Variability of electrocardiographic precordial lead placement: a method to improve accuracy and reliability.

M V Herman1, D A Ingram, J A Levy, J R Cook, R J Athans.   

Abstract

Variability in precordial lead placement is a recognized source of electrocardiographic inaccuracy and lack of reproducibility. In an attempt to reduce error, we evaluated a new device to facilitate and guide precordial lead placement. This study involved three phases: (1) comparison of device-guided electrocardiogram with ECGs obtained by deliberate misplacement of precordial leads on the same patient; (2) electrocardiograms obtained by using the precordial lead device versus those obtained by standard technician methods; (3) reproducibility of precordial electrocardiographic leads between two technicians using the device to guide lead placement. Deliberate misplacement of precordial leads by 2 cm resulted in significant electrocardiographic interpretation changes in all patients. Comparing electrocardiograms obtained after device-guided precordial placement with those obtained after technician placement resulted in variations in 60% of patients including changes in R-wave amplitude, ST segments, Q waves, and transition zone. Significant Q-wave appearance/disappearance and/or significant ST-segment elevation/depression occurred in 19% of patients in Phase II. Sixteen percent of electrocardiograms showed significant changes when analyzed by an experienced electrocardiographer and 10% when interpreted by computer. Variable lead placements and resulting electrocardiographic alterations were not seen by either of two technicians when the device was used. This study confirms the widespread variability in precordial electrocardiograms secondary to lead misplacement. The use of a device to assist in the placement of precordial leads ensures accuracy and reproducibility of electrocardiography. Improved precision and quality control in this laboratory test have important implications in health care and its costs.

Entities:  

Mesh:

Year:  1991        PMID: 1810683

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  5 in total

1.  The Effect of Diuresis on the Paced QRS Complexes in Pacing-Dependent Patients with Heart Failure.

Authors:  Fuqiang Sheng; Bin Chen; Maorong He; Meilin Zhang; Guoying Shen; Xiaowei Zhao
Journal:  Ann Noninvasive Electrocardiol       Date:  2015-06-24       Impact factor: 1.468

2.  Modified precordial lead ECG SafOne on electrocardiography recordings.

Authors:  Wan Nishfa Dewi; Safri Safri; Iswadi Hasyim Rosma
Journal:  Sci Rep       Date:  2022-05-13       Impact factor: 4.996

3.  The Reconstruction of a 12-Lead Electrocardiogram from a Reduced Lead Set Using a Focus Time-Delay Neural Network.

Authors:  Gerard H Smith; Dawie J Van den Heever; Wayne Swart
Journal:  Acta Cardiol Sin       Date:  2021-01       Impact factor: 2.672

Review 4.  Remote and wearable ECG devices with diagnostic abilities in adults: A state-of-the-science scoping review.

Authors:  Zeineb Bouzid; Salah S Al-Zaiti; Raymond Bond; Ervin Sejdić
Journal:  Heart Rhythm       Date:  2022-03-09       Impact factor: 6.779

5.  The effect of precordial lead displacement on ECG morphology.

Authors:  Michał Kania; Hervé Rix; Małgorzata Fereniec; Heriberto Zavala-Fernandez; Dariusz Janusek; Tomasz Mroczka; Günter Stix; Roman Maniewski
Journal:  Med Biol Eng Comput       Date:  2013-10-19       Impact factor: 2.602

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.