Extravascular signal decay rate R2 or R2 * as a function of blood oxygenation, geometry, and field strength was calculated using a Monte Carlo (MC) algorithm for a wider parameter range than hitherto by others. The relaxation rates of gradient-recalled-echo (GRE) and Hahn-spin-echo (HSE) imaging in the presence of blood vessels (ranging from capillaries to veins) have been computed for a wide range of field strengths up to 9.4T and 50% blood deoxygenation. The maximum HSE decay was found to be shifted to lower radii in higher compared to lower field strengths. For GRE, however, the relaxation rate was greatest for large vessels at any field strength. In addition, assessments of computational reliability have been carried out by investigating the influence of the time step, the Monte Carlo step procedure, boundary conditions, the number of angles between the vessel and the exterior field B0, the influence of neighboring vessels having the same orientation as the central vessel, and the number of proton spins. The results were compared with those obtained from a field distribution of the vessel computed by an analytic formula describing the field distribution of an ideal object (an infinitely long cylinder). It was found that the time step is not critical for values equal to or lower than 200 microseconds. The choice of the MC step procedure (three-dimensional Gaussian diffusion, constant one- or three-dimensional diffusion step) also failed to influence the results significantly; in contrast, the free boundary conditions, as well as taking too few angles into account, did introduce errors. Next neighbor vessels with the same orientation as the main vessel did not contribute significantly to signal decay. The total number of particles simulated was also found to play a minor role in computing R2/ R2 *.
Extravascular signal decay rate R2 or R2 * as a function of blood oxygenation, geometry, and field strength was calculated using a Monte Carlo (MC) algorithm for a wider parameter range than hitherto by others. The relaxation rates of gradient-recalled-echo (GRE) and Hahn-spin-echo (HSE) imaging in the presence of blood vessels (ranging from capillaries to veins) have been computed for a wide range of field strengths up to 9.4T and 50% blood deoxygenation. The maximum HSE decay was found to be shifted to lower radii in higher compared to lower field strengths. For GRE, however, the relaxation rate was greatest for large vessels at any field strength. In addition, assessments of computational reliability have been carried out by investigating the influence of the time step, the Monte Carlo step procedure, boundary conditions, the number of angles between the vessel and the exterior field B0, the influence of neighboring vessels having the same orientation as the central vessel, and the number of proton spins. The results were compared with those obtained from a field distribution of the vessel computed by an analytic formula describing the field distribution of an ideal object (an infinitely long cylinder). It was found that the time step is not critical for values equal to or lower than 200 microseconds. The choice of the MC step procedure (three-dimensional Gaussian diffusion, constant one- or three-dimensional diffusion step) also failed to influence the results significantly; in contrast, the free boundary conditions, as well as taking too few angles into account, did introduce errors. Next neighbor vessels with the same orientation as the main vessel did not contribute significantly to signal decay. The total number of particles simulated was also found to play a minor role in computing R2/ R2 *.
The effect of diffusion on signal decay in blood oxygenation level-dependent (BOLD)
imaging is mainly due to the extravascular contribution of spins, especially at
high field strengths >4 T [1]. In the current study, the well-known Monte Carlo (MC) approach modeling Brownian
diffusion of protons in a background magnetic field has been used to compute
extravascular (EV) BOLD signal changes. To this end, the static dipole model
presented previously [2] has been
extended to a dynamic model describing the sampling of phases of the individual
protons moving in the inhomogeneous magnetic field.Earlier studies on the effect of subvoxel variations in magnetic
susceptibility were reported by Fisel et al. [3].
Weisskoff et al. compared MC simulations with experiments with polystyrene
microspheres to demonstrate that enhanced relaxation can be explained
quantitatively for both spin-echo and gradient-echo experiments [4]. The effect of an endogenous
paramagnetic agent (deoxygenated hemoglobin) on image contrast has been
addressed by several authors, for example, Ogawa et al. [5], Kennan et al. [6], and Boxerman et al. [7]. All models
are based on the fact that in the vicinity of capillaries and venules, local
magnetic field distortions are generated by the presence of paramagnetic
deoxyhemoglobin in the blood.Data from the models in the literature so far have mostly been
restricted to a magnetic field strength of 1.5 T, that is, the clinical scanner
field strength in the past, and mostly for GRE only. However, nowadays, scanners
with high or ultra-high field strength for humans up to 9.4 T are available for
research, and EV-BOLD data for these field strengths both for GRE and HSE have not
yet been provided. The aim of the present work, therefore, was to investigate
these issues at such high magnetic field strengths.To examine the
contribution of extravascular spin in isolation from other factors, an
impenetrable vessel wall boundary for extravascular spins was assumed. The
range of investigated susceptibility values was determined using a
deoxygenation content of 5% at 1.5 T as the lowest susceptibility value and up
to 50% at 9.4 T as the highest value. In addition, standard approaches used in
the literature have been evaluated as to how they influence the computed
relaxation rates. In particular, the choice of the time step, the diffusion step,
the number of angles and the influence of neighboring vessels, and the number of
protons were examined.
2. THEORY
Our aim was to study signal decay due to the phase sampling of the
individual spins during their random movement. The spins in the brain
parenchyma are diffusing in a background magnetic field caused by deoxygenated
blood present in capillaries, venules, and veins. The field distribution is
therefore determined by vessels inside a computational volume, filled with
deoxygenated blood. The susceptibility creating the field distribution around
the blood vessel is proportional to the level of blood deoxygenation and to the
exterior field.Weisskoff et al. proposed generalizing their results obtained using a
numerical model by the use of the Bloch-Torrey equation [4]. Fujita also established a dimensionless equation which
is ruled by two parameters [8]. We
briefly recapitulate their arguments in the following paragraph and thereby
show how their theories relate to one other.Because the MC method solution must respect the Bloch-Torrey equations,
generalized scaling laws might be derived in advance to generalize the
numerical solutions [4]. If the
length scale is made dimensionless by , we obtain
If the time scale is made dimensionless by , we obtain
Substituting by and by TE, the general scale independent relation
with and can be established. From this equation, (3), (1), and (2) follow as special cases. The minus sign in (3) indicates
the direction of rotation in the complex plane and therefore can be omitted. In
the present article, we use cgs units with
μm/ms and TE = 40 milliseconds.
3. MATERIAL AND METHODS
3.1. Flow chart
Our model is similar to that of Boxerman et al. [7]. A flow chart illustrating the model is shown in Figure 1. The total signal
is computed from individual complex transverse magnetizations of the protons,
which are accumulating phases in the magnetic background field. The magnetic
background field is described either by an analytic formula, or by integrating
the individual contributions from the discretized susceptibility distribution.
Figure 1
Flow chart of the model. It consists
of the dipole model and a Monte Carlo time-step procedure. The inner of the two
3D loops is substituted in the present work by a known field distribution for
the element (i.e., the field distribution for an infinite vessel).
We used the field distribution caused by a vessel segment approximated
by a paramagnetic cylinder with infinite length to be the susceptibility
distribution for an infinite vessel, which is given by
where is the angle between the cylinder and the
direction of the static magnetic field, r and represent the nondimensional radial coordinate
relative to the vessel radius R and the azimuthal angle in the plane orthogonal
to the cylinder, respectively, and is the susceptibility-induced maximal
frequency shift occurring at the vessel surface, given by
In (5), Hct denotes the haematocrit value, that is, the fraction of the
volume taken up by the red blood cells, is the volume susceptibility difference per
unit Hct between fully oxygenated and fully deoxygenated blood, is the static magnetic field strength in terms
of frequency, and Y is the degree of blood oxygenation.Our parameter space was given by a haematocrit value of Hct = 0.4, a
susceptibility of fully deoxygenated blood of 0.18 ppm (cgs units). Frequency thus varied
from 1.4454 Hz (corresponding to 1−Y = 5% at 1.5 T) to 65 × 1.4454 Hz (90.58 Hz corresponding to 1−Y = 50% at 9.4 T) in steps of 5.7816 Hz. The radii varied from 3 μm to 60 μm in logarithmical
equidistant values.Alternatively, one could discretize the paramagnetic cylinder
representing the vessel into elementary volumes, for example, of size of the
computational cell. In the dipole model, to each elementary volume, a dipole
strength is assigned, where M is
the magnetization proportional to the exterior field strength . The elementary volume then contributes to the field distribution by an
amount of .To compute the total field distortion B
, a spatial integration
to summate all contributions has to be performed. Diffusion was computed
according to 3 models which can be distinguished by the choice of the diffusion
step. They are given by the following definitions.
STEP1D
Take a fixed step () in each direction (along the positive or
negative axis, directions chosen randomly).
STEP3D
Take a fixed step () in a randomly chosen direction.
GAUSS3D
Take a random step with a Gaussian distribution () in a randomly chosen direction.In the formulae, is the diffusion coefficient and is the time step. As diffusion strength, we used
μm/ms, which is
typical for the cerebral cortex [7]. As a random generator, the routine ran1 from
the numerical recipes [9] was
chosen. The various MC step methods were first tested by computing the radial
distribution of the protons after N time steps for M protons. Unless stated
otherwise, the time step was chosen as 10 microseconds and TE was 40 milliseconds at
maximum, and the number of protons was 243.Using these definitions, the routines given by Ogawa et al. [5]
(Step3D), Boxerman et al. [7]
(Gauss1D), and Weisskoff et al. [4]
(Gauss1D) were explored. Gauss1D, of course, is a hybrid of Step1D and Gauss3D.
The arbitrarily chosen direction must be determined from random values lying
inside a sphere (except the origin). To determine the direction, the values have
been projected onto the unit sphere's surface. The time
step was chosen as either 10 microseconds or 200 microseconds for the GE and SE signal relaxation rate computations. Initial positioning and the Monte Carlo steps were ruled by
two separate random generators from Press et al., that is, using the same
starting conditions, the results can be reproduced completely. Computations
have been repeated and mean values have been calculated to reduce numerical
noise due to the finite-sized sample of proton spins.The model consists of the following steps:place a cylinder with a given orientation relative
to B0 in a volume of interest;distribute protons randomly in that volume of
interest;compute the field at the location of each
proton as a superposition of the field generated by the cylinder;(optional) add the gradient fields from the
MR sequence at the location of each proton (e.g., to simulate a CPMG echo train);advance the phase of each proton according to
its local field and, in case of a 180° pulse for HSE, invert the direction of
phase accumulation;advance each proton in an MC step in an
arbitrary direction (in the case of a 3D step chosen from an arbitrary position
on the surface of a unit sphere);if the proton has transgressed the cylinder wall, repeat step (6) with a given probability, which is determined by the vessel
wall porosity (in our case: repeat step (6) always);if the proton has left the volume of
interest, it might reenter, depending on suitable boundary conditions.The resulting signal was computed by summating
all (normalized) complex magnetizations of the EV spins. Spins which did not
contribute (spins inside the vessel) were ignored.In (4), the cosine of the angle is computed as a scalar vector
product. This, together with expressing the trigonometric functions by a cosine
function, prevents the evaluation of trigonometric series, which is highly time consuming.Instead of one cylinder, we also used 5 and 9 cylinders as illustrated,
for example, by Kennan et al. for 5 cylinders [6].As conditions at the limits of the computational domain, we used
periodical boundary conditions (with the spins re-entering the computational
domain from the adjacent side of the domain), reflecting boundary conditions
(with the spins being reflected at the limits of the domain), and free boundary
conditions (where the spins are free to leave the domain).Background gradients can be added to the exterior field at any time
during the computations, so that diffusion-sensitive sequences like CPMG can be
established.The occurrence of vessel direction varies with ,
where is the angle between the exterior field and
the vessel orientation. For our computations, we used 6, 9, and 18 angles,
equally distributed between 0 and 90°. The total signal from vessels with
varying orientation thus becomes [5]
The relaxation rate R2 was computed using a two-point evaluation
function at TE = 16 milliseconds and TE = 40 milliseconds assuming [5].
4. NUMERICAL RESULTS
Testing of the diffusion process is shown in Figures 2and 3: Figure 2 shows the distribution of spins after a
fixed diffusion time TD together with the theoretical distribution. Figure 3 shows the CPMG signal decay together
with its analytical course. Both the local spin distribution as well as the
CPMG signal decay follows the theoretical predictions with some random noise
due to the limited number of protons.
Figure 2
The radial spin position
distribution shown after diffusion at time interval TD. The protons all started
from the origin at T = 0 and propagated in space corresponding to the different
Monte Carlo step procedures Step1D, Step3D, and Gauss3D.
Figure 3
The analytical versus the computed
decay curve of the transverse magnetization shown for a CPMG set of parameters.
Results for the computation of R2/R2* for GRE and HSE imaging are shown
in Figure 4. The curves are the mean
values of 8 computations, each with a different random series initialization. Time
step length has been set at 10 microseconds within these simulations. As can be seen, HSE
relaxation rates are greatest for small radii, whereas for GRE large vessel radii
have the highest relaxation rates. In addition, as expected, the relaxation
rate both for GRE and HSE increase with blood oxygenation and field strength.
Higher susceptibility for HSE results in shifting the maximal relaxation rate
to even smaller radii.
Figure 4
R2/R2* relaxation rate (in s
−1)
for SE, GE, from deoxygenation corresponding to a frequency of 1.4454 Hz to 65
× 1.4454 Hz. Data are shown in steps of 8 × 1.4454 Hz. The time step length in
these computations was 10 microseconds. The mean value for 8 computations with different
random series initialization is shown.
Comparison of these results to computations with a time step length of 200 microseconds is shown in Figure 5. They agree with the results using the
smaller time steps. The 200-microsecond data are slightly noisier since there is no averaging
as in the 10-microsecond data, which have been computed 8 times for different random
series initialization. Comparisons to a Gauss3D diffusion step computation and to
a free boundary condition computation are also shown (time step 200 microseconds in both).
Only the data for the free boundary computation are slightly inconsistent with
the others.
Figure 5
R2/R2* relaxation rate (in s
−1)
for time step length of 10 microseconds versus 200 microseconds. Data for the Gauss3D step and for
the free boundary conditions are also shown. The free boundary condition data
clearly deviate from the other data.
Figure 6 compares the results of the 8
computations with a computation using 9 or 18 angles instead of 6 (time step
200 microseconds each). A nonnegligible deviation of the GE curve, and a slightly
smaller deviation also for the SE curve can be seen (in the static dephasing
regime: a deviation of 5% for 6 angles, i.e., a deviation of 2.5% for 9 compared
to 18 angles). Again, data from the 9-angle versus 18-angle comparison have been
computed only once.
Figure 6
R2/R2* relaxation rates (in s
−1)
shown for 9 angles and 18 angles compared to the data for 6 angles between the
exterior field and the vessel direction.
In Figure 7, neighbor rods are taken into account
and the result is compared to the result of the 8 computations with different
random series initialization. Data for the 5 or 9 next neighbors have been
computed once. There are no apparent differences compared to the computations
with only one vessel.
Figure 7
R2/R2* relaxation rates (in s
−1)
shown for 5 vessels (central vessel and 4 next neighbors), respectively, shown for
9 vessels (8 next neighbors) versus data for one single vessel. All neighbor
vessels possess the same direction as the central vessel.
Finally, in Figure 8, the data for different
numbers of spins in the computational domain can be found. The numbers of spins
distributed in the computational domain were 243, 303, 363, 423, and 483 (time step 200 microseconds each). A change in the number of spins does not result in any significant
difference between the calculated relaxation rates.
Figure 8
R2/R2* relaxation rates (in s
−1)
shown for the computational set up with edge lengths of the computational cell
of 24, 30, 36, 42, and 48 mean spin-to-spin distances along each edge,
corresponding to 243, 303, 363, 423 and 483spins in each
computational cell.
5. DISCUSSION
When compared to lower field strengths, the maxima of the spin-echo
decay were found to be shifted to lower radii. A set-up with too few angles as
well as the use of free boundary conditions introduces errors and therefore
should be avoided. All other parameters (time-step, number of protons, and
influence of neighboring vessels of the same orientation, choice of the MC step
model used) did not influence the results in our computations.The current loop approach according to Biot Savart proved the use of the
dipole formulae for a finite voxel element to be correct [2]. The plot of the field distribution of an infinite vessel is in agreement with the field distribution of a very long
discretized vessel. However, we found that results for a discretized cylinder
slightly deviate from the results for an infinitely long cylinder. For
consistency, we show only results obtained for the infinite cylinder in the
present work. Moreover, the CPMG decay experiment showed that our diffusion
modeling works quite well for all MC step methods investigated.As discussed by Ogawa et al. [5],
at the limit of infinite numbers of spins and time steps, the signal in (6) would be real. However, because we
use a finite number of proton spins, there is a residual complex part of the
signal. In (6), we forced S to be real by taking its absolute value.
However, as the number of spins increases, the residual part should diminish
and, at the limit of infinite numbers of protons, should vanish.The precursor modeling approaches according to Balac et al. [10], Bhagwandien et al. [11],
Lüdeke et al. [12], and
Bakker et al. [13] are all limited in
that they rely either on singular analytical solutions for spheres and cylinders,
or on complicated procedures such as boundary element methods (BEMs), finite
differences (FDs), or finite element methods (FEMs). The dipole model, in contrast,
uses arbitrarily chosen elements (discretized to a grid) in combination with
simple geometries (spheres, finite cylinders, parallelepipeds). This allows the
design of special interventional instruments (e.g., a biopsy needle with
markers) and the computation of the decay of the transverse magnetization in “real
world” geometry as in a trabecular bone model, or in a vessel network to
model signal decay in brain parenchyma. Branches of vessels
might be simulated using our model. Use of analytic field distributions remains
possible, thereby allowing modeling similar to Boxerman et al. [7], Ogawa et al. [5],
Kennan et al. [6], or Fujita [8].Discretization of the susceptibility distribution can be made more
precise than the mean distance between protons by introducing distance factors [14]. An estimate of arbitrarily high
resolution can then be achieved by Richardson Extrapolation [9].Intra- and extravascular pools were separated by a routine which tests
whether a spin is intra- or extravascular. The same routine can be used to
detect the transgression of a boundary, and in the case of impenetrable vessel
walls, the diffusion step can then be rejected until a transgression-free step has
been attained. This is achieved by testing whether a spin was extravascular
before the MC step but becomes intravascular after it, or vice versa. Actually,
vessel walls are to be regarded as impenetrable, because exchange rates are
much longer (typically 500 milliseconds) than TE (typically 100 milliseconds) as discussed in Fisel
et al. [3], and in Boxerman et al. [7]. Using a rejection
with probability P (drawn from a random series), vessel walls might be modeled
as partially penetrable. Spins belonging to the intravascular pool are not
considered to contribute to the BOLD signal.The boundaries of the computational domain might be treated as periodical
boundary conditions, reflection of spins at the boundaries or by no conditions
at all. However, we observed that tests with such penetrable boundaries (also
called “free boundary conditions”) lead to slightly false results.Let us summarize an important, specific finding revealed in Figure 4 of Section 3. At high field
strength (9.4 T), spin-echo imaging is especially sensitive at about 3 μm
radii. The maximum of the spin-echo decay is shifted to lower radii than with
lower field strengths. This makes spin-echo imaging especially sensitive to
small vessels and capillaries.The computation with 18 angles revealed a source of discretization error.
The 6 angles used regularly in our computations essentially yield only an
approximate result. However, since the effort to compute with 18 angles is 3
times greater, we have restricted our analysis to 6 angles. One should bear in
mind that this leads to a (somehow limited and therefore minor) error in R2/R2*. Moreover, we found that placing vessels in the next 6 or 9 neighbor
volumes does not change the result. Vessels with orientations different from
the orientation of the central vessel in the next neighbor volumes have not
been investigated and are left for future work.From 243 protons on, the
result does not change significantly with increasing numbers of protons.
However, to obtain smoother curves (less numerical noise due to a limited
number of protons), computations might be repeated using different random
initializations for the MC step procedure. Computing time largely depends on
the hardware resources available. For the computations with TE = 40 milliseconds in steps
of 10 microseconds, computing the complete parameter-space for one random series
initialization value with 243 protons took approximately
8 days on a Pentium machine with a clock rate of 1.8 GHz.The model might be further improved by introducing Bloch's equations instead
of considering a constant transverse magnetization. Pulses and gradients can
then be simulated to model a real pulse sequence with all its implications regarding
magnetization. The diffusion constant might also be replaced by a diffusion
tensor, making the treatment of problems based on diffusion anisotropy
possible.
Authors: C R Fisel; J L Ackerman; R B Buxton; L Garrido; J W Belliveau; B R Rosen; T J Brady Journal: Magn Reson Med Date: 1991-02 Impact factor: 4.668