Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems.

Slawomir Weretka1, Jochen Michaelsen, Ruediger Becker, Christoph A Karle, Frederik Voss, Thomas Hilbel, Brigitte R Osswald, Malte L Bahner, Julia C Senges, Wolfgang Kuebler, Wolfgang Schoels.   

Abstract

UNLABELLED: Modern dual chamber ICD systems are able to overcome various sensing problems. However, improvement of their performance is still required. The aim of this study was to assess the sensing function in 101 consecutive patients (84 men, 17 women; mean age 63 +/- 12 years; mean follow-up 24 +/- 4 months) implanted with dual chamber defibrillators and integrated (IB) or dedicated bipolar (DB) lead systems. Follow-up data were analyzed for the presence of ventricular oversensing. Oversensing occurred in 25 (25%) patients, significantly more frequent in patients implanted with IB compared to DB lead systems (21/52 vs 4/49, P = 0.0002). Patients with cardiomyopathies (CMs) were more prone to sensing malfunctions than patients with no CM (12/30 vs 13/71, P = 0.04). T wave oversensing (n = 14), respirophasic ventricular oversensing (n = 4), and P wave oversensing (n = 6) were the most common pitfalls of ventricular sensing. P wave oversensing was unique to the IB lead system. CT scans performed in these patients disclosed the position of the RV coil to be proximal to the tricuspid area. Four patients received inappropriate ICD shocks due to oversensing. In all but two patients who received lead revision, oversensing was resolved by noninvasive means. IN
CONCLUSION: (1) ventricular oversensing is a common problem occurring in up to 25% of patients with dual chamber ICDs; (2) P wave oversensing is a ventricular sensing problem affecting function of 11% of dual chamber devices with IB lead systems; (3) IB leads are significantly more susceptible to T wave and P wave oversensing than DB leads; and (4) patients with cardiomyopathies are more prone to oversensing than patients with other heart diseases.

Entities:  

Mesh:

Year:  2003        PMID: 12685142     DOI: 10.1046/j.1460-9592.2003.00152.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  16 in total

1.  Use of "IEGM Online" in ICD patients--early detection of inappropriate classified ventricular tachycardia via home monitoring.

Authors:  Oliver Ritter; Wolfgang R Bauer
Journal:  Clin Res Cardiol       Date:  2006-06-20       Impact factor: 5.460

2.  Inappropriate ICD discharge due to T-wave oversensing in a patient with the Brugada syndrome.

Authors:  Abolfath Alizadeh; Majid Haghjoo; Arash Arya; Amir Farjam Fazelifar; Mohammad Alasti; Ata Alah Bagherzadeh; Mohammad Ali Sadr-Ameli
Journal:  J Interv Card Electrophysiol       Date:  2006-01       Impact factor: 1.900

3.  [Progression of heart failure during cardiac resynchronization therapy: what is the cause?].

Authors:  C W Israel; M Britten
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-03

4.  Myocardial ischemic hyperacute T-wave oversensing leading to a defibrillator shock storm.

Authors:  Lovely Chhabra; Nauman Khalid; Jeffrey Kluger
Journal:  Proc (Bayl Univ Med Cent)       Date:  2015-04

5.  [Avoidance of intermittent T-wave oversensing with device programming].

Authors:  M Wieczorek; R Hoeltgen; I Djajadisastra
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2006-06

6.  [Current status and problems of the entirely subcutaneous ICD (S-ICD®)].

Authors:  F Reinke; A Löher; J Köbe; L Eckardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-09

7.  Unmasking of myopotential oversensing by an integrated bipolar defibrillator lead following AV node ablation.

Authors:  M Dorenkamp; L-H Boldt; F Blaschke; Y Kühnle; W Haverkamp; M Roser
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06

8.  An unusual resolution of T-wave oversensing in an implantable cardioverter defibrillator in a child with long QT syndrome.

Authors:  Mitchell I Cohen; Petros V Anagnostopoulos; Andrew Papez
Journal:  J Interv Card Electrophysiol       Date:  2009-01-16       Impact factor: 1.900

9.  T-wave Oversensing with Inappropriate Therapy in Remote Monitoring.

Authors:  Manuel Frutos; Alonso Pedrote; Eduardo Arana; Juan Sanchez-Brotons
Journal:  Indian Pacing Electrophysiol J       Date:  2010-06-05

10.  Inadvertent transposition of defibrillator coil terminal pins causing inappropriate ICD therapies.

Authors:  Ziad F Issa
Journal:  J Interv Card Electrophysiol       Date:  2008-02-12       Impact factor: 1.900

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.