Literature DB >> 18238836

Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study.

Y Enomoto1, T Asai, T Arai, K Kamishima, Y Okuda.   

Abstract

BACKGROUND: We studied whether laryngoscopy and tracheal intubation were easier when using the Pentax-AWS (Tokyo, Japan), a new videolaryngoscope, than when applying the Macintosh laryngoscope, during manual in-line neck stabilization.
METHODS: In 203 anaesthetized patients with manual in-line neck stabilization, we inserted the Pentax-AWS and a Macintosh laryngoscope, in turn, and recorded the view of the glottis and time taken to laryngoscopy. The success rate of tracheal intubation (within 120 s) and time to intubation were also recorded.
RESULTS: The view of the glottis was significantly better with the Pentax-AWS than with the Macintosh laryngoscope (P<0.001). For the Macintosh laryngoscope, the view was obscured in 22 of 203 patients (11%) (Grade 3 in 21 patients and Grade 4 in one patient), whereas for the Pentax-AWS, the glottis was always clearly seen (Grade 1). Time taken to see the glottis with the Pentax-AWS [mean (sd): 6.0 (3.1) s] was significantly shorter than with the Macintosh laryngoscope [11.0 (5.0) s] (95% CI for difference: 4-6 s). The success rate of tracheal intubation with the Pentax-AWS (all of 99 patients) was significantly higher than with the Macintosh laryngoscope (93 of 104 patients) (P=0.001). Time taken for intubation was similar between the Macintosh laryngoscope [51 (27) s] and the Pentax-AWS [54 (14) s] (95% CI for difference: -9 to 3 s).
CONCLUSIONS: In patients with stabilized neck, the Pentax-AWS provided a better view of the glottis and a higher success rate of tracheal intubation, compared with the conventional Macintosh laryngoscope.

Entities:  

Mesh:

Year:  2008        PMID: 18238836     DOI: 10.1093/bja/aen002

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  32 in total

1.  Difficult tracheal intubation using the Airway Scope in a patient with unexpected mouth-opening difficulty.

Authors:  Kenji Kayashima; Hitomi Matsushita; Koji Murashima
Journal:  J Anesth       Date:  2011-12-10       Impact factor: 2.078

2.  Assessing the efficacy of video versus direct laryngoscopy through retrospective comparison of 436 emergency intubation cases.

Authors:  Benjamen M Jones; Ankit Agrawal; Thomas E Schulte
Journal:  J Anesth       Date:  2013-06-13       Impact factor: 2.078

Review 3.  [Indirect laryngoscopy : Alternatives to securing the airway].

Authors:  R R Noppens; C Werner; T Piepho
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

4.  Comparison of Pentax-AWS Airwayscope video laryngoscope, Airtraq optic laryngoscope, and Macintosh laryngoscope during cardiopulmonary resuscitation under cervical stabilization: a manikin study.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Hanako Kohama; Shin-ichi Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2011-09-07       Impact factor: 2.078

5.  Influence of clinical experience of the Macintosh laryngoscope on performance with the Pentax-AWS Airway Scope(®), a rigid video-laryngoscope, by paramedics in Japan.

Authors:  Kohei Ota; Takuma Sadamori; Shinji Kusunoki; Tadatsugu Otani; Tomoko Tamura; Kazunobu Une; Yoshiko Kida; Junji Itai; Yasumasa Iwasaki; Nobuyuki Hirohashi; Masakazu Nakao; Koichi Tanigawa
Journal:  J Anesth       Date:  2014-12-20       Impact factor: 2.078

6.  Are video laryngoscopes useful for paramedics during cardiopulmonary resuscitation?

Authors:  Takashi Asai
Journal:  J Anesth       Date:  2015-02-05       Impact factor: 2.078

7.  Clinical evaluation of C-MAC videolaryngoscope with or without use of stylet for endotracheal intubation in patients with cervical spine immobilization.

Authors:  Nidhi Gupta; Girija Prasad Rath; Hemanshu Prabhakar
Journal:  J Anesth       Date:  2013-03-11       Impact factor: 2.078

8.  Massive subcutaneous emphysema, bilateral pneumothorax, pneumomediastinum, pneumoperitoneum, pneumoretroperitoneum, and pneumoscrotum after multiple direct laryngoscopies: an autopsy case report.

Authors:  Yuko Ono; Yoshinori Okubo; Katsuhiko Hashimoto; Ryota Inokuchi; Hajime Odajima; Choichiro Tase; Kazuaki Shinohara
Journal:  J Anesth       Date:  2015-03-18       Impact factor: 2.078

9.  Comparison of the conventional Macintosh laryngoscope, the Pentax Airwayscope, and the McGrath MAC video laryngoscope under restricted cervical motion: a manikin study.

Authors:  Atsushi Kotera; Hiroki Irie; Shinsuke Iwashita; Junichi Taniguchi; Shunji Kasaoka; Yoshihiro Kinoshita
Journal:  Acute Med Surg       Date:  2014-10-20

10.  Central venous catheter malposition in the azygos vein and difficult endotracheal intubation in severe ankylosing spondylitis: a case report.

Authors:  Eunjin Moon; Hyungmo Jeong; Junyoung Chung; Jaewoo Yi
Journal:  Int J Clin Exp Med       Date:  2015-11-15