Literature DB >> 18307145

Integrated orbital implants--a comparison of hydroxyapatite and porous polyethylene implants.

S Ahmed Sadiq1, Lakbir S Mengher, John Lowry, Richard Downes.   

Abstract

PURPOSE: A retrospective case note review was undertaken to elucidate any differences in the cosmetic results and rate of serious complications between hydroxyapatite (HA) and porous polyethylene (PP) orbital implants.
METHODS: Patients who had undergone orbital implant surgery during the period 1993 to 1997 by a single surgeon were identified. Twenty-six patients had an HA implant and twenty-six had received a PP implant. All patients were reviewed in the ocular prosthetics department in a masked manner. The orbits were scored for post-enucleation socket syndrome (PESS) score, and quantification of implant mobility was done. Data analysis was conducted using Mann-Whitney U test for two independent samples.
RESULTS: There was no significant difference in PESS scores in patients with an HA or a PP implant. Implant elevation was better in patients with secondary (p = 0.029) and exchange (p = 0.042) PP implants. Only one patient in each of the HA and PP groups suffered significant implant exposure that required dermis fat and mucous membrane graft cover.
CONCLUSIONS: The complication rates and PESS scores were identical between the HA and PP implant groups. The implant mobility was better in the PP group. In clinical practice, the choice of the type of orbital implant inserted may also be governed by other factors such as surgeon experience, ease of use, and cost.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18307145     DOI: 10.1080/01676830701512585

Source DB:  PubMed          Journal:  Orbit        ISSN: 0167-6830


  5 in total

1.  [Porous orbital implants].

Authors:  B Cleres; H W Meyer-Rüsenberg
Journal:  Ophthalmologe       Date:  2014-06       Impact factor: 1.059

2.  Enucleation in malignant choroidal melanoma - results in 15 years of using a new material in the prosthesis of the orbital cavity.

Authors:  C P Tataru; M D Pop
Journal:  J Med Life       Date:  2012-06-18

Review 3.  Current concepts of ocular adnexal surgery.

Authors:  Maria Borrelli; Gerd Geerling
Journal:  GMS Interdiscip Plast Reconstr Surg DGPW       Date:  2013-02-27

4.  Comparison of pre-bent titanium mesh versus polyethylene implants in patient specific orbital reconstructions.

Authors:  Marcin Kozakiewicz; Piotr Szymor
Journal:  Head Face Med       Date:  2013-10-29       Impact factor: 2.151

Review 5.  Ocular implants-methods of ocular reconstruction following radical surgical interventions.

Authors:  Corina Teodora Catalu; Sânziana Luminiţa Istrate; Liliana Mary Voinea; Costin Mitulescu; Viorela Popescu; Ciuluvică Radu
Journal:  Rom J Ophthalmol       Date:  2018 Jan-Mar