| Literature DB >> 18295772 |
R Nageswara Rao1, N Venkateswarlu, R Narsimha.
Abstract
A robust and sensitive solid-phase extraction followed by liquid chromatography-electrospray ionization mass spectrometric (LC-ESI-MS) method for determination of antibiotics viz., fluoroquinolones, sulfamethoxazole, trimethoprim and cephalosporines in surface waters was developed. The sample recoveries on Oasis HLB cartridges were found to be >80%. Identification was carried out by LC-ESI-MS/MS. The positive ion ESI mass spectra containing the peaks of quasimolecular ions [M+H](+) allowed the determination of molecular masses whereas the fragment ions obtained by MS/MS of [M+H](+) ions permitted the structural assignment. Quantification was carried out by selective ion monitoring (SIM) using the quasimolecular ions [M+H](+) of the parent compounds. The detection and quantification limits were found to be in the range of 0.6-8.1 and 2.0-24.0 microg/L. The surface waters of different lakes and tanks of Hyderabad, India were found to contain a few antibiotics.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18295772 DOI: 10.1016/j.chroma.2008.02.021
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759