Literature DB >> 18230601

Comparison of different approaches for optimization of atrioventricular and interventricular delay in biventricular pacing.

Michel Zuber1, Stefan Toggweiler, Markus Roos, Richard Kobza, Peiman Jamshidi, Paul Erne.   

Abstract

AIMS: It has been shown that optimizing atrioventricular (AV) and interventricular (VV) delay improves cardiac performance in patients with biventricular pacemakers. However, there is no standard method for optimization available yet. The aim of this study was to compare echocardiographic parameters-displacement imaging, A wave duration, and aortic velocity time integral (VTI)-and acoustic cardiography derived electromechanical activation time (EMAT) using different approaches of AV and VV delay optimization. We tested whether the initial optimization of the AV interval followed by VV optimization at that optimal AV interval or initial optimization of the VV interval followed by AV optimization at the determined optimal VV interval was accurate and consistent, and how this compared to testing every conceivable combination of AV and VV intervals available. METHODS AND
RESULTS: A group of 20 patients with biventricular pacemakers was included. Displacement imaging, A wave duration, and aortic VTI were determined at different combinations of AV (100, 150, 200, 250 ms) and VV (RV40, 0, LV40 ms) intervals. If AV duration was determined first, displacement imaging identified the best setting in 8/20, aortic VTI in 10/20, A duration in 13/20, and EMAT in 18/20 patients. With VV duration determined first, the best setting was more difficult to identify regardless of the method used. There was a poor agreement in optimal AV and VV delays of the different methods, and there was no single patient in whom all four methods yielded the same delay combination.
CONCLUSION: It is advisable to measure a full grid of AV and VV delays to identify optimal settings rather than optimizing one of the two delays first. Different techniques for delay optimization resulted in different optimal delay combinations.

Entities:  

Mesh:

Year:  2008        PMID: 18230601     DOI: 10.1093/europace/eum287

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  6 in total

Review 1.  Current role of echocardiography in cardiac resynchronization therapy.

Authors:  Donato Mele; Matteo Bertini; Michele Malagù; Marianna Nardozza; Roberto Ferrari
Journal:  Heart Fail Rev       Date:  2017-11       Impact factor: 4.214

2.  Effects of AV-delay optimization on hemodynamic parameters in patients with VDD pacemakers.

Authors:  Konstantin A Krychtiuk; Michael Nürnberg; Romana Volker; Linda Pachinger; Rudolf Jarai; Matthias K Freynhofer; Johann Wojta; Kurt Huber; Thomas W Weiss
Journal:  Wien Klin Wochenschr       Date:  2014-03-21       Impact factor: 1.704

Review 3.  Cardiac resynchronization therapy: Dire need for targeted left ventricular lead placement and optimal device programming.

Authors:  Sokratis Pastromas; Antonis S Manolis
Journal:  World J Cardiol       Date:  2014-12-26

4.  Characterization of cardiac time intervals in healthy bonnet macaques (Macaca radiata) by using an electronic stethoscope.

Authors:  Haroon Kamran; Louis Salciccioli; Sergei Pushilin; Paraag Kumar; John Carter; John Kuo; Carol Novotney; Jason M Lazar
Journal:  J Am Assoc Lab Anim Sci       Date:  2011-03       Impact factor: 1.232

5.  Simultaneous variation of ventricular pacing site and timing with biventricular pacing in acute ventricular failure improves function by interventricular assist.

Authors:  T Alexander Quinn; Santos E Cabreriza; Marc E Richmond; Alan D Weinberg; Jeffrey W Holmes; Henry M Spotnitz
Journal:  Am J Physiol Heart Circ Physiol       Date:  2009-10-23       Impact factor: 4.733

6.  The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols.

Authors:  Charlotte H Manisty; Ali Al-Hussaini; Beth Unsworth; Resham Baruah; Punam A Pabari; Jamil Mayet; Alun D Hughes; Zachary I Whinnett; Darrel P Francis
Journal:  Circ Arrhythm Electrophysiol       Date:  2011-11-17
  6 in total