Literature DB >> 18273913

Image-enhanced surgical navigation for endoscopic sinus surgery: evaluating calibration, registration and tracking.

R Lapeer1, M S Chen, G Gonzalez, A Linney, G Alusi.   

Abstract

BACKGROUND: Endoscopic sinus surgery (ESS) is generally applied to treat sinusitis when medication is not effective in eliminating the symptoms. Images captured by the endoscope are viewed on a monitor placed near the surgeon. Due to the separation of the handling of the endoscope from the viewing of the image, ESS requires surgeons to have well-trained hand-eye coordination. Unlike the use of the stereo surgical microscope in ENT, the endoscope does not provide the stereo cue for depth perception, hence a surgeon can only perceive depth through motion and shading, which may affect the accuracy of tool placement. Whilst the skill and experience of the surgeon are important factors to the success of ESS, the assistance of image-enhanced surgical navigation (IESN) can further reassure the surgeon's judgement and enhance surgical performance.
METHODS: We developed and validated an IESN system (ARView) for a rigid zero-degree endoscope, typically used for ESS. We present the interface, and calibration and registration (pre-operative and intra-operative) methods of the system. We then quantitatively assess the performance of each of the steps needed to generate the overlay of a real endoscope image with its 'virtual' counterpart, obtained from computed tomography (CT) image data of a real skull. These steps include calibration, registration, motion tracking and final overlay.
RESULTS: Calibration results using a planar calibration object displayed optimized object space errors of 0.025 +/- 0.013 mm, whilst a non-planar calibration object displayed errors of 0.12 +/- 0.08 mm. Target registration errors (TREs) near the region of interest (ROI), using our pre-operative registration method with the calibration object located near the mouth of the patient (skull), were 2.3 +/- 0.4 mm. The proposed photo-consistency method for intra-operative registration has not yet yielded satisfactory results for ESS-based IESN. (RMS) values for tracking accuracy were found to be around 1.2 mm in a typical workspace of 400 x 400 mm. Object space overlay errors in a small measurement volume of 10 x 10 x 10 mm were found to be around 0.4 +/- 0.02 mm.
CONCLUSIONS: We conclude that, in agreement with individual experiments, the current overall overlay accuracy is of the order of 2-3 mm in the x-y plane, which is in line with current conventional SN systems. The method which is most in need of improvement is registration, hence we wish to investigate the application of the proposed photo-consistency method further.

Entities:  

Mesh:

Year:  2008        PMID: 18273913     DOI: 10.1002/rcs.175

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  9 in total

1.  Overlay visualization in endoscopic ENT surgery.

Authors:  Christian Winne; Martin Khan; Fabian Stopp; Emanuel Jank; Erwin Keeve
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-06-26       Impact factor: 2.924

2.  [Clinical experiences with an automatic collision warning system: instrument navigation in endoscopic transnasal surgery].

Authors:  G Strauss; S Schaller; B Zaminer; S Heininger; M Hofer; D Manzey; J Meixensberger; A Dietz; T C Lüth
Journal:  HNO       Date:  2011-05       Impact factor: 1.284

3.  Intra-operative virtual endoscopy for image guided endonasal transsphenoidal pituitary surgery.

Authors:  Florian Schulze; Katja Bühler; André Neubauer; Armin Kanitsar; Leslie Holton; Stefan Wolfsberger
Journal:  Int J Comput Assist Radiol Surg       Date:  2009-09-04       Impact factor: 2.924

4.  [Virtual endoscopy of the nose and paranasal sinuses in real-time. Surgical planning system "Sinus endoscopy" (SPS-SE)].

Authors:  G Strauss; E Limpert; M Fischer; M Hofer; C Kubisch; A Krüger; A Dietz; J Meixensberger; C Trantakis; M Strauss; B Preim
Journal:  HNO       Date:  2009-08       Impact factor: 1.284

5.  Transorbital target localization with augmented ophthalmologic surgical endoscopy.

Authors:  Michael P DeLisi; Louise A Mawn; Robert L Galloway
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-09-12       Impact factor: 2.924

6.  Image-guided transorbital procedures with endoscopic video augmentation.

Authors:  Michael P DeLisi; Louise A Mawn; Robert L Galloway
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

7.  Toward Video-Based Navigation for Endoscopic Endonasal Skull Base Surgery.

Authors:  Daniel Mirota; Hanzi Wang; Russell H Taylor; Masaru Ishii; Gregory D Hager
Journal:  Med Image Comput Comput Assist Interv       Date:  2009-01-01

8.  Evaluation of a system for high-accuracy 3D image-based registration of endoscopic video to C-arm cone-beam CT for image-guided skull base surgery.

Authors:  Daniel J Mirota; Ali Uneri; Sebastian Schafer; Sajendra Nithiananthan; Douglas D Reh; Masaru Ishii; Gary L Gallia; Russell H Taylor; Gregory D Hager; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2013-01-28       Impact factor: 10.048

9.  Computed tomography-guided navigation assisted drainage for inaccessible deep neck abscess: A case report.

Authors:  Sung Min Han; Hee Sung Chae; Ha Neul Lee; Hyun Jong Jeon; Jeong Pyo Bong; Ji Hoon Kim
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.817