Blade source effect on laser in situ keratomileusis flap thickness with the Amadeus I microkeratome.

Adrienne L Ruth1, Michael J Lynn, J Bradley Randleman, R Doyle Stulting.   

Abstract

PURPOSE: To determine the effect of different blades on laser in situ keratomileusis (LASIK) flap thickness created with the Amadeus I microkeratome (Ziemer Ophthalmic Systems).
SETTING: Emory University Department of Ophthalmology and Emory Vision, Atlanta, Georgia, USA.
METHODS: This retrospective nonrandomized comparative case study from January 2005 through June 2006 compared LASIK flap thickness created with blades from 2 manufacturers: the Surepass from Surgical Instrument Systems and distributed by AMO and the ML7090 CLB distributed by Med-Logics, Inc. Sex, preoperative corneal thickness, surgical-eye sequence, flap thickness and variance, and residual stromal bed were evaluated in each group.
RESULTS: This study evaluated 424 eyes of 226 patients. Surepass blades were used in 238 eyes and ML7090 CLB blades in 186 eyes. There were no significant differences between the 2 blade groups in preoperative corneal thickness, sex, or cases with corneal thickness greater than 550 microm. Mean flap thickness and variance were significantly lower in the ML7090 CLB group than in the Surepass group (P<.0001). There were no significant differences in flap thickness in either group based on sex; however, in both groups, flap thickness was significantly lower in second eyes and in eyes with a preoperative thickness less than 550 microm (P<.001).
CONCLUSIONS: The Amadeus I microkeratome created thinner, more consistent LASIK flaps with the ML7090 CLB blade than with the Surepass blade. Preoperative corneal thickness and eye sequence affected flap thickness, while sex did not.

Entities:  

Mesh:

Year:  2008        PMID: 18299064      PMCID: PMC3750739          DOI: 10.1016/j.jcrs.2007.11.020

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  18 in total

1.  Unexpected flap thickness in laser in situ keratomileusis.

Authors:  Osama Giledi; Sheraz M Daya
Journal:  J Cataract Refract Surg       Date:  2003-09       Impact factor: 3.351

2.  Clinical feature of unintended thin corneal flap in LASIK: 1-year follow-up.

Authors:  Hwan-Eok Yeo; Byung-Joo Song
Journal:  Korean J Ophthalmol       Date:  2002-12

3.  Accuracy and precision of the amadeus microkeratome in producing LASIK flaps.

Authors:  David W Jackson; Li Wang; Douglas D Koch
Journal:  Cornea       Date:  2003-08       Impact factor: 2.651

4.  Flap thickness accuracy: comparison of 6 microkeratome models.

Authors:  Kerry D Solomon; Eric Donnenfeld; Helga P Sandoval; Oday Al Sarraf; Terrance J Kasper; Mike P Holzer; Elizabeth H Slate; David T Vroman
Journal:  J Cataract Refract Surg       Date:  2004-05       Impact factor: 3.351

5.  Reproducibility of LASIK flap thickness using the Hansatome microkeratome.

Authors:  Osama Giledi; Mark G Mulhern; Marcela Espinosa; Andrea Kerr; Sheraz M Daya
Journal:  J Cataract Refract Surg       Date:  2004-05       Impact factor: 3.351

6.  Risk assessment for ectasia after corneal refractive surgery.

Authors:  J Bradley Randleman; Maria Woodward; Michael J Lynn; R Doyle Stulting
Journal:  Ophthalmology       Date:  2007-07-12       Impact factor: 12.079

7.  Influence of flap thickness on results of laser in situ keratomileusis for myopia.

Authors:  Barbara Prandi; Julio Baviera; Miguel Morcillo
Journal:  J Refract Surg       Date:  2004 Nov-Dec       Impact factor: 3.573

8.  Predictability of corneal flap thickness in laser in situ keratomileusis using three different microkeratomes.

Authors:  Gabi Shemesh; Gad Dotan; Isaac Lipshitz
Journal:  J Refract Surg       Date:  2002 May-Jun       Impact factor: 3.573

9.  Safety of laser in situ keratomileusis performed under ultra-thin corneal flaps.

Authors:  Robert T Lin; Stephanie Lu; Lan Lan Wang; Elli S Kim; Jarrod Bradley
Journal:  J Refract Surg       Date:  2003 Mar-Apr       Impact factor: 3.573

10.  Precision of flap measurements for laser in situ keratomileusis in 4428 eyes.

Authors:  Gerald W Flanagan; Perry S Binder
Journal:  J Refract Surg       Date:  2003 Mar-Apr       Impact factor: 3.573

View more
  3 in total

1.  Influence of flap thickness on visual and refractive outcomes after laser in situ keratomileusis performed with a mechanical keratome.

Authors:  Alok S Bansal; Terrence Doherty; J Bradley Randleman; R Doyle Stulting
Journal:  J Cataract Refract Surg       Date:  2010-05       Impact factor: 3.351

2.  Flap thickness in eyes with ectasia after laser in situ keratomileusis.

Authors:  J Bradley Randleman; Carolyn B Hebson; Paul M Larson
Journal:  J Cataract Refract Surg       Date:  2012-03-16       Impact factor: 3.351

3.  Analysis of microkeratome thin flap architecture using Fourier-domain optical coherence tomography.

Authors:  Karolinne Maia Rocha; J Bradley Randleman; R Doyle Stulting
Journal:  J Refract Surg       Date:  2011-08-19       Impact factor: 3.573

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.