Literature DB >> 18244194

Design and evaluation of a feedback based phased array system for ultrasound surgery.

D R Daum1, M T Buchanan, T Fjield, K Hynynen.   

Abstract

A driving system has been designed for phased array ultrasound applicators. The system is designed to-operate in the bandwidth 1.2 to 1.8 MHz, with independent channel power control up to 60 W (8 bit resolution) for each array element. To reduce power variation between elements, the system utilizes switching regulators in a feedback loop to automatically adjust the DC supply of a class D/E power converter. This feedback reduces the RF electrical power variation from 20% to 1% into a 16 element array. DC-to-RF efficiencies close to 70% for all power levels eliminates the need for large heat sinks. In addition to power control, each channel may be phase shifted 360 degrees with a minimum of 8 bit resolution. To ensure proper operation while driving ultrasound arrays with varying element sizes, each RF driving channel implements phase feedback such that proper phase of the driving signal is produced either at the amplifier output before the matching circuitry or after the matching circuitry at the transducer face. This feedback has been experimentally shown to increase the focal intensities by 20 to 25% of two tested phased arrays without array calibration using a hydrophone.

Year:  1998        PMID: 18244194     DOI: 10.1109/58.660153

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  13 in total

1.  Adaptive Real-Time Closed-Loop Temperature Control for Ultrasound Hyperthermia Using Magnetic Resonance Thermometry.

Authors:  L Sun; C M Collins; J L Schiano; M B Smith; N B Smith
Journal:  Concepts Magn Reson Part B Magn Reson Eng       Date:  2005-10-28       Impact factor: 1.176

2.  Longitudinal and shear mode ultrasound propagation in human skull bone.

Authors:  P J White; G T Clement; K Hynynen
Journal:  Ultrasound Med Biol       Date:  2006-07       Impact factor: 2.998

3.  Local frequency dependence in transcranial ultrasound transmission.

Authors:  P J White; G T Clement; K Hynynen
Journal:  Phys Med Biol       Date:  2006-04-19       Impact factor: 3.609

4.  Magnetic resonance acoustic radiation force imaging.

Authors:  Nathan McDannold; Stephan E Maier
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

5.  Transcostal high-intensity-focused ultrasound: ex vivo adaptive focusing feasibility study.

Authors:  J-F Aubry; M Pernot; F Marquet; M Tanter; M Fink
Journal:  Phys Med Biol       Date:  2008-05-12       Impact factor: 3.609

6.  Transcranial passive acoustic mapping with hemispherical sparse arrays using CT-based skull-specific aberration corrections: a simulation study.

Authors:  Ryan M Jones; Meaghan A O'Reilly; Kullervo Hynynen
Journal:  Phys Med Biol       Date:  2013-06-27       Impact factor: 3.609

Review 7.  Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.

Authors:  Kullervo Hynynen; Ryan M Jones
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

8.  HIFU Power Monitoring Using Combined Instantaneous Current and Voltage Measurement.

Authors:  Chris Adams; James R McLaughlan; Thomas M Carpenter; Steven Freear
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-09-12       Impact factor: 2.725

9.  Lateral mode coupling to reduce the electrical impedance of small elements required for high power ultrasound therapy phased arrays.

Authors:  Kullervo Hynynen; Jianhua Yin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-03       Impact factor: 2.725

10.  Field Characterization and Compensation of Vibrational Nonuniformity for a 256-Element Focused Ultrasound Phased Array.

Authors:  Mohamed A Ghanem; Adam D Maxwell; Wayne Kreider; Bryan W Cunitz; Vera A Khokhlova; Oleg A Sapozhnikov; Michael R Bailey
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-06-27       Impact factor: 2.725

View more