Literature DB >> 18158311

Effect of chest compressions on the time taken to insert airway devices in a manikin.

J J Gatward1, M J C Thomas, J P Nolan, T M Cook.   

Abstract

BACKGROUND: Resuscitation guidelines recommend that chest compressions should continue throughout attempts to place airway devices. Few data support the use of the tracheal tube over supraglottic airway devices (SADs) during cardiopulmonary arrest. This study was designed to evaluate the speed with which different airway devices could be placed with and without interrupting chest compressions.
METHODS: Forty volunteer doctors regularly involved in cardiopulmonary resuscitation (CPR) were timed inserting four different airway devices [tracheal tube (TT), LMA Classic (cLMA), LMA ProSeal (PLMA), and igel] into a manikin, with and without stopping chest compressions.
RESULTS: Chest compressions delayed the placement of the TT only (3.3 s, P<0.0001). Comparison of the speed of insertion of the different airway devices during CPR enabled ranking of the devices: igel (fastest), PLMA (second), and TT and cLMA (joint slowest). The igel was inserted approximately 50% faster than the other devices. Doctors who had previously inserted more than 50 tracheal tubes were significantly faster at intubating the trachea, but no faster at inserting SADs.
CONCLUSIONS: Our results show that continuing chest compressions has a minor effect on time for tracheal intubation and until clear human data are available the recommendation to intubate without interrupting CPR is therefore justified. The PLMA and igel (SADs with a gastric drain tube) were both faster to insert than the cLMA and offer additional benefits. They should be considered for use in CPR.

Entities:  

Mesh:

Year:  2007        PMID: 18158311     DOI: 10.1093/bja/aem364

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  24 in total

Review 1.  Cardiopulmonary resuscitation and management of cardiac arrest.

Authors:  Jerry P Nolan; Jasmeet Soar; Volker Wenzel; Peter Paal
Journal:  Nat Rev Cardiol       Date:  2012-06-05       Impact factor: 32.419

2.  Validation of the Pentax-AWS Airwayscope utility as an intubation device during cardiopulmonary resuscitation on the ground.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Motoi Itani; Shin-ichi Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2010-05-19       Impact factor: 2.078

3.  Comparison of Supreme(®) and Soft Seal(®) laryngeal masks for airway management during cardiopulmonary resuscitation in novice doctors: a manikin study.

Authors:  Hanako Kohama; Nobuyasu Komasawa; Ryusuke Ueki; Aoi Samma; Masashi Nakagawa; Shin-ichi Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2010-12-01       Impact factor: 2.078

4.  Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Hajime Nomura; Motoi Itani; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2010-02-03       Impact factor: 2.078

5.  Comparison of tube-guided and guideless videolaryngoscope for tracheal intubation during chest compression in a manikin: a randomized crossover trial.

Authors:  Daisuke Okada; Nobuyasu Komasawa; Shunsuke Fujiwara; Toshiaki Minami
Journal:  J Anesth       Date:  2014-10-28       Impact factor: 2.078

6.  Utility of the Pentax-AWS Airwayscope and Macintosh laryngoscope for airway management during chest compressions in 27° left-lateral tilt: a manikin simulation study of maternal cardiopulmonary resuscitation.

Authors:  Hanako Kohama; Nobuyasu Komasawa; Ryusuke Ueki; Noriyasu Yamamoto; Chikara Tashiro; Yoshiroh Kaminoh; Shin-ichi Nishi
Journal:  J Anesth       Date:  2013-04-26       Impact factor: 2.078

7.  Evaluation of chest compression effect on airway management with air-Q, aura-i, i-gel, and Fastrack intubating supraglottic devices by novice physicians: a randomized crossover simulation study.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Yoshiroh Kaminoh; Shin-Ichi Nishi
Journal:  J Anesth       Date:  2014-02-22       Impact factor: 2.078

8.  Benefits of using an endotracheal tube introducer as an adjunct to a Macintosh laryngoscope for endotracheal intubation performed by inexperienced doctors during mechanical CPR: A randomized prospective crossover study.

Authors:  Hüseyin Cahit Halhalli; Asım Enes Özbek; Emrah Çelİk; Yavuz Yİğİt; Serkan Yilmaz; Müge Çardak
Journal:  World J Emerg Med       Date:  2019

9.  Assessment of Success and Ease of Insertion of ProSeal™ Laryngeal Mask Airway versus I-gel™ Insertion by Paramedics in Simulated Difficult Airway Using Cervical Collar in Different Positions in Manikins.

Authors:  Arushi Gupta; Ankita Kabi; Divya Gaur
Journal:  Anesth Essays Res       Date:  2021-05-06

10.  Randomised comparison of the effectiveness of the laryngeal mask airway supreme, i-gel and current practice in the initial airway management of prehospital cardiac arrest (REVIVE-Airways): a feasibility study research protocol.

Authors:  Jonathan Richard Benger; Sarah Voss; David Coates; Rosemary Greenwood; Jerry Nolan; Steven Rawstorne; Megan Rhys; Matthew Thomas
Journal:  BMJ Open       Date:  2013-02-13       Impact factor: 2.692

View more