Literature DB >> 18263952

A computationally efficient method for automatic registration of orthogonal x-ray images with volumetric CT data.

Xin Chen1, Martin R Varley, Lik-Kwan Shark, Glyn S Shentall, Mike C Kirby.   

Abstract

The paper presents a computationally efficient 3D-2D image registration algorithm for automatic pre-treatment validation in radiotherapy. The novel aspects of the algorithm include (a) a hybrid cost function based on partial digitally reconstructed radiographs (DRRs) generated along projected anatomical contours and a level set term for similarity measurement; and (b) a fast search method based on parabola fitting and sensitivity-based search order. Using CT and orthogonal x-ray images from a skull and a pelvis phantom, the proposed algorithm is compared with the conventional ray-casting full DRR based registration method. Not only is the algorithm shown to be computationally more efficient with registration time being reduced by a factor of 8, but also the algorithm is shown to offer 50% higher capture range allowing the initial patient displacement up to 15 mm (measured by mean target registration error). For the simulated data, high registration accuracy with average errors of 0.53 mm +/- 0.12 mm for translation and 0.61 +/- 0.29 degrees for rotation within the capture range has been achieved. For the tested phantom data, the algorithm has also shown to be robust without being affected by artificial markers in the image.

Entities:  

Mesh:

Year:  2008        PMID: 18263952     DOI: 10.1088/0031-9155/53/4/010

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Automated 2D-3D registration of portal images and CT data using line-segment enhancement.

Authors:  Reshma Munbodh; Zhe Chen; David A Jaffray; Douglas J Moseley; Jonathan P S Knisely; James S Duncan
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

2.  A comparative analysis of intensity-based 2D-3D registration for intraoperative use in pedicle screw insertion surgeries.

Authors:  Hooman Esfandiari; Carolyn Anglin; Pierre Guy; John Street; Simon Weidert; Antony J Hodgson
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-07-10       Impact factor: 2.924

3.  Validation for 2D/3D registration. I: A new gold standard data set.

Authors:  S A Pawiro; P Markelj; F Pernus; C Gendrin; M Figl; C Weber; F Kainberger; I Nöbauer-Huhmann; H Bergmeister; M Stock; D Georg; H Bergmann; W Birkfellner
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

4.  Fast volumetric registration method for tumor follow-up in pulmonary CT exams.

Authors:  José Silvestre Silva; João Cancela; Luísa Teixeira
Journal:  J Appl Clin Med Phys       Date:  2011-02-02       Impact factor: 2.102