

Comparison of calculated TCP, based on LQ equation, for varying ISS and ICS for GOV plans, of ideal HDR single plane interstitial implants.

To calculate the NTCP for normal tissue, the normal tissue / organ is divided into two regions, (i) the region that receives a dose less than the reference dose, and (ii) the region that receives a dose equal to or greater than the reference dose. For demonstration purpose and to simulate the lung complications in breast HDR implants, the BEEUDs values were calculated for each region of the normal tissue / organ using derived values of the parameters 10, N0 = 3.93, k =1.03, for combined set of lung tolerance data, and α = 0.075 Gy-1, for lung tolerance data of Emami et al 17 and published values of α/β = 6.9 Gy 19, 20. The plots of NTCP and ISS for different ICS setting are shown in Figure 8, where it is seen that the value of NTCP increases with increasing ICS and ISS and highest value was found at ICS = 2.0 cm and ISS = 2.0 cm. Similar results were obtained for double plan implants, but to avoid repetition, the figures have not been included.
Figure 8

Comparison of calculated NTCP, based on LQ equation, for varying ISS and ICS GOV plans, of ideal HDR single plane interstitial implants.

DISCUSSION

A number of quality indices have been proposed to evaluate LDR and HDR interstitial implants, such as, DHI and DNR proposed by Saw and Suntharalingam 21, 22 and Saw et al 12 for LDR interstitial brachytherapy and was adopted by Meertens et al 11 for the evaluation of HDR interstitial implants. Hence, in this study we used the QI values as defined by Meertens et al 11. The expressions for TCP and NTCP incorporating above defined QIs were derived in this work, and the effects of the variation in ISS and ICS were investigated in GOV plans of ideal HDR interstitial implants. Figures – 7 and 8 show the effect of variation in ISS and ICS on TCP and NTCP. The calculations of TCP and NTCP, done by most of the investigators 8, 9, 10, 14, 15, 16, based on either entire target or normal tissue volume with a single dose or by dividing the entire volume in small voxels. In this work, we have opted different approach, where TV and NTV are divided into 4 and 2 parts to define target and normal tissue related QIs, respectively. The expressions of BEEUD were derived for these parts of TV and NTV, and were incorporated into the expression of the TCP and NTCP. In the Paris dosimetry system, designed for Ir – 192 wires and ribbons, suggests that to obtain a better coverage of the TV, one have to increase the active length of the catheters, and peripheral catheters have to be placed outside the target volume. But by doing so, this also increases the EI which consequently will increase the NTCP. Many researchers investigated this aspect and stated that active length of the catheters can be reduced compared to non optimized plans with uniform dwell times 5, 23, 24, 25 by properly optimizing the implant, because in optimization the dwell times of the dwell positions at the ends of the catheters and peripheral catheters are increased to compensate for the lack of source locations beyond the outermost dwell positions. Kwan et al 6 have reported that with respect to the dose homogeneity, within the implants, the optimal source and ribbon separation for single – plane implants was found to be 1.0 cm, and the ribbon and plane separation of 1.5 cm was found for double plane implants, maintaining a 1.0 cm source separation. Zwicker et al 7 found that interplanar spacing in Quimby type breast implants was implant size dependent. Major et al 26 have studied the effect of source step size and catheter separation on DNR for non-optimized and optimized interstitial breast HDR implants. In their study, the lowest value of DNR is reported for 10 mm source step size. The effect of catheter separation is studied at 5 mm source step size. The catheter separation was increased from 10 to 20 mm, for which the value of DNR reported to be increased from 0.15 to 0.22. While in the present study, QIs were evaluated for the above mentioned values of ISS and ICS and lowest DNR is found at ISS = 1.0 cm and ICS = 1.0 cm, where ISS =1.0 cm is same as 10 mm source step size. From the results of the present study, it can be concluded that in HDR ISBT implants the GOV provides an optimum outcome with regard to afore mentioned QIs, TCP, and NTCP, and is best achieved nearly uniform dose distribution within the implanted volume. While HDR ISBT implants done using classical dosimetry systems or non- optimization process lead to unsatisfactory results. It is also seen that on the basis of CI, EI and NTCP plots, no conclusion can be drawn as to what values of ICS and ISS (and IPS) would be used to optimize the single and double plane HDR implants. While DHI, ODI, DNR and TCP clearly reveal that their optimal values are at ICS = 1.0 cm and ISS = 1.0 cm, for single plane implant and ICS = 1.0 cm, ISS = 1.0 cm, and IPS = 0.75 to 1.25 cm, for double plane implants as shown in the study.
  21 in total

Review 1.  Tolerance of normal tissue to therapeutic irradiation.

Authors:  B Emami; J Lyman; A Brown; L Coia; M Goitein; J E Munzenrider; B Shank; L J Solin; M Wesson
Journal:  Int J Radiat Oncol Biol Phys       Date:  1991-05-15       Impact factor: 7.038

2.  Quantitative assessment of interstitial implants.

Authors:  C B Saw; N Suntharalingam
Journal:  Int J Radiat Oncol Biol Phys       Date:  1991-01       Impact factor: 7.038

Review 3.  The linear-quadratic formula and progress in fractionated radiotherapy.

Authors:  J F Fowler
Journal:  Br J Radiol       Date:  1989-08       Impact factor: 3.039

4.  Reference dose rates for single- and double-plane 192Ir implants.

Authors:  C B Saw; N Suntharalingam
Journal:  Med Phys       Date:  1988 May-Jun       Impact factor: 4.071

5.  Pulsed dose rate and fractionated high dose rate brachytherapy: choice of brachytherapy schedules to replace low dose rate treatments.

Authors:  A G Visser; G J van den Aardweg; P C Levendag
Journal:  Int J Radiat Oncol Biol Phys       Date:  1996-01-15       Impact factor: 7.038

6.  The Paris system in interstitial radiation therapy.

Authors:  B Pierquin; A Dutreix; C H Paine; D Chassagne; G Marinello; D Ash
Journal:  Acta Radiol Oncol Radiat Phys Biol       Date:  1978

7.  The clinical implications of non-standard fractionation.

Authors:  M Overgaard
Journal:  Int J Radiat Oncol Biol Phys       Date:  1985-06       Impact factor: 7.038

8.  Planning of Ir-192 seed implants for boost irradiation to the breast.

Authors:  R D Zwicker; R Schmidt-Ullrich; B Schiller
Journal:  Int J Radiat Oncol Biol Phys       Date:  1985-12       Impact factor: 7.038

Review 9.  Tumour and normal tissue responses to fractionated non-uniform dose delivery.

Authors:  P Källman; A Agren; A Brahme
Journal:  Int J Radiat Biol       Date:  1992-08       Impact factor: 2.694

10.  Spontaneous radiation-induced rib fractures in breast cancer patients treated with postmastectomy irradiation. A clinical radiobiological analysis of the influence of fraction size and dose-response relationships on late bone damage.

Authors:  M Overgaard
Journal:  Acta Oncol       Date:  1988       Impact factor: 4.089

View more
  5 in total

1.  Effect of applicator removal from target volume for cervical cancer patients treated with Venezia high-dose-rate brachytherapy applicator.

Authors:  ZhengZheng Xu; Bryan J Traughber; Eleanor Harris; Tarun K Podder
Journal:  J Contemp Brachytherapy       Date:  2022-03-24

2.  Role of high dose rate interstitial brachytherapy in early and locally advanced squamous cell carcinoma of buccal mucosa.

Authors:  Parthasarathy Vedasoundaram; Aravind Kumar Prasanna; Reddy Ks; Gangothri Selvarajan; Mourougan Sinnatamby; Seenisamy Ramapandian; Saravanan Kandasamy
Journal:  Springerplus       Date:  2014-10-09

3.  Biological effective doses in the intracavitary high dose rate brachytherapy of cervical cancer.

Authors:  B Arunkumar Sharma; Th Tomcha Singh; L Jaichand Singh; Y Indibor Singh; Y Sobita Devi
Journal:  J Contemp Brachytherapy       Date:  2011-12-30

4.  Treatment planning methods in high dose rate interstitial brachytherapy of carcinoma cervix: a dosimetric and radiobiological analysis.

Authors:  Surega Anbumani; Pichandi Anchineyan; Arunainambiraj Narayanasamy; Siddanna R Palled; Sajitha Sathisan; Punitha Jayaraman; Muthu Selvi; Ramesh S Bilimagga
Journal:  ISRN Oncol       Date:  2014-01-23

5.  A brachytherapy plan evaluation tool for interstitial applications.

Authors:  Surega Anbumani; N Arunai Nambiraj; Sridhar Dayalan; Kalaivany Ganesh; Pichandi Anchineyan; Ramesh S Bilimagga
Journal:  Adv Bioinformatics       Date:  2014-02-09