| Literature DB >> 18215960 |
Abstract
A good statistical model of speckle formation is useful for designing a good adaptive filter for speckle reduction In ultrasound B-scan images. Previously, statistical models have been used, but they failed to account for the log compression of the echo envelope employed by clinical ultrasound systems. Log-compression helps in reducing the dynamic range of the B-scan Images for display on a monitor as well as enhancing weak backscatters. In this article, statistics of log-compressed echo images, using the K-distribution statistical model for the echo envelope, are used to derive a parameter that can be used to quantify the extent of speckle formation. This speckle quantification can be used with an unsharp masking filter to adaptively reduce speckle. The effectiveness of the filter is demonstrated on images of contrast detail phantoms and on in-vivo abdominal images obtained by a clinical ultrasound system with log-compression.Year: 1996 PMID: 18215960 DOI: 10.1109/42.544498
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048