Evaluation of Al2O3:C optically stimulated luminescence (OSL) dosimeters for passive dosimetry of high-energy photon and electron beams in radiotherapy.

E G Yukihara1, G Mardirossian, M Mirzasadeghi, S Guduru, S Ahmad.   

Abstract

This article investigates the performance of Al2O3: C optically stimulated luminescence dosimeters (OSLDs) for application in radiotherapy. Central-axis depth dose curves and optically stimulated luminescence (OSL) responses were obtained in a water phantom for 6 and 18 MV photons, and for 6, 9, 12, 16, and 20 MeV electron beams from a Varian 21EX linear accelerator. Single OSL measurements could be repeated with a precision of 0.7% (one standard deviation) and the differences between absorbed doses measured with OSLDs and an ionization chamber were within +/- 1% for photon beams. Similar results were obtained for electron beams in the low-gradient region after correction for a 1.9% photon-to-electron bias. The distance-to-agreement values were of the order of 0.5-1.0 mm for electrons in high dose gradient regions. Additional investigations also demonstrated that the OSL response dependence on dose rate, field size, and irradiation temperature is less than 1% in the conditions of the present study. Regarding the beam energy/quality dependence, the relative response of the OSLD for 18 MV was (0.51 +/- 0.48)% of the response for the 6 MV photon beam. The OSLD response for the electron beams relative to the 6 MV photon beam. The OSLD response for the electron beams relative to the 6 MV photon beam was in average 1.9% higher, but this result requires further confirmation. The relative response did not seem to vary with electron energy at dmax within the experimental uncertainties (0.5% in average) and, therefore, a fixed correction factor of 1.9% eliminated the energy dependence in our experimental conditions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18293581     DOI: 10.1118/1.2816106

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry.

Authors:  Kevin E Casey; Paola Alvarez; Stephen F Kry; Rebecca M Howell; Ann Lawyer; David Followill
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

2.  Angular dependence of the nanoDot OSL dosimeter.

Authors:  James R Kerns; Stephen F Kry; Narayan Sahoo; David S Followill; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  Characterisation of energy response of Al(2)O(3):C optically stimulated luminescent dosemeters (OSLDs) using cavity theory.

Authors:  S B Scarboro; S F Kry
Journal:  Radiat Prot Dosimetry       Date:  2012-05-31       Impact factor: 0.972

4.  Measurement of characteristic prompt gamma rays emitted from oxygen and carbon in tissue-equivalent samples during proton beam irradiation.

Authors:  Jerimy C Polf; Rajesh Panthi; Dennis S Mackin; Matt McCleskey; Antti Saastamoinen; Brian T Roeder; Sam Beddar
Journal:  Phys Med Biol       Date:  2013-08-06       Impact factor: 3.609

5.  Flexible optically stimulated luminescence band for 1D in vivo radiation dosimetry.

Authors:  Tae Jin Kim; Kyung Oh Jung; Benjamin Fahimian; Guillem Pratx
Journal:  Phys Med Biol       Date:  2018-08-10       Impact factor: 3.609

6.  Evaluating the peak-to-valley dose ratio of synchrotron microbeams using PRESAGE fluorescence.

Authors:  N Annabell; N Yagi; K Umetani; C Wong; M Geso
Journal:  J Synchrotron Radiat       Date:  2012-03-15       Impact factor: 2.616

7.  On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy.

Authors:  Fasihah Hanum Yusof; Ngie Min Ung; Jeannie Hsiu Ding Wong; Wei Loong Jong; Vannyat Ath; Vincent Chee Ee Phua; Siew Ping Heng; Kwan Hoong Ng
Journal:  PLoS One       Date:  2015-06-08       Impact factor: 3.240

8.  Validation of measurement-guided 3D VMAT dose reconstruction on a heterogeneous anthropomorphic phantom.

Authors:  Daniel Opp; Benjamin E Nelms; Geoffrey Zhang; Craig Stevens; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

9.  Response of Nanodot Optically Stimulated Luminescence Dosimeters to Therapeutic Electron Beams.

Authors:  Y Retna Ponmalar; Ravikumar Manickam; S Sathiyan; K M Ganesh; R Arun; Henry Finlay Godson
Journal:  J Med Phys       Date:  2017 Jan-Mar

10.  Recent developments of optically stimulated luminescence materials and techniques for radiation dosimetry and clinical applications.

Authors:  A S Pradhan; J I Lee; J L Kim
Journal:  J Med Phys       Date:  2008-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.