Assessment of response function in two PET scanners with and without interplane septa.

C Michel1, A Bol, T Spinks, D Townsend, D Bailey, S Grootoonk, T Jones.   

Abstract

The authors have assessed the response function both experimentally and theoretically for two commercial tomographs: CTI 931/08-12 and CTI 953B with and without interplane septa. Monte Carlo simulations were undertaken using the GEANT package from CERN. Spatial resolution (tomographic and axial) was calculated for line sources at various positions in the field of view. Sensitivity and scatter fraction (SF) were calculated for various source geometries as a function of energy discrimination. A very realistic response function in positron emission tomography (PET) is obtained by Monte Carlo methods, using global parameters to account for unsimulated phenomena such as scintillation light transport inside a detector block and its sharing among the various phototubes. Minor discrepancies remain for sensitivity and SF at high energy thresholds and may probably be explained by introducing the observed dispersion in the energy response for the various crystals within a detector block.

Year:  1991        PMID: 18222823     DOI: 10.1109/42.97571

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  2 in total

1.  Investigation of OPET Performance Using GATE, a Geant4-Based Simulation Software.

Authors:  Fernando R Rannou; Vandana Kohli; David L Prout; Arion F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2004-10       Impact factor: 1.679

2.  A study on 3D Monte Carlo modeling of photon propagation through tissue.

Authors:  M K Kiymik
Journal:  J Med Syst       Date:  1995-08       Impact factor: 4.460

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.