In this approach, a pneumatic flow injection-tandem spectrometer system, without a delivery pump, has been developed to study chromium speciation. In this system, suction force of pneumatic nebulizer of a flame atomic absorption spectrometer has been used for solution delivery through the manifold. Cr(VI) and total Cr concentrations were determined using UV-Vis and FAAS spectrometers, respectively. The Cr(III) was determined by difference. The calibration curves were linear up to 10 mug mL-1 and 20 mug mL-1 for Cr(VI) and total Cr with detection limit of 0.12 mug mL-1 and 0.07 mug mL-1 for Cr(VI) and Cr(III), respectively. The midrange precision and accuracy are less than 1.98% and +/- 2.50% for two species, respectively, at a sampling rate of 100 h-1. This system was applied for the determination of the chromium species in spiked and natural waters as well as industrial waters.
In this approach, a pneumatic flow injection-tandem spectrometer system, without a delivery pump, has been developed to study chromium speciation. In this system, suction force of pneumatic nebulizer of a flame atomic absorption spectrometer has been used for solution delivery through the manifold. Cr(VI) and total Cr concentrations were determined using UV-Vis and FAAS spectrometers, respectively. The Cr(III) was determined by difference. The calibration curves were linear up to 10 mug mL-1 and 20 mug mL-1 for Cr(VI) and total Cr with detection limit of 0.12 mug mL-1 and 0.07 mug mL-1 for Cr(VI) and Cr(III), respectively. The midrange precision and accuracy are less than 1.98% and +/- 2.50% for two species, respectively, at a sampling rate of 100 h-1. This system was applied for the determination of the chromium species in spiked and natural waters as well as industrial waters.
The demand
for fast, reliable analytical methods for determining
different species of an element in environmental samples has rapidly increased, because in most cases
the different biological, nutritional, or toxicological properties depend
critically on chemical form or oxidation state [1, 2]. Chromium is a naturally
occurring element, mainly found in minerals, rocks, plants, soil, and water in
volcano dust and gases. Although, chromium is known to exist in all oxidation
states from Cr(0) to Cr(VI), the Cr(III) and Cr(VI) species are the most
widespread in the nature. Trivalent Cr(III) and hexavalent Cr(VI) enter the
environment as a result of effluent discharge, electroplating, tanning industries
and oxidative dyeing [3]. Chromium(III) is considered as an
essential micronutrient for humans playing a role in the maintenance of normal
glucose, cholesterol, and fatty acid metabolism, whereas Cr(VI) is highly toxic
than Cr(III). Its acute toxic effects include immediate cardiovascular shock,
with later effects on kidney, liver, and blood-forming organs. The toxic nature
of the Cr(VI) ions is attributed to their high oxidation potential and their
relatively small size, which enables them to penetrate through biological cell
membranes [4]. Hence, the speciation of chromium has special importance in
several fields of life.Up to now, several methods for speciation of chromium(VI) and chromium(III) have been investigated and published in different journals such as catalytic
cathodic stripping voltammetry (CCSV) [5], X-ray absorption
near-edge structure (XANES) spectroscopy [6], the separation or preconcentration methods using capillary electrophoresis [7], sorption [8, 9], liquid-liquid extraction [10, 11], extraction using ion-pair formation [12, 13], coprecipitation [14], extraction
using supported liquid membranes (SLM) [15], solid phase extraction [16, 17, 19],
ion exchange [18], cloud point extraction [20] followed by instrumental
analysis such as UV-Vis [7, 13], energy dispersive X-ray fluorescence
spectrometry [9], GFAAS [10, 11, 15–17], FAAS [14, 19], ion
chromatography [18],
and high-performance liquid chromatography (HPLC) [20]. The
flow injection methods with conjunction by several detection techniques were also
applied for the chromium speciation [3, 4, 21–28]. However, there are several disadvantages of using these
methods such as using complicated chemical systems and techniques, expensive
reagents and instruments such as ICP and ICP/MS, low sample frequency, and also
two determination needs for each sample; first determining one of the species,
followed by reduction/oxidation of the corresponding redox form and
quantification of the total amounts of chromium or other species.Flow techniques present several useful
analytical features, in particular the possibility to attain high sample
throughput, little sample handling, or manipulation and this eliminates many of
the stringent clean particles often necessary for standard chromium determinations.
FI constitutes the most advanced form of solution manipulation available to
analytical chemists for mixing and transporting the reagents and products of a
chemical reaction to the point of measurement. Four basic sections of an FIA
system are as follows pump, injector valve-loop, reactor,
and detector. Pump is the heart of an FIA system. Different kinds of pumps such as
peristaltic
pump, HPLC pump, and syringe pump have been
used in all FIA systems [29, 30].In a flame atomic absorption spectroscopy, an
aqueous solution of the sample is nebulized as a fine spray and then mixed with
gaseous fuel and oxidant that carry it into a burner. The most commonly used
nebulization devices are pneumatic devices in which a jet of compressed gas
aspirates and nebulizes the solution. The transport of solution to the
nebulizer tip is known as aspiration. With the concentric nebulizer, the
nebulization gas flows through an opening that concentrically surrounds the
capillary tube, causing a reduced pressure at the tip and thus suction of the
sample solution from the container. It is known as Bernoulli effect. In most cases,
the flow of solution is laminar, and the aspiration rate is proportional to the
pressure drop along the capillary and to the fourth power of the capillary diameter;
it is inversely proportional to the viscosity of solution. The solution drawn
up by the capillary tube encounters the high velocity of the nebulizing gas, which
causes the formation of droplets. The efficiency and the droplet size
distribution mostly depend on the diameter and the relative position of the end
of the capillary and the nose-piece. The effect of the sample uptake rate on
the absorbance has been studied previously [31]. A maximum is obtained between
flow rates of 2–6 mL min−1 when the efficiency is about 10%.Here, the authors have designed a simple,
cheap, and fast pneumatic flow injection analysis-tandem spectrometer (PFIA-TS)
system to work without usual pumps for the speciation of Cr(VI) and
Cr(III). The basic element of PFIA-TS is using suction force of a flame atomic
absorption spectrometer (FAAS) pneumatic nebulizer for solution delivery
(carrier) through the FI manifold. So, usual pumps in FI systems have been
eliminated. The potentials of the automated methodology were evaluated using
the spectrophotometric monitoring of chromium as a model of chemistry. The
method is suitable for chromium speciation with only one injection and without
need to use of an oxidant/reductant in carrier stream. This technique has been
designed with the aim to combine the advantages of FIA system and two kinds of
spectrometry. In this system, Cr(VI) and total Cr concentrations were determined
using UV-Vis spectrometer and FAAS, respectively. The Cr(III) concentration is
determined by difference.
2. EXPERIMENTAL
2.1. Reagents
All reagents were prepared from analytical
reagent grade chemicals unless specified otherwise and purchased from Merck Company. All aqueous solutions were prepared with double-distilled
water (DDW). A Cr(VI) solution stock containing 0.1 g L−1 Cr(VI)
was prepared by dissolving 0.3735 g pure potassium chromate in 1 L of water. A
Cr(III) standard solution was prepared by the dilution of a Titrisol stock
solution (0.1 g L−1) Cr as chromium chloride. Working standard
solutions of Cr(VI) and Cr(III) used for calibration were prepared by the
appropriate dilution of the above solutions. The 0.002 M sym-diphenylcarbazide
(sym-DPC) with 0.015% (w/w) KNO3(as ionization suppressor) solution
was used as carrier solution in this system. This solution was prepared by
dissolving 0.625 g of sym-DPC in 500 mL acetone and 15 g KNO3 diluted to 1000 mL by 0.5 M nitric acid solution. Stock solutions of
interfering ions were prepared by dissolving the appropriate salts in water.
2.2. Apparatus
A scheme of the single-line PFIA-TS system is
presented in Figure 1. In this system, suction force of a Philips FAAS
(Model PU 9110X) pneumatic nebulizer with an N2O-acetyle flame has been used for
solution delivery (carrier) through the manifold. A six-way injection valve
(Rheodyne, Model 7125, USA) allowed the sample to be loaded directly into a 400 L
loop, and subsequently injected into the carrier stream. Manifold lines
consisted of 0.8 mm i.d. polyethylene tubing. The injection valve was kept at
the loading position for each first 5 seconds of every run to load the sample. After
that it was switched to the injection position to inject the sample to the
carrier stream. The valve was kept in the injection position for further 30 seconds
to ensure that the entire sample was flushed out of the sample loop. Next, the
valve was again switched to the loading position to fill the sample loop for
the next run. The generated products in reactor were channeled to the 10 mm in-length optical flow cell, a variable wavelength UV-Vis spectrophotometer detector (Knauer, Germany) at 548 nm,
then an FAAS with a 5 cm optical path-length (burner). The carrier flow rate in this manifold was 2.4 mL min−1. The signals from UV-Vis detector were performed with
computer via the Chromstar software (Bruker,
Germany). The light source for FAAS was a Cr hollow cathode lamp. The wavelength was set to 357.9 nm with a spectral slit-width of 0.5 nm and a lamp current of 12 mA.
Figure 1
Schematic diagram of the PFIA-TS system used for the speciation of chromium. C: carrier;
INJ: injector loop valve; R: reactor; UV-Vis: UV-Vis detector; FAAS: flame atomic absorption
spectrometer; Com: computer; W: waste.
3. RESULTS AND DISCUSSION
3.1. Optimization of the experimental condition
The performance of the proposed flow injection
system depended on the efficiency of the reaction occurring between sym-DPC and
chromium(VI) in the reactor, as well as on the reactor length, reactor
diameter (reactor i.d.), and the loop volume.The reactor length (with i.d. 0.8 mm) was
varied between 20 cm and 160 cm. The optimizations for two detectors are shown in
Figure 2. The analytical signal for UV-Vis detector significantly increases for
lengths ranging between 20 cm and 100 cm and remains nearly constant afterwards. The
response for FAAS detector decreases slightly for lengths ranging between 20 and 100
and then decreased with increasing in reactor length. The best results were
obtained at the length of 100 cm.
Figure 2
Influence of the reactor length on analytical signal: (a) UV-Vis detector,
[Cr(VI)] = 5.00 g mL−1, reactor i.d. = 0.8 mm, and loop volume = 400 L and (b) FAAS detector, [Cr] = 10.00 g mL−1, reactor i.d. = 0.8 mm, and loop volume = 400 L.
The influences of reactors i.d. were studied in
range 0.2–1.2 mm. The results for two detectors are shown in Figure 3. In reactors having an i.d. smaller than 0.8 mm, the interactions between the sample and the DPC were weak and the peak
height decreases for UV-Vis detector signal. The response of the FAAS increases
from 0.2 mm i.d. to an i.d. of 0.8 mm, remaining nearly constant afterwards.
Also, the carrier flow rate decreases with decreasing of reactors i.d. An i.d.
larger than 0.8 mm led to increase dispersion coefficients. A 0.8 mm i.d. was
chosen as the optimum diameter.
Figure 3
Influence of the reactor diameter on analytical signal: (a) UV-Vis detector,
[Cr(VI)] = 5.00 g mL−1, reactor length = 100 cm, and loop volume = 400 L and (b) FAAS detector, [Cr] = 10.00 g mL−1, reactor length = 100 cm, and loop volume = 400 L.
The response of the system was also studied for
various sample volumes injected (loop volume): 20, 80, 100, 150, 220, 300, 400,
500, 1000, and 2000 L. These optimizations are shown in Figure 4. The analytical
signal increased with an increase of the volume. Larger volumes resulted in
higher, yet broader peaks. Consequently, the sampling frequency and the peak
capacity sharply decreased. A 400 L volume was selected as the loop volume.
Figure 4
Influence of the loop volume on analytical signal: (a) UV-Vis detector,
[Cr(VI)] = 5.00 g mL−1, reactor length = 100 cm, reactor diameter =
0.8 mm and (b) FAAS detector, [Cr] = 10.00 g mL−1, reactor length =
100 cm, and reactor diameter = 0.8 mm.
The optimization ranges of the experimental conditions and
optimum value for PFIA-TS system are summarized in Table 1.
Table 1
Optimization of the experimental condition PFIA-TS system for chromium speciation.
Parameter
Study range
Optimum value
Reactor length (cm)
20–160
100
Reactor diameter (mm)
0.2–1.2
0.8
Loop volume (μL)
20–2000
400
3.2. Sample matrix interference
The effect of potentially interfering ionic
species on the determination of Cr(VI) with UV-Vis detector was investigated.
The tolerance limit was defined as the concentration of foreign ions resulting
in the error in the determination of 5 g mL−1 Cr(VI). It was
found that even a 250-fold excess of Ca2+, Mg2+, Cl−,
Br−, I−, F−, 200-fold excess of , Zn2+,
Al3+, Mn2+, Ni2+, , 120-fold excess of Cu2+, Co2+, ,
, ,
, and 45-fold
excess of Fe3+, Cd2+, VO3
−, 12-fold
excess of Hg2+, Pb2+, and CH3COO−did not
interfere.
3.3. Calibration plot and precision
A typical response of UV-Vis detector in optimum
conditions was shown in Figure 5. The results show the stability of the flow
rate and repeatability of the proposed system. The linearity of the response
was evaluated for the analyte concentration ranging from 1 to 100 g mL-1 under the above experimental conditions. The calibration graphs were linear up
to 10 g mL−1 and 20 g mL−1 for Cr(VI) and total Cr and are described by the following
equations:
where P. H.
is peak height in arbitrary unit (for UV-Vis detector signals), (Cr(VI)) and (Cr)
are the chromium(VI) and total chromium
concentrations, is the number of experimental points, and is the correlation
coefficient. The detection limits (DL) were defined as the analytes
concentration equal to 3 times the standard deviation of the most diluted
standard solution. DL were found to be 0.12 g mL−1 and 0.07 g mL−1 for Cr(VI) and Cr(III), respectively. Through validation and applications to evaluate
the analytical applicability of the method and to check its accuracy and
precision, synthetic aqueous mixtures containing Cr(VI) and Cr(III) were
analyzed. The precision and accuracy of the method were determined by 7
replicate analyses of standard solutions under optimum conditions. The
midrange precision and accuracy are and for two species,
respectively, at a sampling rate of 100 h−1. According to the results
presented in Table 2, the concentrations of Cr(VI) and Cr(III) obtained with
the proposed method are in good agreement with the expected values.
Figure 5
Typical analytical signals of UV-Vis detector for determination of Cr(VI)
with PFIA-TS system. Values above peaks are concentrations of Cr(VI) in g mL−1.
Table 2
Determination results for Cr(VI) and Cr(III)
in synthetic aqueous mixtures.
Added/μg mL−1
Total found/μg mL−1
Cr6+/μg mL−1
Cr3+/μg mL−1
Cr6+
Cr3+
Found
RSD (%)
RE (%)
Found
RSD (%)
RE (%)
0.5
4.00
4.39
0.49
1.98
−2.00
3.90
1.47
−2.50
3.00
3.00
5.98
3.05
1.51
1.66
2.93
1.69
−2.33
5.00
6.00
11.00
4.90
1.73
−2.00
6.10
0.98
1.66
7.00
2.00
8.95
6.91
1.67
−1.28
2.04
1.33
2.00
The proposed method was applied to the analyses
of various water and leather treatment plant samples collected from different
locations in Iran
.
The samples were filtered (Whatman filter no. 1) and determined by the proposed
method. Initial concentration of chromium in these samples was determined
before spiking. After spiking the samples with the known amounts of Cr(VI) and
Cr(III), excellent recoveries were obtained and no matrix interference was
observed (Table 3).
Recoveries for Cr(VI) and Cr(III) were found to be in the range of
.
Table 3
Determination results and recoveries for
Cr(VI) and Cr(II) speciations in water and west water samples collected
at different locations in Iran.
Sample
Added/μg mL−1
Cr(III) found/μg mL−1
Recovery (%)
Cr(VI) found/μg mL−1
Recovery (%)
Cr(III)
Cr(VI)
(n = 5)
(n = 5)
Drinking water Zahedan city
0.00
0.00
0.00
—
0.00
—
2.00
3.00
1.98
99.0
3.04
101.3
Well water
0.00
0.00
0.00
—
0.00
—
3.00
3.00
2.97
99.0
3.02
100.7
4.00
4.00
3.94
98.5
4.05
101.3
Lether treatment plant 1
0.00
0.00
1.33
—
0.00
—
4.00
2.00
5.40
101.8
1.96
98.0
2.00
3.00
3.39
103.0
2.90
96.7
Lether treatment plant 2
0.00
0.00
3.71
—
0.00
—
3.00
2.00
6.65
98.0
2.05
102.5
1.00
5.00
4.75
104.0
4.90
98.0
Caspian sea
0.00
0.00
0.00
—
0.00
—
3.00
5.00
3.04
101.3
4.89
97.8
5.00
2.00
4.95
99.0
1.97
98.5
4. CONCLUSIONS
The application of the flow injection
analysis-tandem spectrometer to the speciation analysis of trace Cr(VI) and
Cr(III) in water samples has been demonstrated. In the proposed system, the
pump from usual FIA manifold has been eliminated and suction force of a flame
atomic absorption spectrometry (FAAS) pneumatic nebulizer has been used for
solution delivery through the manifold. In this system, Cr(VI) and the total Cr concentrations were determined
using UV-Vis and flame atomic absorption spectrometers, respectively. The
method is very simple, rapid, accurate, and has good sensitivity and selectivity. It
requires only small sample volumes (400 L). It provides good reproducibility
of the results (% RSD < 1.98) which is superior to other speciation methods
already described in the literature. The PFIA-TS is inexpensive, stable, and
available in every laboratory.