Literature DB >> 18252903

Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas.

Stuart A Weinzimer1, Garry M Steil, Karena L Swan, Jim Dziura, Natalie Kurtz, William V Tamborlane.   

Abstract

OBJECTIVE: The most promising beta-cell replacement therapy for children with type 1 diabetes is a closed-loop artificial pancreas incorporating continuous glucose sensors and insulin pumps. The Medtronic MiniMed external physiological insulin delivery (ePID) system combines an external pump and sensor with a variable insulin infusion rate algorithm designed to emulate the physiological characteristics of the beta-cell. However, delays in insulin absorption associated with the subcutaneous route of delivery inevitably lead to large postprandial glucose excursions. RESEARCH DESIGN AND METHODS: We studied the feasibility of the Medtronic ePID system in youth with type 1 diabetes and hypothesized that small manual premeal "priming" boluses would reduce postprandial excursions during closed-loop control. Seventeen adolescents (aged 15.9 +/- 1.6 years; A1C 7.1 +/- 0.8%) underwent 34 h of closed-loop control; 8 with full closed-loop (FCL) control and 9 with hybrid closed-loop (HCL) control (premeal priming bolus).
RESULTS: Mean glucose levels were 135 +/- 45 mg/dl in the HCL group versus 141 +/- 55 mg/dl in the FCL group (P = 0.09); daytime glucose levels averaged 149 +/- 47 mg/dl in the HCL group versus 159 +/- 59 mg/dl in the FCL group (P = 0.03). Peak postprandial glucose levels averaged 194 +/- 47 mg/dl in the HCL group versus 226 +/- 51 mg/dl in the FCL group (P = 0.04). Nighttime control was similar in both groups (111 +/- 27 vs. 112 +/- 28 mg/dl).
CONCLUSIONS: Closed-loop glucose control using an external sensor and insulin pump provides a means to achieve near-normal glucose concentrations in youth with type 1 diabetes during the overnight period. The addition of small manual priming bolus doses of insulin, given 15 min before meals, improves postprandial glycemic excursions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18252903     DOI: 10.2337/dc07-1967

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  196 in total

1.  Clinical hurdles and possible solutions in the implementation of closed-loop control in type 1 diabetes mellitus.

Authors:  Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

2.  Postprandial glycemic excursions with the use of a closed-loop platform in subjects with type 1 diabetes: a pilot study.

Authors:  Arianne C van Bon; Jeroen Hermanides; Robin Koops; Joost B L Hoekstra; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2010-07-01

Review 3.  Toward closing the loop: an update on insulin pumps and continuous glucose monitoring systems.

Authors:  Tandy Aye; Jen Block; Bruce Buckingham
Journal:  Endocrinol Metab Clin North Am       Date:  2010-09       Impact factor: 4.741

4.  Pramlintide but Not Liraglutide Suppresses Meal-Stimulated Glucagon Responses in Type 1 Diabetes.

Authors:  Alfonso Galderisi; Jennifer Sherr; Michelle VanName; Lori Carria; Melinda Zgorski; Eileen Tichy; Kate Weyman; Eda Cengiz; Stuart Weinzimer; William Tamborlane
Journal:  J Clin Endocrinol Metab       Date:  2018-03-01       Impact factor: 5.958

5.  Combining glucose monitoring and insulin delivery into a single device: current progress and ongoing challenges of the artificial pancreas.

Authors:  Kathleen H Ang; William V Tamborlane; Stuart A Weinzimer
Journal:  Expert Opin Drug Deliv       Date:  2015-09-07       Impact factor: 6.648

6.  Preliminary evaluation of a new semi-closed-loop insulin therapy system over the prandial period in adult patients with type 1 diabetes: the WP6.0 Diabeloop study.

Authors:  Marie Aude Quemerais; Maeva Doron; Florent Dutrech; Vincent Melki; Sylvia Franc; Michel Antonakios; Guillaume Charpentier; Helene Hanaire; Pierre Yves Benhamou
Journal:  J Diabetes Sci Technol       Date:  2014-08-04

7.  Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas.

Authors:  Daniel P Howsmon; Nihat Baysal; Bruce A Buckingham; Gregory P Forlenza; Trang T Ly; David M Maahs; Tatiana Marcal; Lindsey Towers; Eric Mauritzen; Sunil Deshpande; Lauren M Huyett; Jordan E Pinsker; Ravi Gondhalekar; Francis J Doyle; Eyal Dassau; Juergen Hahn; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2018-02-01

8.  Postprandial fuzzy adaptive strategy for a hybrid proportional derivative controller for the artificial pancreas.

Authors:  Aleix Beneyto; Josep Vehi
Journal:  Med Biol Eng Comput       Date:  2018-05-03       Impact factor: 2.602

9.  Engineering Synthetically Modified Insulin for Glucose-Responsive Diabetes Therapy.

Authors:  Matthew J Webber; Daniel G Anderson; Robert Langer
Journal:  Expert Rev Endocrinol Metab       Date:  2015-07-18

10.  Evaluating the accuracy and large inaccuracy of two continuous glucose monitoring systems.

Authors:  Lalantha Leelarathna; Marianna Nodale; Janet M Allen; Daniela Elleri; Kavita Kumareswaran; Ahmad Haidar; Karen Caldwell; Malgorzata E Wilinska; Carlo L Acerini; Mark L Evans; Helen R Murphy; David B Dunger; Roman Hovorka
Journal:  Diabetes Technol Ther       Date:  2012-12-20       Impact factor: 6.118