Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties.

Piero Ceruti1, Roberta Morbio, Michele Marraffa, Giorgio Marchini.   

Abstract

PURPOSE: To compare the measurements of intraocular pressure (IOP) with dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in eyes with corneal graft and to evaluate the influence of corneal thickness (CCT), corneal curvature (CC), and astigmatism on these methods.
DESIGN: Prospective, observational cross-sectional study.
METHODS: Eighteen eyes of 18 patients after penetrating keratoplasty (PKP) and 14 eyes of 14 patients after deep lamellar keratoplasty (DLKP) underwent IOP evaluation with DCT and GAT, and measurements of CCT, CC, and astigmatism. Bland-Altman plots were used to evaluate the agreement between tonometers. Multivariate regression analysis was used to evaluate the influence of ocular structural factors and running suture on IOP measurements obtained with both tonometers.
RESULTS: IOP values obtained by DCT and GAT were strongly correlated in all eyes (r = .91; P < .001). DCT values measured 2.5 +/- 1.7 mm Hg higher than GAT readings (P < .001). A reduction of the mean IOP difference between DCT and GAT with an increase in IOP values (P < .001) was found. Regression analysis showed no effect of CCT, CC, astigmatism, and running suture on both DCT and GAT readings, either in DLKP or in PKP eyes.
CONCLUSIONS: We found a good overall correlation between both tonometers but the agreement between instruments differs in high or low IOP ranges. The wide and varying 95% limits of agreement between DCT and GAT indicates that DCT provides IOP measurements on deep lamellar and penetrating keratoplasties which can be used in the clinical practice.

Entities:  

Mesh:

Year:  2008        PMID: 18222191     DOI: 10.1016/j.ajo.2007.10.003

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  8 in total

1.  Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty.

Authors:  M L Salvetat; M Zeppieri; F Miani; C Tosoni; L Parisi; P Brusini
Journal:  Eye (Lond)       Date:  2011-03-25       Impact factor: 3.775

2.  Evaluation of a contact lens-embedded sensor for intraocular pressure measurement.

Authors:  Michael D Twa; Cynthia J Roberts; Huikai J Karol; Ashraf M Mahmoud; Paul A Weber; Robert H Small
Journal:  J Glaucoma       Date:  2010-08       Impact factor: 2.503

3.  Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions.

Authors:  Caterina Schweier; James V M Hanson; Jens Funk; Marc Töteberg-Harms
Journal:  BMC Ophthalmol       Date:  2013-09-05       Impact factor: 2.209

4.  Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Authors:  Mahmoud Rateb; Mahmoud Abdel-Radi; Zeiad Eldaly; Mohamed Nagy Elmohamady; Asaad Noor El Din
Journal:  J Ophthalmol       Date:  2019-05-09       Impact factor: 1.909

Review 5.  Challenges in the complex management of post-keratoplasty glaucoma.

Authors:  Lisa-Marie Anders; Zisis Gatzioufas; Matthias C Grieshaber
Journal:  Ther Adv Ophthalmol       Date:  2021-07-20

Review 6.  Intraocular Pressure Measurement after Penetrating Keratoplasty.

Authors:  Otilia-Maria Dumitrescu; Sinziana Istrate; Mioara-Laura Macovei; Alina Gabriela Gheorghe
Journal:  Diagnostics (Basel)       Date:  2022-01-19

Review 7.  Glaucoma after penetrating keratoplasty.

Authors:  Mihail Zemba; Alina-Cristina Stamate
Journal:  Rom J Ophthalmol       Date:  2017 Jul-Sep

8.  Early intraocular pressure changes following different keratoplasty techniques and association with cornea parameters and anterior chamber depth.

Authors:  Gulsah Gumus; Cigdem Altan; Yusuf Yildirim; Nilay Kandemir Beşek; Selim Genç; Ahmet Kirgiz; Gonul Karatas Durusoy; Alper Ağca
Journal:  Ther Adv Ophthalmol       Date:  2022-03-19