An ultrasonic applicator, which utilizes both electronic and variable geometric focusing, for deep-localized hyperthermia is investigated. The applicator is based around a linear phased array that furnishes its electronic focusing capability. The output of the array radiates through a spherical liquid-lens that provides the applicator a variable geometric focusing capability as well. A lens of this type adds dynamic focusing to the elevation dimension of the linear phased array. By controlling the volume of liquid in the lens (and thus the radius of curvature of its membrane), dynamic control of the geometrical focus can be achieved. Comparisons of computer simulations and experimental measurements of the field intensity distribution of a small-scale prototype applicator are presented. Important design parameters, such as the choice of the liquid for the lens and the size and number of array elements, are examined.Entities:
Year: 1992 PMID: 18263147 DOI: 10.1109/58.139125
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725