Literature DB >> 18266095

Inadvertent transposition of defibrillator coil terminal pins causing inappropriate ICD therapies.

Ziad F Issa1.   

Abstract

We report the case of a 65-year-old man with chronic atrial fibrillation (AF) and severe ischemic cardiomyopathy who underwent implantation of a prophylactic single-chamber implantable cardioverter-defibrillator (ICD). The patient experienced inappropriate ICD therapies due to oversensing of pectoral muscle myopotential secondary to reversal of the defibrillator coil terminal pins in the ICD header. Recognizing this possibility is important to avoid misinterpretation of spontaneous oversensing as hardware failure (e.g., lead fracture or insulation breech) and potentially unnecessary ICD system surgical intervention, including lead extraction.

Entities:  

Mesh:

Year:  2008        PMID: 18266095     DOI: 10.1007/s10840-008-9211-8

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  3 in total

1.  Provoked and spontaneous high-frequency, low-amplitude, respirophasic noise transients in patients with implantable cardioverter defibrillators.

Authors:  M O Sweeney; K E Ellison; J B Shea; J B Newell
Journal:  J Cardiovasc Electrophysiol       Date:  2001-04

2.  Inappropriate sensing of pectoral myopotentials by an implantable cardioverter defibrillator system.

Authors:  Sree Karanam; Roy M John
Journal:  Pacing Clin Electrophysiol       Date:  2004-05       Impact factor: 1.976

3.  Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems.

Authors:  Slawomir Weretka; Jochen Michaelsen; Ruediger Becker; Christoph A Karle; Frederik Voss; Thomas Hilbel; Brigitte R Osswald; Malte L Bahner; Julia C Senges; Wolfgang Kuebler; Wolfgang Schoels
Journal:  Pacing Clin Electrophysiol       Date:  2003-01       Impact factor: 1.976

  3 in total
  1 in total

1.  Unmasking of myopotential oversensing by an integrated bipolar defibrillator lead following AV node ablation.

Authors:  M Dorenkamp; L-H Boldt; F Blaschke; Y Kühnle; W Haverkamp; M Roser
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.